COMPARISON OF DIFFERENT TECHNIQUES FOR EVALUATION OF DOSE DISTRIBUTIONS IN RADIOTHERAPY USING RADIOCHROMIC EBT3 FILMS
V. Dufek, I. Horakova, I. Koniarova,
Jazyk angličtina Země Velká Británie
Typ dokumentu srovnávací studie, hodnotící studie, časopisecké články
PubMed
31711200
DOI
10.1093/rpd/ncz231
Knihovny.cz E-zdroje
- MeSH
- algoritmy MeSH
- celková dávka radioterapie MeSH
- fantomy radiodiagnostické * MeSH
- filmová dozimetrie přístrojové vybavení metody MeSH
- kalibrace MeSH
- lidé MeSH
- nádory prostaty radioterapie MeSH
- plánování radioterapie pomocí počítače metody MeSH
- radioterapie s modulovanou intenzitou metody MeSH
- záření gama MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- hodnotící studie MeSH
- srovnávací studie MeSH
In radiotherapy, radiochromic films can be used for verification of delivery of dose distributions calculated by treatment planning systems. The main objective of this work was to compare three different techniques for evaluation of dose distributions for prostate cancer treatment plans using radiochromic EBT3 films. These techniques are: red channel evaluation taking into account only a response of irradiated film (R), red channel evaluation taking into account a response of unirradiated and irradiated film (Rcor) and multichannel evaluation in FilmQA software (RGB). Also comparison between film and MatriXX measurement was performed. Comparison showed that gamma analysis passing rates strongly depend on evaluation technique and on a model of scanner for digitizing films. The highest gamma passing rates were obtained with red channel evaluation taking into account a response of unirradiated and irradiated film using Epson V750 scanner (Rcor) and multichannel evaluation in FilmQA using Epson 11000XL scanner.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc20025371
- 003
- CZ-PrNML
- 005
- 20201222153851.0
- 007
- ta
- 008
- 201125s2019 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1093/rpd/ncz231 $2 doi
- 035 __
- $a (PubMed)31711200
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Dufek, V $u Medical Exposure Section, National Radiation Protection Institute, Prague, Czech Republic. Department of Radiological Physics, Na Bulovce Hospital, Prague, Czech Republic.
- 245 10
- $a COMPARISON OF DIFFERENT TECHNIQUES FOR EVALUATION OF DOSE DISTRIBUTIONS IN RADIOTHERAPY USING RADIOCHROMIC EBT3 FILMS / $c V. Dufek, I. Horakova, I. Koniarova,
- 520 9_
- $a In radiotherapy, radiochromic films can be used for verification of delivery of dose distributions calculated by treatment planning systems. The main objective of this work was to compare three different techniques for evaluation of dose distributions for prostate cancer treatment plans using radiochromic EBT3 films. These techniques are: red channel evaluation taking into account only a response of irradiated film (R), red channel evaluation taking into account a response of unirradiated and irradiated film (Rcor) and multichannel evaluation in FilmQA software (RGB). Also comparison between film and MatriXX measurement was performed. Comparison showed that gamma analysis passing rates strongly depend on evaluation technique and on a model of scanner for digitizing films. The highest gamma passing rates were obtained with red channel evaluation taking into account a response of unirradiated and irradiated film using Epson V750 scanner (Rcor) and multichannel evaluation in FilmQA using Epson 11000XL scanner.
- 650 _2
- $a algoritmy $7 D000465
- 650 _2
- $a kalibrace $7 D002138
- 650 _2
- $a filmová dozimetrie $x přístrojové vybavení $x metody $7 D005373
- 650 _2
- $a záření gama $7 D005720
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 12
- $a fantomy radiodiagnostické $7 D019047
- 650 _2
- $a nádory prostaty $x radioterapie $7 D011471
- 650 _2
- $a celková dávka radioterapie $7 D011879
- 650 _2
- $a plánování radioterapie pomocí počítače $x metody $7 D011880
- 650 _2
- $a radioterapie s modulovanou intenzitou $x metody $7 D050397
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a hodnotící studie $7 D023362
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Horakova, I $u Medical Exposure Section, National Radiation Protection Institute, Prague, Czech Republic.
- 700 1_
- $a Koniarova, I $u Medical Exposure Section, National Radiation Protection Institute, Prague, Czech Republic.
- 773 0_
- $w MED00004029 $t Radiation protection dosimetry $x 1742-3406 $g Roč. 186, č. 2-3 (2019), s. 357-361
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/31711200 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20201125 $b ABA008
- 991 __
- $a 20201222153847 $b ABA008
- 999 __
- $a ok $b bmc $g 1599516 $s 1116057
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2019 $b 186 $c 2-3 $d 357-361 $e 2019Dec31 $i 1742-3406 $m Radiation protection dosimetry $n Radiat Prot Dosimetry $x MED00004029
- LZP __
- $a Pubmed-20201125