Retroauricular Emphysema as a Late Complication After Bonebridge Implantation: Case Report

. 2021 May ; 100 (4) : 233-236. [epub] 20210317

Jazyk angličtina Země Spojené státy americké Médium print-electronic

Typ dokumentu kazuistiky, časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid33729895

Bonebridge (BB) is the first active implantation system for bone conduction that is placed fully under the skin. Experience suggests that BB is characterized by low incidence of postoperative complications. This case report presents a rare case of a 16-year-old girl with incidence of emphysema occurring over the implant 1 year after operation. We performed a computed tomography scan that showed pockets of gas above the floating mass transducer so we provided the revision surgery and sealed the artificial opening with fat from the earlobe and fibrin glue. Since that time, no air has collected in the retroauricular area and the implant has been fully functional.