• Je něco špatně v tomto záznamu ?

Using single-isomer octa(6-O-sulfo)-γ-cyclodextrin for fast capillary zone electrophoretic enantioseparation of pindolol: Determination of complexation constants, software-assisted optimization, and method validation

L. Kanizsová, M. Ansorge, I. Zusková, P. Dubský,

. 2018 ; 1568 (-) : 214-221. [pub] 20180705

Jazyk angličtina Země Nizozemsko

Typ dokumentu časopisecké články, validační studie

Perzistentní odkaz   https://www.medvik.cz/link/bmc18032986

The present study describes a rapid and effective capillary electrophoresis (CE) method for the enantioseparation of pindolol using single-isomer octa(6-O-sulfo)-γ-cyclodextrin. The complexation parameters were determined under neutral and high pH conditions to identify optimal separation conditions using a theoretical model. Baseline separation of pindolol enantiomers was achieved within 6 min in a sodium/MOPS buffer, pH 7.2, with a selector concentration of 6 mM. The method was validated according to the ICH guidelines using imidazole as an internal standard. Low limits of detection and quantification were found, specifically 1.2 μg/mL and 4 μg/mL (0.6 μg/mL and 2 μg/mL per enantiomer), respectively. The calibration curves showed good linearity, with a coefficient of determination R2 ≥ 0.999 over a 5 - 55 μg/mL concentration range and over a 50 - 300 μg/mL concentration range of the racemic mixture. The relative standard deviations (%RSD) of intra-day and inter-day precision were lower than 8% at LOQ level, lower than 3% at 50 μg/mL level and lower than 1.5% at 300 μg/mL level. Accuracy ranged from 95 to 103% (106% at LOQ level). The proposed method was successfully tested on a medical formulation of Visken® Sandoz intravenous solution and Visken® Teofarma pills for oral use.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc18032986
003      
CZ-PrNML
005      
20181008122204.0
007      
ta
008      
181008s2018 ne f 000 0|eng||
009      
AR
024    7_
$a 10.1016/j.chroma.2018.07.020 $2 doi
035    __
$a (PubMed)30122164
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a ne
100    1_
$a Kanizsová, Lívia $u Charles University in Prague, Faculty of Science, Department of Physical and Macromolecular Chemistry, Prague, Czech Republic.
245    10
$a Using single-isomer octa(6-O-sulfo)-γ-cyclodextrin for fast capillary zone electrophoretic enantioseparation of pindolol: Determination of complexation constants, software-assisted optimization, and method validation / $c L. Kanizsová, M. Ansorge, I. Zusková, P. Dubský,
520    9_
$a The present study describes a rapid and effective capillary electrophoresis (CE) method for the enantioseparation of pindolol using single-isomer octa(6-O-sulfo)-γ-cyclodextrin. The complexation parameters were determined under neutral and high pH conditions to identify optimal separation conditions using a theoretical model. Baseline separation of pindolol enantiomers was achieved within 6 min in a sodium/MOPS buffer, pH 7.2, with a selector concentration of 6 mM. The method was validated according to the ICH guidelines using imidazole as an internal standard. Low limits of detection and quantification were found, specifically 1.2 μg/mL and 4 μg/mL (0.6 μg/mL and 2 μg/mL per enantiomer), respectively. The calibration curves showed good linearity, with a coefficient of determination R2 ≥ 0.999 over a 5 - 55 μg/mL concentration range and over a 50 - 300 μg/mL concentration range of the racemic mixture. The relative standard deviations (%RSD) of intra-day and inter-day precision were lower than 8% at LOQ level, lower than 3% at 50 μg/mL level and lower than 1.5% at 300 μg/mL level. Accuracy ranged from 95 to 103% (106% at LOQ level). The proposed method was successfully tested on a medical formulation of Visken® Sandoz intravenous solution and Visken® Teofarma pills for oral use.
650    _2
$a pufry $7 D002021
650    _2
$a kalibrace $7 D002138
650    _2
$a elektroforéza kapilární $x metody $7 D019075
650    _2
$a koncentrace vodíkových iontů $7 D006863
650    _2
$a limita detekce $7 D057230
650    _2
$a pindolol $x izolace a purifikace $7 D010869
650    _2
$a reprodukovatelnost výsledků $7 D015203
650    12
$a software $7 D012984
650    _2
$a stereoizomerie $7 D013237
650    _2
$a časové faktory $7 D013997
650    _2
$a gama-cyklodextriny $x chemie $7 D047408
655    _2
$a časopisecké články $7 D016428
655    _2
$a validační studie $7 D023361
700    1_
$a Ansorge, Martin $u Charles University in Prague, Faculty of Science, Department of Physical and Macromolecular Chemistry, Prague, Czech Republic.
700    1_
$a Zusková, Iva $u Charles University in Prague, Faculty of Science, Department of Physical and Macromolecular Chemistry, Prague, Czech Republic.
700    1_
$a Dubský, Pavel $u Charles University in Prague, Faculty of Science, Department of Physical and Macromolecular Chemistry, Prague, Czech Republic. Electronic address: pavel.dubsky@natur.cuni.cz.
773    0_
$w MED00004962 $t Journal of chromatography. A $x 1873-3778 $g Roč. 1568, č. - (2018), s. 214-221
856    41
$u https://pubmed.ncbi.nlm.nih.gov/30122164 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y a $z 0
990    __
$a 20181008 $b ABA008
991    __
$a 20181008122650 $b ABA008
999    __
$a ok $b bmc $g 1339334 $s 1029980
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2018 $b 1568 $c - $d 214-221 $e 20180705 $i 1873-3778 $m Journal of chromatography. A, Including electrophoresis and other separation methods $n J Chromatogr A $x MED00004962
LZP    __
$a Pubmed-20181008

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace