Accuracy and Feasibility of Real-time Continuous Glucose Monitoring in Critically Ill Patients After Abdominal Surgery and Solid Organ Transplantation

B. Voglová Hagerf, M. Protus, L. Nemetova, M. Mraz, E. Kieslichova, E. Uchytilova, V. Indrova, J. Lelito, P. Girman, M. Haluzík, J. Franekova, V. Svirlochova, DC. Klonoff, MA. Kohn, A. Jabor

. 2024 ; 47 (6) : 956-963. [pub] 20240601

Language English Country United States

Document type Journal Article

Grant support
National Institute for Research of Metabolic and Cardiovascular Diseases (Programme EXCELES, ID Project No. LX22NPO5104) - Funded by the European Union - Next Generation EU) MH CZ - DRO (Institute for Clinical and Experimental Medicine - IKEM, IN 00023001)
Cooperatio Programme, section Medical Diagnostics and Basic Medical Sciences, Charles University, Prague, Czech Republic

OBJECTIVE: Glycemia management in critical care is posing a challenge in frequent measuring and adequate insulin dose adjustment. In recent years, continuous glucose measurement has gained accuracy and reliability in outpatient and inpatient settings. The aim of this study was to assess the feasibility and accuracy of real-time continuous glucose monitoring (CGM) in ICU patients after major abdominal surgery. RESEARCH DESIGN AND METHODS: We included patients undergoing pancreatic surgery and solid organ transplantation (liver, pancreas, islets of Langerhans, kidney) requiring an ICU stay after surgery. We used a Dexcom G6 sensor, placed in the infraclavicular region, for real-time CGM. Arterial blood glucose measured by the amperometric principle (ABL 800; Radiometer, Copenhagen, Denmark) served as a reference value and for calibration. Blood glucose was also routinely monitored by a StatStrip bedside glucose meter. Sensor accuracy was assessed by mean absolute relative difference (MARD), bias, modified Bland-Altman plot, and surveillance error grid for paired samples of glucose values from CGM and acid-base analyzer (ABL). RESULTS: We analyzed data from 61 patients and obtained 1,546 paired glucose values from CGM and ABL. Active sensor use was 95.1%. MARD was 9.4%, relative bias was 1.4%, and 92.8% of values fell in zone A, 6.1% fell in zone B, and 1.2% fell in zone C of the surveillance error grid. Median time in range was 78%, with minimum (<1%) time spent in hypoglycemia. StatStrip glucose meter MARD compared with ABL was 5.8%. CONCLUSIONS: Our study shows clinically applicable accuracy and reliability of Dexcom G6 CGM in postoperative ICU patients and a feasible alternative sensor placement site.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc24013771
003      
CZ-PrNML
005      
20240905133342.0
007      
ta
008      
240725s2024 xxu f 000 0|eng||
009      
AR
024    7_
$a 10.2337/dc23-1663 $2 doi
035    __
$a (PubMed)38412005
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxu
100    1_
$a Voglová Hagerf, Barbora $u Department of Diabetes, Diabetes Centre, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $u First Faculty of Medicine, Charles University, Prague, Czech Republic $1 https://orcid.org/0000000171325134
245    10
$a Accuracy and Feasibility of Real-time Continuous Glucose Monitoring in Critically Ill Patients After Abdominal Surgery and Solid Organ Transplantation / $c B. Voglová Hagerf, M. Protus, L. Nemetova, M. Mraz, E. Kieslichova, E. Uchytilova, V. Indrova, J. Lelito, P. Girman, M. Haluzík, J. Franekova, V. Svirlochova, DC. Klonoff, MA. Kohn, A. Jabor
520    9_
$a OBJECTIVE: Glycemia management in critical care is posing a challenge in frequent measuring and adequate insulin dose adjustment. In recent years, continuous glucose measurement has gained accuracy and reliability in outpatient and inpatient settings. The aim of this study was to assess the feasibility and accuracy of real-time continuous glucose monitoring (CGM) in ICU patients after major abdominal surgery. RESEARCH DESIGN AND METHODS: We included patients undergoing pancreatic surgery and solid organ transplantation (liver, pancreas, islets of Langerhans, kidney) requiring an ICU stay after surgery. We used a Dexcom G6 sensor, placed in the infraclavicular region, for real-time CGM. Arterial blood glucose measured by the amperometric principle (ABL 800; Radiometer, Copenhagen, Denmark) served as a reference value and for calibration. Blood glucose was also routinely monitored by a StatStrip bedside glucose meter. Sensor accuracy was assessed by mean absolute relative difference (MARD), bias, modified Bland-Altman plot, and surveillance error grid for paired samples of glucose values from CGM and acid-base analyzer (ABL). RESULTS: We analyzed data from 61 patients and obtained 1,546 paired glucose values from CGM and ABL. Active sensor use was 95.1%. MARD was 9.4%, relative bias was 1.4%, and 92.8% of values fell in zone A, 6.1% fell in zone B, and 1.2% fell in zone C of the surveillance error grid. Median time in range was 78%, with minimum (<1%) time spent in hypoglycemia. StatStrip glucose meter MARD compared with ABL was 5.8%. CONCLUSIONS: Our study shows clinically applicable accuracy and reliability of Dexcom G6 CGM in postoperative ICU patients and a feasible alternative sensor placement site.
