Newly Designed 3D-Printed Sonication Test Cell Optimized for In Vitro Sonication Experiments
M. Snehota, M. Kolarikova, J. Vachutka, B. Hosikova, K. Balazova, L. Dolezal, H. Kolarova
Language English Country England, Great Britain
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Printing, Three-Dimensional MeSH
- High-Intensity Focused Ultrasound Ablation * methods MeSH
- Silicones MeSH
- Ultrasonography MeSH
- Sonication * methods MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
OBJECTIVE: Precise control over the ultrasound field parameters experienced by biological samples during sonication experiments in vitro may be quite challenging. The main goal of this work was to outline an approach to construction of sonication test cells that would minimize the interaction between the test cells and ultrasound. METHODS: Optimal dimensions of the test cell were determined through measurements conducted in a water sonication tank using 3D-printed test objects. The offset of local acoustic intensity variability inside the sonication test cell was set to value of ±50% of the reference value (i.e., local acoustic intensity measured at last axial maximum in the free-field condition). The cytotoxicity of several materials used for 3D printing was determined using the MTT (3-[4,5-dimethylthiazol-2-yl]-2,5 diphenyl tetrazolium bromide) assay. RESULTS: The sonication test cells were 3D printed from polylactic acid material, which was not toxic to the cells. Silicone membrane HT-6240, which was used to construct the bottom of the test cell, was found to reduce ultrasound energy minimally. Final ultrasound profiles inside the sonication test cells indicated the desired variability of local acoustic intensity. The cell viability in our sonication test cell was comparable to that of commercial culture plates with bottoms constructed with silicone membrane. CONCLUSION: An approach to construction of sonication test cells minimizing the interaction of the test cell and ultrasound has been outlined.
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23016409
- 003
- CZ-PrNML
- 005
- 20231026105905.0
- 007
- ta
- 008
- 231013s2023 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.ultrasmedbio.2023.05.009 $2 doi
- 035 __
- $a (PubMed)37330389
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Snehota, Martin $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 245 10
- $a Newly Designed 3D-Printed Sonication Test Cell Optimized for In Vitro Sonication Experiments / $c M. Snehota, M. Kolarikova, J. Vachutka, B. Hosikova, K. Balazova, L. Dolezal, H. Kolarova
- 520 9_
- $a OBJECTIVE: Precise control over the ultrasound field parameters experienced by biological samples during sonication experiments in vitro may be quite challenging. The main goal of this work was to outline an approach to construction of sonication test cells that would minimize the interaction between the test cells and ultrasound. METHODS: Optimal dimensions of the test cell were determined through measurements conducted in a water sonication tank using 3D-printed test objects. The offset of local acoustic intensity variability inside the sonication test cell was set to value of ±50% of the reference value (i.e., local acoustic intensity measured at last axial maximum in the free-field condition). The cytotoxicity of several materials used for 3D printing was determined using the MTT (3-[4,5-dimethylthiazol-2-yl]-2,5 diphenyl tetrazolium bromide) assay. RESULTS: The sonication test cells were 3D printed from polylactic acid material, which was not toxic to the cells. Silicone membrane HT-6240, which was used to construct the bottom of the test cell, was found to reduce ultrasound energy minimally. Final ultrasound profiles inside the sonication test cells indicated the desired variability of local acoustic intensity. The cell viability in our sonication test cell was comparable to that of commercial culture plates with bottoms constructed with silicone membrane. CONCLUSION: An approach to construction of sonication test cells minimizing the interaction of the test cell and ultrasound has been outlined.
- 650 12
- $a vibrace ultrazvukové $x metody $7 D013010
- 650 12
- $a ablace intenzivním ultrazvukovým paprskem $x metody $7 D057086
- 650 _2
- $a ultrasonografie $7 D014463
- 650 _2
- $a 3D tisk $7 D066330
- 650 _2
- $a silikony $7 D012828
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Kolarikova, Marketa $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 700 1_
- $a Vachutka, Jaromir $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 700 1_
- $a Hosikova, Barbora $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic. Electronic address: barbora.hosikova@upol.cz
- 700 1_
- $a Balazova, Klara $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 700 1_
- $a Dolezal, Ladislav $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 700 1_
- $a Kolarova, Hana $u Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University Olomouc, Olomouc, Czech Republic
- 773 0_
- $w MED00004601 $t Ultrasound in medicine & biology $x 1879-291X $g Roč. 49, č. 9 (2023), s. 2042-2053
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/37330389 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20231013 $b ABA008
- 991 __
- $a 20231026105859 $b ABA008
- 999 __
- $a ok $b bmc $g 2000119 $s 1202771
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2023 $b 49 $c 9 $d 2042-2053 $e 20230615 $i 1879-291X $m Ultrasound in medicine & biology $n Ultrasound Med Biol $x MED00004601
- LZP __
- $a Pubmed-20231013