Real-time monitoring of biomolecular interactions in blood plasma using a surface plasmon resonance biosensor
Language English Country Germany Media print-electronic
Document type Journal Article, Research Support, Non-U.S. Gov't
- MeSH
- Biosensing Techniques methods MeSH
- Chorionic Gonadotropin blood chemistry MeSH
- Humans MeSH
- Microfluidics * MeSH
- Surface Plasmon Resonance methods MeSH
- Antibodies chemistry MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
- Names of Substances
- Chorionic Gonadotropin MeSH
- Antibodies MeSH
We report a novel approach to biosensor-based observations of biomolecular interactions which enables real-time monitoring of biomolecular interactions in complex media. This approach is demonstrated by investigating the interaction between the human chorionic gonadotropin (hCG) and its antibody in blood plasma using a surface plasmon resonance biosensor and a dispersionless microfluidics system. The real-time binding data obtained in blood plasma are compared with those obtained in buffer and blood plasma using a conventional method. It is also demonstrated that the proposed approach can enhance the capability of the biosensor to detect biomolecules in complex samples in terms of detection time and sensitivity. In the model experiment, this approach is shown to enable direct detection of hCG in blood plasma at levels which are five times lower than those detected using the conventional detection approach.
References provided by Crossref.org
Interactions of 17β-Hydroxysteroid Dehydrogenase Type 10 and Cyclophilin D in Alzheimer's Disease