The portable device for continual measurement of radon progenies on filter using the detector Timepix

B. Bulanek, J. Hulka, K. Jilek, I. Stekl,

. 2015 ; 164 (4) : 493-6. [pub] 20150519

Language English Country England, Great Britain

Document type Comparative Study, Journal Article, Research Support, Non-U.S. Gov't

In this article, a portable device was presented for continual measuring of equilibrium equivalent concentration (EEC) of (222)Rn based on the Timepix detector with 300-µm-thick active layer. In order to have a portable device, a filtration head was developed for collecting short-lived radon progenies attached on aerosols. The short-lived progenies are estimated from analysing alphas from decay of (218,214)Po from Millipore filter after termination of filtration. Comparison with beta measurement was done as well. The dependence of EEC on an air flow and filtration time was studied. The low-level detection limit for EEC was estimated from the last 10 min of 3-h decay measurement and was found in the range of 40-70 Bq m(-3). EEC was measured in National Radiation Protection Institute radon chamber, and results were compared with the commercial detector Fritra4.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc16020775
003      
CZ-PrNML
005      
20160722121411.0
007      
ta
008      
160722s2015 enk f 000 0|eng||
009      
AR
024    7_
$a 10.1093/rpd/ncv343 $2 doi
024    7_
$a 10.1093/rpd/ncv343 $2 doi
035    __
$a (PubMed)25990115
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a enk
100    1_
$a Bulanek, Boris $u National Radiation Protection Institute, Bartoskova 28, Prague 140 00, Czech Republic boris.bulanek@suro.cz.
245    14
$a The portable device for continual measurement of radon progenies on filter using the detector Timepix / $c B. Bulanek, J. Hulka, K. Jilek, I. Stekl,
520    9_
$a In this article, a portable device was presented for continual measuring of equilibrium equivalent concentration (EEC) of (222)Rn based on the Timepix detector with 300-µm-thick active layer. In order to have a portable device, a filtration head was developed for collecting short-lived radon progenies attached on aerosols. The short-lived progenies are estimated from analysing alphas from decay of (218,214)Po from Millipore filter after termination of filtration. Comparison with beta measurement was done as well. The dependence of EEC on an air flow and filtration time was studied. The low-level detection limit for EEC was estimated from the last 10 min of 3-h decay measurement and was found in the range of 40-70 Bq m(-3). EEC was measured in National Radiation Protection Institute radon chamber, and results were compared with the commercial detector Fritra4.
650    _2
$a aerosoly $7 D000336
650    _2
$a radioaktivní látky znečišťující vzduch $x analýza $7 D000396
650    _2
$a znečištění vzduchu ve vnitřním prostředí $x analýza $7 D016902
650    _2
$a alfa částice $7 D000512
650    _2
$a analogově digitální konverze $7 D000703
650    _2
$a beta částice $7 D001610
650    _2
$a design vybavení $7 D004867
650    _2
$a monitorování radiace $x přístrojové vybavení $x metody $7 D011834
650    _2
$a radon $x analýza $7 D011886
650    _2
$a dceřiné produkty radonu $x analýza $7 D017687
650    _2
$a počítačové zpracování signálu $7 D012815
650    _2
$a křemík $7 D012825
650    _2
$a ultrafiltrace $7 D014462
655    _2
$a srovnávací studie $7 D003160
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Hulka, Jiri $u National Radiation Protection Institute, Bartoskova 28, Prague 140 00, Czech Republic.
700    1_
$a Jilek, Karel $u National Radiation Protection Institute, Bartoskova 28, Prague 140 00, Czech Republic.
700    1_
$a Stekl, Ivan $u Institute of Experimental and Applied Physics, Czech Technical University in Prague, Horska 3a/22, Prague 128 00, Czech Republic.
773    0_
$w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 164, č. 4 (2015), s. 493-6
856    41
$u https://pubmed.ncbi.nlm.nih.gov/25990115 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20160722 $b ABA008
991    __
$a 20160722121625 $b ABA008
999    __
$a ok $b bmc $g 1155445 $s 945303
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2015 $b 164 $c 4 $d 493-6 $e 20150519 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029
LZP    __
$a Pubmed-20160722

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...