Dlouhodobé zkušenosti s kombinovaným hydroxyapatitovým povrchem ARBOND v osteointegraci implantátu
[Long-term experience with the combined ARBOND hydroxyapatite coating in implant osteointegration]

I. Landor, P. Vavřík, D. Jahoda, D. Pokorný, R. Ballay, A. Sosna

Jazyk čeština Země Česko

Perzistentní odkaz   https://www.medvik.cz/link/bmc07531089

Digitální knihovna NLK
Číslo
Ročník
Zdroj

E-zdroje Online

NLK Free Medical Journals od 2006

PURPOSE OF THE STUDY The hydroxyapatite coating of an implant surface provides osteoactive conditions that can support osteointegration of cement-less joint arthroplasties. However, the possibilities of hydroxyapatite degradation, resorption and delamination that may become responsible for failure of total hip arthroplasty (THA) have been reported. The aim of the study was to assess the properties of Arbond hydroxyapatite coating by comparing the long-term survival of implants identical in construction but different in surface coating. MATERIAL One group (HA) comprised 86 patients (100 THAs) with an average age of 45.14 years (range, 22.3 to 77.4 years) at the time of surgery who received a femoral stem (Walter) with a coating of Arbond sprayed over the proximal half. The other group (control) included 92 patients (100 THAs) with an average age of 49.7 years (range, 33 to 68.7) who had an identical femoral component without coating. In both groups the conical-shaped acetabular cup (Walter) and femoral head made of sintered ceramics were used. The patients in whom one or both components were replaced or extracted were not included in the final clinical evaluation (Harris Hip Score). For the statistical analysis of survival, a stable component still in place at the date of the revision procedure was regarded as surviving; a lose component at the same date was considered a failure. Finally, 71 hips of the HA group followed up for an average of 15.51 (range, 5.6 to 18.56) years and 39 control hips at an average follow-up of 14.19 (range, 6.24 to 18.48) years were clinically evaluated. The data of patients who died in the course of study (HA group, 11; control group, 14) were included in the clinical evaluation with the date of their last follow-up. METHODS For both groups, the Kaplan-Meier survival curves were constructed for overall survival and for the survival of acetabular and femoral components separately. Differences in survival curves were evaluated with the use of Gehan?s Wilcoxon test. Component survival was also calculated using 15-year life-table survivorship analysis. Differences in variables under study were assessed with the use of the two-tailed Student?s t-test. A p value of less than 0.05 was considered significant. RESULTS A total of 29 % hips were revised in the HA group, 27 % for aseptic loosening of the acetabular cup, in 2 % both components were removed because of deep infection. In the control group revision procedures were performed in 61 % of the hips. Except for one case of deep infection (1 %), the reason was cup loosening in 30 %, stem loosening in 12 % and both components loosening in 18 % of the hips. The HA group showed a significantly longer survival of both total hip prostheses and individual components. The final HHS was significantly better than the initial score in both groups. There was no difference in the degree of improvement between the two groups. The radiographic data showed full osteointegration of stems in the HA group. The control group, on the other hand, had 87 % of the stems with translucent lines in zone I and zone VII according to Gruen?s classification. DISCUSSION The significantly longer survival of hips in the HA group gives support to the use of hydroxyapatite coating in total hip arthroplasty. The poorer results in grit-blasted implants, as compared with the literature data, can be explained by alow degree of roughness of the Walter implant surfaces. CONCLUSIONS The combined Arbond hydroxyapatite coating improves conditions for implant osteointegration in the bone.

Long-term experience with the combined ARBOND hydroxyapatite coating in implant osteointegration

Bibliografie atd.

