A Non-Invasive Multichannel Hybrid Fiber-Optic Sensor System for Vital Sign Monitoring

M. Fajkus, J. Nedoma, R. Martinek, V. Vasinek, H. Nazeran, P. Siska,

. 2017 ; 17 (1) : . [pub] 20170108

Language English Country Switzerland

Document type Journal Article

In this article, we briefly describe the design, construction, and functional verification of a hybrid multichannel fiber-optic sensor system for basic vital sign monitoring. This sensor uses a novel non-invasive measurement probe based on the fiber Bragg grating (FBG). The probe is composed of two FBGs encapsulated inside a polydimethylsiloxane polymer (PDMS). The PDMS is non-reactive to human skin and resistant to electromagnetic waves, UV absorption, and radiation. We emphasize the construction of the probe to be specifically used for basic vital sign monitoring such as body temperature, respiratory rate and heart rate. The proposed sensor system can continuously process incoming signals from up to 128 individuals. We first present the overall design of this novel multichannel sensor and then elaborate on how it has the potential to simplify vital sign monitoring and consequently improve the comfort level of patients in long-term health care facilities, hospitals and clinics. The reference ECG signal was acquired with the use of standard gel electrodes fixed to the monitored person's chest using a real-time monitoring system for ECG signals with virtual instrumentation. The outcomes of these experiments have unambiguously proved the functionality of the sensor system and will be used to inform our future research in this fast developing and emerging field.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc18025198
003      
CZ-PrNML
005      
20180710095245.0
007      
ta
008      
180709s2017 sz f 000 0|eng||
009      
AR
024    7_
$a 10.3390/s17010111 $2 doi
035    __
$a (PubMed)28075341
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a sz
100    1_
$a Fajkus, Marcel $u Department of Telecommunications, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, 17 Listopadu 15, Ostrava 70833, Czech Republic. marcel.fajkus@vsb.cz.
245    12
$a A Non-Invasive Multichannel Hybrid Fiber-Optic Sensor System for Vital Sign Monitoring / $c M. Fajkus, J. Nedoma, R. Martinek, V. Vasinek, H. Nazeran, P. Siska,
520    9_
$a In this article, we briefly describe the design, construction, and functional verification of a hybrid multichannel fiber-optic sensor system for basic vital sign monitoring. This sensor uses a novel non-invasive measurement probe based on the fiber Bragg grating (FBG). The probe is composed of two FBGs encapsulated inside a polydimethylsiloxane polymer (PDMS). The PDMS is non-reactive to human skin and resistant to electromagnetic waves, UV absorption, and radiation. We emphasize the construction of the probe to be specifically used for basic vital sign monitoring such as body temperature, respiratory rate and heart rate. The proposed sensor system can continuously process incoming signals from up to 128 individuals. We first present the overall design of this novel multichannel sensor and then elaborate on how it has the potential to simplify vital sign monitoring and consequently improve the comfort level of patients in long-term health care facilities, hospitals and clinics. The reference ECG signal was acquired with the use of standard gel electrodes fixed to the monitored person's chest using a real-time monitoring system for ECG signals with virtual instrumentation. The outcomes of these experiments have unambiguously proved the functionality of the sensor system and will be used to inform our future research in this fast developing and emerging field.
650    12
$a technologie optických vláken $7 D005336
650    _2
$a srdeční frekvence $7 D006339
650    _2
$a lidé $7 D006801
650    _2
$a monitorování fyziologických funkcí $7 D008991
650    _2
$a optická vlákna $7 D055100
650    _2
$a dechová frekvence $7 D056152
655    _2
$a časopisecké články $7 D016428
700    1_
$a Nedoma, Jan $u Department of Telecommunications, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, 17 Listopadu 15, Ostrava 70833, Czech Republic. jan.nedoma@vsb.cz.
700    1_
$a Martinek, Radek $u Department of Cybernetics and Biomedical Engineering, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, 17 Listopadu 15, Ostrava 70833, Czech Republic. radek.martinek@vsb.cz.
700    1_
$a Vasinek, Vladimir $u Department of Telecommunications, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, 17 Listopadu 15, Ostrava 70833, Czech Republic. vladimir.vasinek@vsb.cz.
700    1_
$a Nazeran, Homer $u Department of Electrical and Computer Engineering, University of Texas El Paso, 500W University Ave, El Paso, TX 79968, USA. hnazeran@utep.edu.
700    1_
$a Siska, Petr $u Department of Telecommunications, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava, 17 Listopadu 15, Ostrava 70833, Czech Republic. petr.siska@vsb.cz.
773    0_
$w MED00008309 $t Sensors (Basel, Switzerland) $x 1424-8220 $g Roč. 17, č. 1 (2017)
856    41
$u https://pubmed.ncbi.nlm.nih.gov/28075341 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20180709 $b ABA008
991    __
$a 20180710095535 $b ABA008
999    __
$a ok $b bmc $g 1317329 $s 1022119
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2017 $b 17 $c 1 $e 20170108 $i 1424-8220 $m Sensors $n Sensors Basel $x MED00008309
LZP    __
$a Pubmed-20180709

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...