Miniaturized microdosimeters as LET monitors: First comparison of calculated and experimental data performed at the 62 MeV/u 12C beam of INFN-LNS with four different detectors
Jazyk angličtina Země Itálie Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články
PubMed
30139599
DOI
10.1016/j.ejmp.2018.07.004
PII: S1120-1797(18)31128-1
Knihovny.cz E-zdroje
- MeSH
- celková dávka radioterapie MeSH
- design vybavení MeSH
- fantomy radiodiagnostické MeSH
- izotopy uhlíku terapeutické užití MeSH
- kalibrace MeSH
- metoda Monte Carlo MeSH
- miniaturizace MeSH
- počítačová simulace MeSH
- polymethylmethakrylát MeSH
- radiometrie přístrojové vybavení MeSH
- radioterapie těžkými ionty přístrojové vybavení MeSH
- voda MeSH
- Publikační typ
- časopisecké články MeSH
- srovnávací studie MeSH
- Názvy látek
- izotopy uhlíku MeSH
- polymethylmethakrylát MeSH
- voda MeSH
PURPOSE: The aim of this paper is to investigate the limits of LET monitoring of therapeutic carbon ion beams with miniaturized microdosimetric detectors. METHODS: Four different miniaturized microdosimeters have been used at the 62 MeV/u 12C beam of INFN Southern National Laboratory (LNS) of Catania for this purpose, i.e. a mini-TEPC and a GEM-microdosimeter, both filled with propane gas, and a silicon and a diamond microdosimeter. The y-D (dose-mean lineal energy) values, measured at different depths in a PMMA phantom, have been compared withLET¯D (dose-mean LET) values in water, calculated at the same water-equivalent depth with a Monte Carlo simulation setup based on the GEANT4 toolkit. RESULTS: In these first measurements, no detector was found to be significantly better than the others as a LET monitor. The y-D relative standard deviation has been assessed to be 13% for all the detectors. On average, the ratio between y-D and LET¯D values is 0.9 ± 0.3, spanning from 0.73 ± 0.08 (in the proximal edge and Bragg peak region) to 1.1 ± 0.3 at the distal edge. CONCLUSIONS: All the four microdosimeters are able to monitor the dose-mean LET with the 11% precision up to the distal edge. In the distal edge region, the ratio of y-D to LET¯D changes. Such variability is possibly due to a dependence of the detector response on depth, since the particle mean-path length inside the detectors can vary, especially in the distal edge region.