Analýza oxidačního poškození explantovaných komponent náhrady kyčelního kloubu Beznoska/Poldi
[Analysis of oxidative damage to components removed from Beznoska/Poldi total hip replacements]

P. Fulín, D. Pokorný, M. Šlouf, M. Nevoralová, T. Vacková, J. Dybal, N. Kaspříková, I. Landor

. 2016 ; 83 (3) : 155-162.

Jazyk čeština Země Česko

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc17010345

Grantová podpora
NT12229 MZ0 CEP - Centrální evidence projektů
NV15-31269A MZ0 CEP - Centrální evidence projektů

UNLABELLED: PURPOSE OF THE STUDY Ultrahigh molecular weight polyethylene (UHMWPE) is today the most frequently used bearing surface in total joint replacements (TJR) because of its properties, i.e., excellent biocompatibility, good mechanical and tribological performance and high wear resistance. UHMWPE liners are the most loaded TJR components and, therefore, their properties are decisive for TJR longevity. This study had three objectives: 1) to evaluate the oxidative degradation of explanted UHMWPE components; 2) to look for a statistically significant relationship between the extent of oxidative degradation and the durability of joint replacements; and 3) to investigate whether the durability of a TJR was related to the type of sterilisation used in manufacture. MATERIAL AND METHODS The study included 26 acetabular components obtained at revision arthroplasty between 2004 and 2013 from patients in whom a Beznoska/Poldi total hip replacement was used in the period from 1977 to 2002. The average age of the patients at the time of primary implantation was 57.9 years, the average longevity of the components removed was 18.63 years (range, 6.9 to 27.9 years). Samples of worn out and unworn areas from explanted components were processed in a three-step procedure in order to finally obtain 2-mm microtome sections. These were studied by infrared microspectroscopy. Oxidative damage to UHMWPE was determined as the oxidation index (OI); radiation damage to UHMWPE during sterilization was evaluated as the transvinylene index (VI); oxidation-induced changes in the polymer structure and its properties were characterised as the crystallinity index (CI); and local changes in mechanical properties due to oxidative degradation were assessed as microhardness (MH). Spearman's correlation coefficient and the Wilcoxon two-sample test were used for statistical analysis. RESULTS The OI values (average and maximum) in both worn out and unworn surface areas were related to component longevity. The difference between the oxidation index of ruptured components and that of the other components was statistically significant. Significant differences were also found between the average and maximum OI values of worn out areas and those of unworn surfaces. The relationship between the average oxidation index in both the unworn and worn out areas of UHMWPE components and the longevity of cracked components was statistically significant. DISCUSSION Our results show that the OI values obtained by microspectroscopy correlated with both the microscopic damage and the longevity of UHMWPE liners and the correlation was statistically significant also in relation to the longevity of total replacements. Relationships amongst OI, VI, CI and MH values as well as their relation to failure and longevity of total replacements are discussed. CONCLUSIONS It can be concluded that infrared microspectroscopic measurement of OI values is a simple and fast method to characterise UHMWPE liners. In addition, the IR spectra also show other supplementary characteristics, such as VI and CI indices. These values provide information on the quality of various UHMWPE types currently used in TJR surgery. The types of UHMWPE which exhibit high oxidative degradation should be avoided in clinical practice due to increased risk of early TJR failure. Responsible orthopedic surgeons should be aware of this fact and, if possible, collaborate with an independent, noncommercial laboratory in order to evaluate the quality of various UHMWPE liners used in their hospitals. KEY WORDS: UHMWPE, oxidation, total joint replacement, infrared spectroscopy, microhardness.

Analysis of oxidative damage to components removed from Beznoska/Poldi total hip replacements

000      
00000naa a2200000 a 4500
001      
bmc17010345
003      
CZ-PrNML
005      
20240117103428.0
007      
ta
008      
170320s2016 xr ad f 000 0|cze||
009      
AR
024    7_
$2 doi $a 10.55095/achot2016/023
035    __
$a (PubMed)27484072
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a cze $b eng
044    __
$a xr
100    1_
$a Fulín, Petr $u I. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy v Praze a Fakultní nemocnice Motol, Praha $7 xx0231618
245    10
$a Analýza oxidačního poškození explantovaných komponent náhrady kyčelního kloubu Beznoska/Poldi / $c P. Fulín, D. Pokorný, M. Šlouf, M. Nevoralová, T. Vacková, J. Dybal, N. Kaspříková, I. Landor
246    31
$a Analysis of oxidative damage to components removed from Beznoska/Poldi total hip replacements
520    9_
$a UNLABELLED: PURPOSE OF THE STUDY Ultrahigh molecular weight polyethylene (UHMWPE) is today the most frequently used bearing surface in total joint replacements (TJR) because of its properties, i.e., excellent biocompatibility, good mechanical and tribological performance and high wear resistance. UHMWPE liners are the most loaded TJR components and, therefore, their properties are decisive for TJR longevity. This study had three objectives: 1) to evaluate the oxidative degradation of explanted UHMWPE components; 2) to look for a statistically significant relationship between the extent of oxidative degradation and the durability of joint replacements; and 3) to investigate whether the durability of a TJR was related to the type of sterilisation used in manufacture. MATERIAL AND METHODS The study included 26 acetabular components obtained at revision arthroplasty between 2004 and 2013 from patients in whom a Beznoska/Poldi total hip replacement was used in the period from 1977 to 2002. The average age of the patients at the time of primary implantation was 57.9 years, the average longevity of the components removed was 18.63 years (range, 6.9 to 27.9 years). Samples of worn out and unworn areas from explanted components were processed in a three-step procedure in order to finally obtain 2-mm microtome sections. These were studied by infrared microspectroscopy. Oxidative damage to UHMWPE was determined as the oxidation index (OI); radiation damage to UHMWPE during sterilization was evaluated as the transvinylene index (VI); oxidation-induced changes in the polymer structure and its properties were characterised as the crystallinity index (CI); and local changes in mechanical properties due to oxidative degradation were assessed as microhardness (MH). Spearman's correlation coefficient and the Wilcoxon two-sample test were used for statistical analysis. RESULTS The OI values (average and maximum) in both worn out and unworn surface areas were related to component longevity. The difference between the oxidation index of ruptured components and that of the other components was statistically significant. Significant differences were also found between the average and maximum OI values of worn out areas and those of unworn surfaces. The relationship between the average oxidation index in both the unworn and worn out areas of UHMWPE components and the longevity of cracked components was statistically significant. DISCUSSION Our results show that the OI values obtained by microspectroscopy correlated with both the microscopic damage and the longevity of UHMWPE liners and the correlation was statistically significant also in relation to the longevity of total replacements. Relationships amongst OI, VI, CI and MH values as well as their relation to failure and longevity of total replacements are discussed. CONCLUSIONS It can be concluded that infrared microspectroscopic measurement of OI values is a simple and fast method to characterise UHMWPE liners. In addition, the IR spectra also show other supplementary characteristics, such as VI and CI indices. These values provide information on the quality of various UHMWPE types currently used in TJR surgery. The types of UHMWPE which exhibit high oxidative degradation should be avoided in clinical practice due to increased risk of early TJR failure. Responsible orthopedic surgeons should be aware of this fact and, if possible, collaborate with an independent, noncommercial laboratory in order to evaluate the quality of various UHMWPE liners used in their hospitals. KEY WORDS: UHMWPE, oxidation, total joint replacement, infrared spectroscopy, microhardness.
650    _2
$a senioři $7 D000368
650    _2
$a náhrada kyčelního kloubu $x přístrojové vybavení $7 D019644
650    _2
$a biokompatibilní materiály $x chemie $7 D001672
650    _2
$a lidé $7 D006801
650    _2
$a testování materiálů $x metody $7 D008422
650    _2
$a lidé středního věku $7 D008875
650    _2
$a oxidační stres $x fyziologie $7 D018384
650    _2
$a polyethyleny $x chemie $7 D011095
650    _2
$a selhání protézy $7 D011475
650    _2
$a reoperace $7 D012086
650    _2
$a spektrofotometrie infračervená $7 D013055
655    _2
$a časopisecké články $7 D016428
700    1_
$a Pokorný, David, $d 1968- $7 jn20001103568 $u I. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy v Praze a Fakultní nemocnice Motol, Praha
700    1_
$a Šlouf, Miroslav, $d 1973- $7 xx0107048 $u Ústav makromolekulami chemie Akademie věd ČR, Praha
700    1_
$a Nevoralová, Martina $7 ntka173525 $u Ústav makromolekulami chemie Akademie věd ČR, Praha
700    1_
$a Vacková, Taťana $7 _AN076109 $u Ústav makromolekulami chemie Akademie věd ČR, Praha
700    1_
$a Dybal, Jiří $7 xx0312810 $u Ústav makromolekulami chemie Akademie věd ČR, Praha
700    1_
$a Kaspříková, Nikola $7 jx20120206012 $u Ústav biofyziky a informatiky 1. lékařské fakulty Univerzity Karlovy v Praze
700    1_
$a Landor, Ivan, $d 1956- $7 xx0062301 $u I. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy v Praze a Fakultní nemocnice Motol, Praha
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $x 0001-5415 $g Roč. 83, č. 3 (2016), s. 155-162
910    __
$a ABA008 $b A 8 $c 507 $y 4 $z 0
990    __
$a 20170320 $b ABA008
991    __
$a 20240117103424 $b ABA008
999    __
$a ok $b bmc $g 1197264 $s 971061
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2016 $b 83 $c 3 $d 155-162 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
GRA    __
$a NT12229 $p MZ0
GRA    __
$a NV15-31269A $p MZ0
LZP    __
$b NLK118 $a Pubmed-20170320

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...