New approach for cystic fibrosis diagnosis based on chloride/potassium ratio analyzed in non-invasively obtained skin-wipe sweat samples by capillary electrophoresis with contactless conductometric detection
P. Ďurč, F. Foret, E. Pokojová, L. Homola, J. Skřičková, V. Herout, M. Dastych, H. Vinohradská, P. Kubáň,
Jazyk angličtina Země Německo
Typ dokumentu časopisecké články, validační studie
NLK
ProQuest Central
od 2013-01-01 do Před 1 rokem
Medline Complete (EBSCOhost)
od 2003-01-01 do Před 1 rokem
Health & Medicine (ProQuest)
od 2013-01-01 do Před 1 rokem
- MeSH
- analýza hlavních komponent MeSH
- chloridy analýza MeSH
- cystická fibróza diagnóza MeSH
- dítě MeSH
- dospělí MeSH
- draslík analýza MeSH
- elektroforéza kapilární metody MeSH
- kojenec MeSH
- lidé středního věku MeSH
- lidé MeSH
- mladiství MeSH
- odběr biologického vzorku metody MeSH
- pot chemie MeSH
- předškolní dítě MeSH
- senzitivita a specificita MeSH
- Check Tag
- dítě MeSH
- dospělí MeSH
- kojenec MeSH
- lidé středního věku MeSH
- lidé MeSH
- mladiství MeSH
- předškolní dítě MeSH
- Publikační typ
- časopisecké články MeSH
- validační studie MeSH
A new approach for sweat analysis used in cystic fibrosis (CF) diagnosis is proposed. It consists of a noninvasive skin-wipe sampling followed by analysis of target ions using capillary electrophoresis with contactless conductometric detection (C4D). The skin-wipe sampling consists of wiping a defined skin area with precleaned cotton swab moistened with 100 μL deionized water. The skin-wipe sample is then extracted for 3 min into 400 μL deionized water, and the extract is analyzed directly. The developed sampling method is cheap, simple, fast, and painless, and can replace the conventional pilocarpine-induced sweat chloride test commonly applied in CF diagnosis. The aqueous extract of the skin-wipe sample content is analyzed simultaneously by capillary electrophoresis with contactless conductometric detection using a double opposite end injection. A 20 mmol/L L-histidine/2-(N-morpholino)ethanesulfonic acid and 2 mmol/L 18-crown-6 at pH 6 electrolyte can separate all the major ions in less than 7 min. Skin-wipe sample extracts from 30 study participants-ten adult patients with CF (25-50 years old), ten pediatric patients with CF (1-15 years old), and ten healthy control individuals (1-18 years old)-were obtained and analyzed. From the analyzed ions in all samples, a significant difference between chloride and potassium concentrations was found in the CF patients and healthy controls. We propose the use of the Cl-/K+ratio rather than the absolute Cl-concentration and a cutoff value of 4 in skin-wipe sample extracts as an alternative to the conventional sweat chloride analysis. The proposed Cl-/K+ion ratio proved to be a more reliable indicator, is independent of the patient's age, and allows better differentiation between non-CF individuals and CF patients having intermediate values on the Cl-sweat test. Figure New approach for cystic fibrosis diagnosis based on skin-wipe sampling of forearm and analysis of ionic content (Cl-/K+ratio) in skin-wipe extracts by capillary electrophoresis with contactless conductometric detection.
Department of Clinical Biochemistry University Hospital Brno Jihlavská 20 625 00 Brno Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc18010713
- 003
- CZ-PrNML
- 005
- 20180528163616.0
- 007
- ta
- 008
- 180404s2017 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1007/s00216-017-0318-6 $2 doi
- 035 __
- $a (PubMed)28357484
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Ďurč, Pavol $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00, Brno, Czech Republic. Department of Chemistry, Masaryk University, Kotlářská 2, 611 37, Brno, Czech Republic.
