Drop coating deposition Raman (DCDR) spectroscopy of contaminants

. 2021 Dec 05 ; 262 () : 120109. [epub] 20210624

Jazyk angličtina Země Anglie, Velká Británie Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid34214739
Odkazy

PubMed 34214739
DOI 10.1016/j.saa.2021.120109
PII: S1386-1425(21)00686-7
Knihovny.cz E-zdroje

Raman spectroscopy is a useful technique to identify small organic molecules, including contaminants. The drop coating deposition Raman (DCDR) is more sensitive than conventional Raman spectroscopy from solution. It is based on Raman measurement from a small drop dried on a hydrophobic surface where studied molecules are preconcentrated. In this paper, DCDR spectra of dried drops of selected contaminants (food contaminant melamine, fungicide thiram, herbicides bentazon and picloram) on the hydrophobic substrate were acquired for the first time, whereas Raman spectra from stock solutions were impossible to obtain under the same experimental conditions. The lowest DCDR detected concentrations were determined as 6.4 µM, 0.31 µM, 20 µM and 2 µM in deposited concentrations for melamine, thiram, bentazon and picloram, respectively. Therefore, DCDR spectroscopy can serve to detect these molecules in concentrations relevant in food/groundwater contaminations.