PURPOSE OF THE STUDY Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). MATERIAL AND METHODS Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to assess the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. A dedicated load cell and a monoaxial extensometer were utilised. Every measurement was repeated at least 10 times. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. RESULTS Mean stiffness ratio (±1SD) of pelvic structure was 0.638 ± 0.005 for TIFI-C, 0.722 ± 0.014 for TIFI-A and 0.720 ± 0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p < 0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112). CONCLUSIONS Results of the mechanical analysis using pelvic models indicate that for linear vertical transforaminal sacral fracture without comminuted zone, an application of either TIFI-A or DTIFI provides significantly higher stiffness of the lateral pelvic segment than application of TIFI-C. Key words: transforaminal sacral fracture, transiliac internal fixator, dual TIFI, stability, biomechanics, digital image correlation.

Srovnání modifikovaných transiliakálních vnitřních fixátorů: mechanické testování na modelech pánve

Bibliografie atd.

Literatura

000      
00000naa a2200000 a 4500
001      
bmc21029375
003      
CZ-PrNML
005      
20220228121627.0
007      
ta
008      
211206s2021 xr ad f 000 0|eng||
009      
AR
035    __
$a (PubMed)34738891
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xr
100    1_
$a Salášek, Martin $7 xx0129311 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
245    10
$a Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models / $c M. Salášek, L. Lobovský, J. Hluchá, J. Krystek, J. Křen
246    31
$a Srovnání modifikovaných transiliakálních vnitřních fixátorů: mechanické testování na modelech pánve
504    __
$a Literatura
520    9_
$a PURPOSE OF THE STUDY Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). MATERIAL AND METHODS Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to assess the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. A dedicated load cell and a monoaxial extensometer were utilised. Every measurement was repeated at least 10 times. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. RESULTS Mean stiffness ratio (±1SD) of pelvic structure was 0.638 ± 0.005 for TIFI-C, 0.722 ± 0.014 for TIFI-A and 0.720 ± 0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p < 0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112). CONCLUSIONS Results of the mechanical analysis using pelvic models indicate that for linear vertical transforaminal sacral fracture without comminuted zone, an application of either TIFI-A or DTIFI provides significantly higher stiffness of the lateral pelvic segment than application of TIFI-C. Key words: transforaminal sacral fracture, transiliac internal fixator, dual TIFI, stability, biomechanics, digital image correlation.
650    _2
$a biomechanika $7 D001696
650    _2
$a kostní šrouby $7 D001863
650    _2
$a vnitřní fixace fraktury $7 D005593
650    12
$a fraktury kostí $7 D050723
650    _2
$a lidé $7 D006801
650    _2
$a interní fixátory $7 D016268
650    _2
$a mechanické testy $7 D000083522
650    12
$a pánevní kosti $x diagnostické zobrazování $x chirurgie $7 D010384
650    _2
$a pánev $7 D010388
650    _2
$a křížová kost $x chirurgie $7 D012447
655    _2
$a časopisecké články $7 D016428
700    1_
$a Lobovský, Libor $7 mzk2016922576 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
700    1_
$a Hluchá, Jana $7 xx0267925 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
700    1_
$a Krystek, Jan, $d 1984- $7 xx0191919 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
700    1_
$a Křen, Jiří, $d 1947- $7 ola2002146730 $u NTIS - New Technologies for the Information Society, Faculty of Applied Sciences, University of West Bohemia, Plzeň, Czech Republic
773    0_
$w MED00011021 $t Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca $x 0001-5415 $g Roč. 88, č. 5 (2021), s. 333-338
856    41
$u https://achot.cz/pdfs/ach/2021/05/01.pdf $y plný text volně přístupný
910    __
$a ABA008 $b A 8 $c 507 $y p $z 0
990    __
$a 20211206 $b ABA008
991    __
$a 20220228121625 $b ABA008
999    __
$a ok $b bmc $g 1742357 $s 1149928
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2021 $b 88 $c 5 $d 333-338 $e 20211020 $i 0001-5415 $m Acta chirurgiae orthopaedicae et traumatologiae Čechoslovaca $n Acta chir. orthop. traumatol. Čechoslovaca $x MED00011021
LZP    __
$b NLK118 $a Pubmed-20211206

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace