Six-month electrical performance of the first dual-chamber leadless pacemaker

. 2024 Oct ; 21 (10) : 1929-1938. [epub] 20240430

Jazyk angličtina Země Spojené státy americké Médium print-electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid38697271
Odkazy

PubMed 38697271
DOI 10.1016/j.hrthm.2024.04.091
PII: S1547-5271(24)02525-6
Knihovny.cz E-zdroje

BACKGROUND: The first dual-chamber leadless pacemaker (DC-LP) system consists of 2 separate atrial and ventricular devices that communicate to maintain synchronous atrioventricular pacing and sensing. The initial safety and efficacy were previously reported. OBJECTIVE: The purpose of this study was to evaluate the chronic electrical performance of the DC-LP system. METHODS: Patients meeting standard dual-chamber pacing indications were enrolled and implanted with the DC-LP system (Aveir DR, Abbott), including right atrial and ventricular helix-fixation LPs (atrial leadless pacemaker [ALP], ventricular leadless pacemaker [VLP]). Pacing capture threshold, sensed amplitude, and pacing impedance were collected using the device programmer at prespecified timepoints from 0-6 months postimplant. RESULTS: De novo devices were successfully implanted in 381 patients with complete 6-month data (62% male; age 69 ± 14 years; weight 82 ± 20 kg; 65% sinus nodal dysfunction, 30% atrioventricular block). ALPs were implanted predominantly in the right atrial appendage anterior base and VLPs primarily at the mid-to-apical right ventricular septum. From implant to 1 month, pacing capture thresholds (0.4-ms pulse width) improved in both ALPs (2.4 ± 1.5 V to 0.8 ± 0.8 V; P <.001) and VLPs (0.8 ± 0.6 V to 0.6 ± 0.4 V; P <.001). Sensed amplitudes improved in both ALPs (1.8 ± 1.3 mV to 3.4 ± 1.9 mV; P <.001) and VLPs (8.8 ± 4.0 mV to 11.7 ± 4.2 mV; P <.001). Impedances were stable in ALPs (334 ± 68 Ω to 329 ± 52 Ω; P = .17) and reduced in VLPs (789 ± 351 Ω to 646 ± 190 Ω; P <.001). Electrical measurements remained relatively stable from 1-6 months postimplant. No differences in electrical metrics were observed among ALP or VLP implant locations. CONCLUSION: This first in-human evaluation of the new dual-chamber leadless pacemaker system demonstrated reliable electrical performance throughout the initial 6-month evaluation period.

Citace poskytuje Crossref.org

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...