Lit.: 27

000      
00000naa 2200000 a 4500
001      
bmc07531089
003      
CZ-PrNML
005      
20250121143204.0
008      
091014s2009 xr e cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2009/033
040    __
$a ABA008 $b cze $c ABA008 $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Landor, Ivan, $d 1956- $7 xx0062301
245    10
$a Dlouhodobé zkušenosti s kombinovaným hydroxyapatitovým povrchem ARBOND v osteointegraci implantátu / $c I. Landor, P. Vavřík, D. Jahoda, D. Pokorný, R. Ballay, A. Sosna
246    11
$a Long-term experience with the combined ARBOND hydroxyapatite coating in implant osteointegration
314    __
$a I. ortopedická klinika 1.LF UK, FN Motol, Praha
504    __
$a Lit.: 27
520    9_
$a PURPOSE OF THE STUDY The hydroxyapatite coating of an implant surface provides osteoactive conditions that can support osteointegration of cement-less joint arthroplasties. However, the possibilities of hydroxyapatite degradation, resorption and delamination that may become responsible for failure of total hip arthroplasty (THA) have been reported. The aim of the study was to assess the properties of Arbond hydroxyapatite coating by comparing the long-term survival of implants identical in construction but different in surface coating. MATERIAL One group (HA) comprised 86 patients (100 THAs) with an average age of 45.14 years (range, 22.3 to 77.4 years) at the time of surgery who received a femoral stem (Walter) with a coating of Arbond sprayed over the proximal half. The other group (control) included 92 patients (100 THAs) with an average age of 49.7 years (range, 33 to 68.7) who had an identical femoral component without coating. In both groups the conical-shaped acetabular cup (Walter) and femoral head made of sintered ceramics were used. The patients in whom one or both components were replaced or extracted were not included in the final clinical evaluation (Harris Hip Score). For the statistical analysis of survival, a stable component still in place at the date of the revision procedure was regarded as surviving; a lose component at the same date was considered a failure. Finally, 71 hips of the HA group followed up for an average of 15.51 (range, 5.6 to 18.56) years and 39 control hips at an average follow-up of 14.19 (range, 6.24 to 18.48) years were clinically evaluated. The data of patients who died in the course of study (HA group, 11; control group, 14) were included in the clinical evaluation with the date of their last follow-up. METHODS For both groups, the Kaplan-Meier survival curves were constructed for overall survival and for the survival of acetabular and femoral components separately. Differences in survival curves were evaluated with the use of Gehan?s Wilcoxon test. Component survival was also calculated using 15-year life-table survivorship analysis. Differences in variables under study were assessed with the use of the two-tailed Student?s t-test. A p value of less than 0.05 was considered significant. RESULTS A total of 29 % hips were revised in the HA group, 27 % for aseptic loosening of the acetabular cup, in 2 % both components were removed because of deep infection. In the control group revision procedures were performed in 61 % of the hips. Except for one case of deep infection (1 %), the reason was cup loosening in 30 %, stem loosening in 12 % and both components loosening in 18 % of the hips. The HA group showed a significantly longer survival of both total hip prostheses and individual components. The final HHS was significantly better than the initial score in both groups. There was no difference in the degree of improvement between the two groups. The radiographic data showed full osteointegration of stems in the HA group. The control group, on the other hand, had 87 % of the stems with translucent lines in zone I and zone VII according to Gruen?s classification. DISCUSSION The significantly longer survival of hips in the HA group gives support to the use of hydroxyapatite coating in total hip arthroplasty. The poorer results in grit-blasted implants, as compared with the literature data, can be explained by alow degree of roughness of the Walter implant surfaces. CONCLUSIONS The combined Arbond hydroxyapatite coating improves conditions for implant osteointegration in the bone.
650    _2
$a dospělí $7 D000328
650    _2
$a senioři $7 D000368
650    _2
$a náhrada kyčelního kloubu $7 D019644
650    _2
$a biokompatibilní potahované materiály $7 D020099
650    _2
$a hydroxyapatit $7 D017886
650    _2
$a kyčelní protézy $7 D006622
650    _2
$a lidé $7 D006801
650    _2
$a Kaplanův-Meierův odhad $7 D053208
650    _2
$a lidé středního věku $7 D008875
650    _2
$a osteointegrace $7 D016348
650    _2
$a protézy - design $7 D011474
650    _2
$a selhání protézy $7 D011475
650    _2
$a reoperace $7 D012086
700    1_
$a Vavřík, Pavel, $d 1951- $7 nlk19990073980
700    1_
$a Jahoda, David $7 xx0014721
700    1_
$a Pokorný, David, $d 1968- $7 jn20001103568
700    1_
$a Ballay, Rastislav $7 xx0327926
700    1_
$a Sosna, Antonín, $d 1943- $7 jn20000710566
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae čechoslovaca $g Roč. 76, č. 3 (2009), s. 172-178 $x 0001-5415
910    __
$a ABA008 $b A 8 $c 507 $y 9 $z 0
990    __
$a 20091014142643 $b ABA008
991    __
$a 20250121143201 $b ABA008
999    __
$a ok $b bmc $g 689942 $s 551823
BAS    __
$a 3
BMC    __
$a 2009 $b 76 $c 3 $d 172-178 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x MED00011021
LZP    __
$a 2009-53/vtme

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace