A worldwide experience of the management of battery failures and chronic device retrieval of the Nanostim leadless pacemaker
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články, multicentrická studie, randomizované kontrolované studie
PubMed
28705736
DOI
10.1016/j.hrthm.2017.07.004
PII: S1547-5271(17)30846-9
Knihovny.cz E-zdroje
- Klíčová slova
- Battery advisory, Battery failure, Extraction, Leadless pacemaker, Patient management, Retrieval,
- MeSH
- bezpečnost pacientů MeSH
- časové faktory MeSH
- design vybavení MeSH
- kardiostimulátor škodlivé účinky MeSH
- lidé MeSH
- management nemoci * MeSH
- následné studie MeSH
- odstranění implantátu metody MeSH
- retrospektivní studie MeSH
- selhání zařízení MeSH
- senioři MeSH
- srdeční arytmie terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- randomizované kontrolované studie MeSH
BACKGROUND: The Nanostim leadless pacemaker (LP) met the primary endpoints in an investigational device exemption trial, and was shown to be fully retrievable percutaneously. In October 2016, St Jude Medical issued a worldwide alert of a battery malfunction that caused lost pacing output and LP communication. OBJECTIVE: To report the battery failure mechanism and incidence and the worldwide patient management, including device retrieval experiences. METHODS: The affected LP battery is a custom lithium-carbon monofluoride cell. These were returned after failure and underwent analysis assessing electronics and battery performance. Data were collected in ongoing clinical studies when LPs were abandoned or retrieved. RESULTS: Of 1423 LPs implanted worldwide, there were 34 battery failures, occurring at 2.9 ± 0.4 years with no instances of associated patient injury. Analysis of returned batteries revealed an increase in battery resistance caused by insufficient electrolyte availability at the cathode/anode interface. A total of 66 of 73 retrieval attempts were successful (90.4%; implant duration range: 0.2-4.0 years). The LP docking button was inaccessible in 6 patients, and the docking button detached from the LP during retrieval in 1 patient. There was 1 case of arteriovenous fistula and another case of the LP docking button migrating into the pulmonary artery. There were also 115 non-LP retrieval patients after the advisory who received an additional pacemaker, with no adverse device-to-device interactions reported. CONCLUSION: As with standard pacers, LPs can have critical battery failures. Chronic retrieval of LPs is safe and efficacious.
Amsterdam Medical Center Amsterdam The Netherlands
CHRU Albert Michallon Grenoble France
Duke University Medical Center Durham North Carolina
Hospital Universitario de la Paz Madrid Spain
Kansas University Medical Center Kansas City Kansas
Kerckhoff Klinik Bad Nauheim Germany
Libin Cardiovascular Institute of Alberta Calgary Canada
Mayo Clinic Rochester Minnesota
Mount Sinai Medical Center New York New York
Na Homolce Hospital Prague Czech Republic
Sparrow Clinical Research Institute Lansing Michigan
Citace poskytuje Crossref.org
Transcatheter non-acute retrieval of the tine-based leadless ventricular pacemaker
Retrieval and replacement of a helix-fixation leadless pacemaker at 9 years post-implant