Most cited article - PubMed ID 37069277
Clinical outcomes and survival comparison between NexGen all-poly and its metal-backed equivalent in total knee arthroplasty
BACKGROUND: Total knee arthroplasty (TKA) with all-polyethylene tibial (APT) components has shown comparable survivorship and clinical outcomes to that with metal-backed tibial (MBT). Although MBT is more frequently implanted, APT equivalents are considered a low-cost variant for elderly patients. A biomechanical analysis was assumed to be suitable to compare the response of the periprosthetic tibia after implantation of TKA NexGen APT and MBT equivalent. METHODS: A standardised load model was used representing the highest load achieved during level walking. The geometry and material models were created using computed tomography data. In the analysis, a material model was created that represents a patient with osteopenia. RESULTS: The equivalent strain distribution in the models of cancellous bone with an APT component showed values above 1000 με in the area below the medial tibial section, with MBT component were primarily localised in the stem tip area. For APT variants, the microstrain values in more than 80% of the volume were in the range from 300 to 1500 με, MBT only in less than 64% of the volume. CONCLUSION: The effect of APT implantation on the periprosthetic tibia was shown as equal or even superior to that of MBT despite maximum strain values occurring in different locations. On the basis of the strain distribution, the state of the bone tissue was analysed to determine whether bone tissue remodelling or remodelling would occur. Following clinical validation, outcomes could eventually modify the implant selection criteria and lead to more frequent implantation of APT components.
- Keywords
- All-polyethylene tibial component, Computational modeling, FEA, Finite element method, Knee replacement, Metal-backed tibial component, TKR, Total knee arthroplasty,
- MeSH
- Finite Element Analysis MeSH
- Metals MeSH
- Humans MeSH
- Polyethylene MeSH
- Prosthesis Design MeSH
- Knee Prosthesis * MeSH
- Aged MeSH
- Tibia diagnostic imaging surgery MeSH
- Arthroplasty, Replacement, Knee * MeSH
- Check Tag
- Humans MeSH
- Aged MeSH
- Publication type
- Journal Article MeSH
- Names of Substances
- Metals MeSH
- Polyethylene MeSH
Numerous studies have compared metal-backed components (MBTs) and all-polyethylene tibial components (APTs), but none of them specifically analysed the clinical results and the overall patient preference in patients who had undergone a staged bilateral knee replacement. The purpose of this study is to compare clinical results, perceived range of motion, and overall implant preference among patients who had undergone staged bilateral knee replacement with an APT and contralateral knee replacement with MBTs. A dataset of 62 patients from a single centre who underwent staged bilateral TKA between 2009 and 2022 was selected and retrospectively analysed. Tibial component removal was performed in three knees overall, all of which had MBTs. The mean measured Knee Score (KS) of knees with APTs was 78.37 and that of contralateral knees with MBTs was 77.4. The mean measured Function (FS) of knees with APTs was 78.22, and that of contralateral knees with MBs was 76.29. The mean flexion angle of knees with APTs was 103.8 and that for knees with MBTs was 101.04 degrees. A total of 54.8% of the patients preferred the knee that received APTs over contralateral MBTs. In our cohort, TKA with an APT in one knee and an MBT in the contralateral knee recorded similar clinical results and perceived ranges of motion. Patients in general preferred the knee that received an APT over contralateral knee with an MBT.
- Keywords
- all-polyethylene tibia, bilateral knee replacement, metal-backed tibia, staged bilateral knee arthroplasty, total knee arthroplasty,
- Publication type
- Journal Article MeSH