- 245 10
- $a New approach for cystic fibrosis diagnosis based on chloride/potassium ratio analyzed in non-invasively obtained skin-wipe sweat samples by capillary electrophoresis with contactless conductometric detection / $c P. Ďurč, F. Foret, E. Pokojová, L. Homola, J. Skřičková, V. Herout, M. Dastych, H. Vinohradská, P. Kubáň,
- 520 9_
- $a A new approach for sweat analysis used in cystic fibrosis (CF) diagnosis is proposed. It consists of a noninvasive skin-wipe sampling followed by analysis of target ions using capillary electrophoresis with contactless conductometric detection (C4D). The skin-wipe sampling consists of wiping a defined skin area with precleaned cotton swab moistened with 100 μL deionized water. The skin-wipe sample is then extracted for 3 min into 400 μL deionized water, and the extract is analyzed directly. The developed sampling method is cheap, simple, fast, and painless, and can replace the conventional pilocarpine-induced sweat chloride test commonly applied in CF diagnosis. The aqueous extract of the skin-wipe sample content is analyzed simultaneously by capillary electrophoresis with contactless conductometric detection using a double opposite end injection. A 20 mmol/L L-histidine/2-(N-morpholino)ethanesulfonic acid and 2 mmol/L 18-crown-6 at pH 6 electrolyte can separate all the major ions in less than 7 min. Skin-wipe sample extracts from 30 study participants-ten adult patients with CF (25-50 years old), ten pediatric patients with CF (1-15 years old), and ten healthy control individuals (1-18 years old)-were obtained and analyzed. From the analyzed ions in all samples, a significant difference between chloride and potassium concentrations was found in the CF patients and healthy controls. We propose the use of the Cl-/K+ratio rather than the absolute Cl-concentration and a cutoff value of 4 in skin-wipe sample extracts as an alternative to the conventional sweat chloride analysis. The proposed Cl-/K+ion ratio proved to be a more reliable indicator, is independent of the patient's age, and allows better differentiation between non-CF individuals and CF patients having intermediate values on the Cl-sweat test. Figure New approach for cystic fibrosis diagnosis based on skin-wipe sampling of forearm and analysis of ionic content (Cl-/K+ratio) in skin-wipe extracts by capillary electrophoresis with contactless conductometric detection.
- 650 _2
- $a mladiství $7 D000293
- 650 _2
- $a dospělí $7 D000328
- 650 _2
- $a dítě $7 D002648
- 650 _2
- $a předškolní dítě $7 D002675
- 650 _2
- $a chloridy $x analýza $7 D002712
- 650 _2
- $a cystická fibróza $x diagnóza $7 D003550
- 650 _2
- $a elektroforéza kapilární $x metody $7 D019075
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a kojenec $7 D007223
- 650 _2
- $a lidé středního věku $7 D008875
- 650 _2
- $a draslík $x analýza $7 D011188
- 650 _2
- $a analýza hlavních komponent $7 D025341
- 650 _2
- $a senzitivita a specificita $7 D012680
- 650 _2
- $a odběr biologického vzorku $x metody $7 D013048
- 650 _2
- $a pot $x chemie $7 D013542
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a validační studie $7 D023361
- 700 1_
- $a Foret, František $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00, Brno, Czech Republic.
- 700 1_
- $a Pokojová, Eva $u Department of Respiratory Diseases and TB, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic.
- 700 1_
- $a Homola, Lukáš $u Cystic Fibrosis Centre and Clinic of Pediatric Infectious Diseases, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic.
- 700 1_
- $a Skřičková, Jana $u Department of Respiratory Diseases and TB, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic.
- 700 1_
- $a Herout, Vladimír $u Department of Respiratory Diseases and TB, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic.
- 700 1_
- $a Dastych, Milan $u Department of Clinical Biochemistry, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic.
- 700 1_
- $a Vinohradská, Hana, $d 1960- $u Department of Clinical Biochemistry, University Hospital Brno, Jihlavská 20, 625 00, Brno, Czech Republic. $7 xx0224599
- 700 1_
- $a Kubáň, Petr $u Department of Bioanalytical Instrumentation, CEITEC Masaryk University, Veveří 97, 602 00, Brno, Czech Republic. petr.kuban@gmail.com.
- 773 0_
- $w MED00006638 $t Analytical and bioanalytical chemistry $x 1618-2650 $g Roč. 409, č. 14 (2017), s. 3507-3514
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/28357484 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20180404 $b ABA008
- 991 __
- $a 20180528163808 $b ABA008
- 999 __
- $a ok $b bmc $g 1288198 $s 1007525
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2017 $b 409 $c 14 $d 3507-3514 $e 20170329 $i 1618-2650 $m Analytical and bioanalytical chemistry $n Anal Bioanal Chem $x MED00006638
- LZP __
- $a Pubmed-20180404