Detection of bisphenol A using a novel surface plasmon resonance biosensor

. 2010 Nov ; 398 (5) : 1963-6. [epub] 20100812

Jazyk angličtina Země Německo Médium print-electronic

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid20714891

We present a compact surface plasmon resonance (SPR) biosensor for the detection of bisphenol A (BpA), an endocrine-disrupting chemical. The biosensor is based on an SPR sensor platform (SPRCD) and the binding inhibition detection format. The detection of BpA in PBS and wastewater was performed at concentrations ranging from 0.05 to 1,000 ng/ml. The limit of detection for BpA in PBS and wastewater was estimated to be 0.08 and 0.14 ng/ml, respectively. It was also demonstrated that the biosensor can be regenerated for repeated use. Results achieved with the SPR biosensor are compared with those obtained using ELISA and HPLC methods.

Citace poskytuje Crossref.org

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...