Je něco špatně v tomto záznamu ?
Application of capillary electrophoresis-nano-electrospray ionization-mass spectrometry for the determination of N-nitrosodimethylamine in pharmaceuticals
R. Řemínek, F. Foret, DS. Chung
Jazyk angličtina Země Německo
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
33368407
DOI
10.1002/elps.202000303
Knihovny.cz E-zdroje
- MeSH
- dimethylnitrosamin analýza MeSH
- elektroforéza kapilární metody MeSH
- hmotnostní spektrometrie s elektrosprejovou ionizací metody MeSH
- kontaminace léku * MeSH
- lineární modely MeSH
- nanotechnologie MeSH
- reprodukovatelnost výsledků MeSH
- senzitivita a specificita MeSH
- tablety MeSH
- valsartan chemie MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
After a presence of highly hepatotoxic and potentially carcinogenic N-nitrosodimethylamine was detected in certain lots of sartan, ranitidine, metformin, and other pharmaceuticals, local regulatory authorities issued recalls of suspected products, and concerns of the pharmacotherapy safety were widely discussed. Since then, testing of a representative sample of each produced lot of these pharmaceuticals is required as a part of quality control processes. Hence, an interface-free CE-nanoESI system coupled with MS detection was employed for the development of a simple and economical method for quantitative detection of this contaminant in the valsartan drug substances and finished formulations used as model matrices. In this arrangement, a fused-silica capillary was used as both a separation column and a nanoESI emitter providing high ionization efficiency and sensitivity. The optimized procedure was found to have sufficient selectivity, linearity, accuracy, and precision. The established LOD and LOQ values were 0.3 and 1.0 ng/mL, respectively. The practical applicability of the method was tested by analyses of commercially available Valsacor® tablets. The results obtained prove that the developed procedure represents a promising alternative to currently available GC- and LC-based methods. Furthermore, after an adjustment of the separation conditions, the CE-nanoESI/MS system can be conceptually used for the determination of NDMA in other suspected pharmaceuticals.
Department of Chemistry Seoul National University Seoul Korea
Institute of Analytical Chemistry of the Czech Academy of Sciences v v i Brno Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22004643
- 003
- CZ-PrNML
- 005
- 20220127145108.0
- 007
- ta
- 008
- 220113s2021 gw f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1002/elps.202000303 $2 doi
- 035 __
- $a (PubMed)33368407
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a gw
- 100 1_
- $a Řemínek, Roman $u Institute of Analytical Chemistry of the Czech Academy of Sciences, v. v. i., Brno, Czech Republic
- 245 10
- $a Application of capillary electrophoresis-nano-electrospray ionization-mass spectrometry for the determination of N-nitrosodimethylamine in pharmaceuticals / $c R. Řemínek, F. Foret, DS. Chung
- 520 9_
- $a After a presence of highly hepatotoxic and potentially carcinogenic N-nitrosodimethylamine was detected in certain lots of sartan, ranitidine, metformin, and other pharmaceuticals, local regulatory authorities issued recalls of suspected products, and concerns of the pharmacotherapy safety were widely discussed. Since then, testing of a representative sample of each produced lot of these pharmaceuticals is required as a part of quality control processes. Hence, an interface-free CE-nanoESI system coupled with MS detection was employed for the development of a simple and economical method for quantitative detection of this contaminant in the valsartan drug substances and finished formulations used as model matrices. In this arrangement, a fused-silica capillary was used as both a separation column and a nanoESI emitter providing high ionization efficiency and sensitivity. The optimized procedure was found to have sufficient selectivity, linearity, accuracy, and precision. The established LOD and LOQ values were 0.3 and 1.0 ng/mL, respectively. The practical applicability of the method was tested by analyses of commercially available Valsacor® tablets. The results obtained prove that the developed procedure represents a promising alternative to currently available GC- and LC-based methods. Furthermore, after an adjustment of the separation conditions, the CE-nanoESI/MS system can be conceptually used for the determination of NDMA in other suspected pharmaceuticals.
- 650 _2
- $a dimethylnitrosamin $x analýza $7 D004128
- 650 12
- $a kontaminace léku $7 D004340
- 650 _2
- $a elektroforéza kapilární $x metody $7 D019075
- 650 _2
- $a lineární modely $7 D016014
- 650 _2
- $a nanotechnologie $7 D036103
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a senzitivita a specificita $7 D012680
- 650 _2
- $a hmotnostní spektrometrie s elektrosprejovou ionizací $x metody $7 D021241
- 650 _2
- $a tablety $7 D013607
- 650 _2
- $a valsartan $x chemie $7 D000068756
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Foret, František $u Institute of Analytical Chemistry of the Czech Academy of Sciences, v. v. i., Brno, Czech Republic
- 700 1_
- $a Chung, Doo Soo $u Department of Chemistry, Seoul National University, Seoul, Korea
- 773 0_
- $w MED00001508 $t Electrophoresis $x 1522-2683 $g Roč. 42, č. 4 (2021), s. 334-341
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33368407 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220113 $b ABA008
- 991 __
- $a 20220127145104 $b ABA008
- 999 __
- $a ok $b bmc $g 1751951 $s 1155792
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 42 $c 4 $d 334-341 $e 20210114 $i 1522-2683 $m Electrophoresis $n Electrophoresis $x MED00001508
- LZP __
- $a Pubmed-20220113