650    _2
$a lidé $7 D006801
650    _2
$a mužské pohlaví $7 D008297
650    12
$a krevní glukóza $x analýza $7 D001786
650    _2
$a lidé středního věku $7 D008875
650    _2
$a ženské pohlaví $7 D005260
650    12
$a kritický stav $7 D016638
650    _2
$a senioři $7 D000368
650    _2
$a břicho $x chirurgie $7 D000005
650    _2
$a transplantace orgánů $7 D016377
650    _2
$a studie proveditelnosti $7 D005240
650    _2
$a dospělí $7 D000328
650    _2
$a monitorování fyziologických funkcí $x metody $7 D008991
650    _2
$a kontinuální monitorování glukózy $7 D000095583
655    _2
$a časopisecké články $7 D016428
700    1_
$a Protus, Marek $u First Faculty of Medicine, Charles University, Prague, Czech Republic $u Department of Anesthesiology, Resuscitation and Intensive Care, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $1 https://orcid.org/0000000265956877
700    1_
$a Nemetova, Lenka $u Department of Diabetes, Diabetes Centre, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
700    1_
$a Mraz, Milos $u Department of Diabetes, Diabetes Centre, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $u First Faculty of Medicine, Charles University, Prague, Czech Republic $1 https://orcid.org/0000000153224612 $7 xx0119173
700    1_
$a Kieslichova, Eva $u First Faculty of Medicine, Charles University, Prague, Czech Republic $u Department of Anesthesiology, Resuscitation and Intensive Care, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $1 https://orcid.org/000000023930623X
700    1_
$a Uchytilova, Eva $u First Faculty of Medicine, Charles University, Prague, Czech Republic $u Department of Anesthesiology, Resuscitation and Intensive Care, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $1 https://orcid.org/0000000204124137
700    1_
$a Indrova, Veronika $u Department of Anesthesiology, Resuscitation and Intensive Care, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
700    1_
$a Lelito, Jan $u Department of Anesthesiology, Resuscitation and Intensive Care, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
700    1_
$a Girman, Peter $u Department of Diabetes, Diabetes Centre, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $1 https://orcid.org/0000000309448189 $7 xx0081892
700    1_
$a Haluzík, Martin $u Department of Diabetes, Diabetes Centre, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $u First Faculty of Medicine, Charles University, Prague, Czech Republic $1 https://orcid.org/0000000202016888 $7 xx0000707
700    1_
$a Franekova, Janka $u Department of Laboratory Methods, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $u Third Faculty of Medicine, Charles University, Prague, Czech Republic $1 https://orcid.org/0000000250947724
700    1_
$a Svirlochova, Veronika $u Department of Laboratory Methods, Institute for Clinical and Experimental Medicine, Prague, Czech Republic
700    1_
$a Klonoff, David C $u Diabetes Research Institute, Mills-Peninsula Medical Center, San Mateo, CA $1 https://orcid.org/0000000163946862 $7 xx0073553
700    1_
$a Kohn, Michael A $u University of California San Francisco, San Francisco, CA $1 https://orcid.org/0000000154595044
700    1_
$a Jabor, Antonin $u Department of Laboratory Methods, Institute for Clinical and Experimental Medicine, Prague, Czech Republic $u Third Faculty of Medicine, Charles University, Prague, Czech Republic $1 https://orcid.org/0000000241443129
773    0_
$w MED00001380 $t Diabetes care $x 1935-5548 $g Roč. 47, č. 6 (2024), s. 956-963
856    41
$u https://pubmed.ncbi.nlm.nih.gov/38412005 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y - $z 0
990    __
$a 20240725 $b ABA008
991    __
$a 20240905133336 $b ABA008
999    __
$a ok $b bmc $g 2143527 $s 1225637
BAS    __
$a 3
BAS    __
$a PreBMC-MEDLINE
BMC    __
$a 2024 $b 47 $c 6 $d 956-963 $e 20240601 $i 1935-5548 $m Diabetes care $n Diabetes Care $x MED00001380
GRA    __
$p National Institute for Research of Metabolic and Cardiovascular Diseases (Programme EXCELES, ID Project No. LX22NPO5104) - Funded by the European Union - Next Generation EU) MH CZ - DRO (Institute for Clinical and Experimental Medicine - IKEM, IN 00023001)
GRA    __
$p Cooperatio Programme, section Medical Diagnostics and Basic Medical Sciences, Charles University, Prague, Czech Republic
LZP    __
$a Pubmed-20240725

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...