Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models [Srovnání modifikovaných transiliakálních vnitřních fixátorů: mechanické testování na modelech pánve]
M. Salášek, L. Lobovský, J. Hluchá, J. Krystek, J. Křen
Jazyk angličtina Země Česko
Typ dokumentu časopisecké články
- MeSH
- biomechanika MeSH
- fraktury kostí * MeSH
- interní fixátory MeSH
- kostní šrouby MeSH
- křížová kost chirurgie MeSH
- lidé MeSH
- mechanické testy MeSH
- pánev MeSH
- pánevní kosti * diagnostické zobrazování chirurgie MeSH
- vnitřní fixace fraktury MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
PURPOSE OF THE STUDY Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). MATERIAL AND METHODS Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to assess the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. A dedicated load cell and a monoaxial extensometer were utilised. Every measurement was repeated at least 10 times. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. RESULTS Mean stiffness ratio (±1SD) of pelvic structure was 0.638 ± 0.005 for TIFI-C, 0.722 ± 0.014 for TIFI-A and 0.720 ± 0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p < 0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112). CONCLUSIONS Results of the mechanical analysis using pelvic models indicate that for linear vertical transforaminal sacral fracture without comminuted zone, an application of either TIFI-A or DTIFI provides significantly higher stiffness of the lateral pelvic segment than application of TIFI-C. Key words: transforaminal sacral fracture, transiliac internal fixator, dual TIFI, stability, biomechanics, digital image correlation.
Srovnání modifikovaných transiliakálních vnitřních fixátorů: mechanické testování na modelech pánve
Literatura
- 000
- 00000naa a2200000 a 4500
- 001
- bmc21029375
- 003
- CZ-PrNML
- 005
- 20220228121627.0
- 007
- ta
- 008
- 211206s2021 xr ad f 000 0|eng||
- 009
- AR
- 035 __
- $a (PubMed)34738891
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Salášek, Martin $7 xx0129311 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
- 245 10
- $a Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models / $c M. Salášek, L. Lobovský, J. Hluchá, J. Krystek, J. Křen
- 246 31
- $a Srovnání modifikovaných transiliakálních vnitřních fixátorů: mechanické testování na modelech pánve
- 504 __
- $a Literatura
- 520 9_
- $a PURPOSE OF THE STUDY Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). MATERIAL AND METHODS Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to assess the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. A dedicated load cell and a monoaxial extensometer were utilised. Every measurement was repeated at least 10 times. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. RESULTS Mean stiffness ratio (±1SD) of pelvic structure was 0.638 ± 0.005 for TIFI-C, 0.722 ± 0.014 for TIFI-A and 0.720 ± 0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p < 0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112). CONCLUSIONS Results of the mechanical analysis using pelvic models indicate that for linear vertical transforaminal sacral fracture without comminuted zone, an application of either TIFI-A or DTIFI provides significantly higher stiffness of the lateral pelvic segment than application of TIFI-C. Key words: transforaminal sacral fracture, transiliac internal fixator, dual TIFI, stability, biomechanics, digital image correlation.
- 650 _2
- $a biomechanika $7 D001696
- 650 _2
- $a kostní šrouby $7 D001863
- 650 _2
- $a vnitřní fixace fraktury $7 D005593
- 650 12
- $a fraktury kostí $7 D050723
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a interní fixátory $7 D016268
- 650 _2
- $a mechanické testy $7 D000083522
- 650 12
- $a pánevní kosti $x diagnostické zobrazování $x chirurgie $7 D010384
- 650 _2
- $a pánev $7 D010388
- 650 _2
- $a křížová kost $x chirurgie $7 D012447
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Lobovský, Libor $7 mzk2016922576 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
- 700 1_
- $a Hluchá, Jana $7 xx0267925 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
- 700 1_
- $a Krystek, Jan, $d 1984- $7 xx0191919 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
- 700 1_
- $a Křen, Jiří, $d 1947- $7 ola2002146730 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
- 773 0_
- $w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca $x 0001-5415 $g Roč. 88, č. 5 (2021), s. 333-338
- 856 41
- $u https://achot.cz/pdfs/ach/2021/05/01.pdf $y plný text volně přístupný
- 910 __
- $a ABA008 $b A 8 $c 507 $y p $z 0
- 990 __
- $a 20211206 $b ABA008
- 991 __
- $a 20220228121625 $b ABA008
- 999 __
- $a ok $b bmc $g 1742357 $s 1149928
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 88 $c 5 $d 333-338 $e 20211020 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
- LZP __
- $b NLK118 $a Pubmed-20211206