Assessment of MR stereotactic imaging and image co-registration accuracy for 3 different MR scanners by 3 different methods/phantoms: phantom and patient study
Jazyk angličtina Země Spojené státy americké Médium print
Typ dokumentu hodnotící studie, časopisecké články
PubMed
30544292
DOI
10.3171/2018.7.gks181527
PII: 2018.7.GKS181527
Knihovny.cz E-zdroje
- Klíčová slova
- CT = computed tomography, MRI = magnetic resonance imaging, PET = positron emission tomography, image co-registration, phantom measurement, stereotactic MR imaging, stereotactic methods, stereotactic radiosurgery,
- MeSH
- anatomické modely MeSH
- artefakty MeSH
- fantomy radiodiagnostické * MeSH
- lebka MeSH
- lidé MeSH
- magnetická rezonanční tomografie přístrojové vybavení metody MeSH
- počítačová rentgenová tomografie MeSH
- počítačové zpracování obrazu MeSH
- reprodukovatelnost výsledků MeSH
- software MeSH
- stereotaktické techniky * přístrojové vybavení MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- hodnotící studie MeSH
OBJECTIVEThe aim of this study was to compare 3 different methods to assess the geometrical distortion of two 1.5-T and one 3-T magnetic resonance (MR) scanners and to evaluate co-registration accuracy. The overall uncertainty of each particular method was also evaluated.METHODSThree different MR phantoms were used: 2 commercial CIRS skull phantoms and PTGR known target phantom and 1 custom cylindrical Perspex phantom made in-house. All phantoms were fixed in the Leksell stereotactic frame and examined by a Siemens Somatom CT unit, two 1.5-T Siemens (Avanto and Symphony) MRI systems, and one 3-T Siemens (Skyra) MRI system. The images were evaluated using Leksell GammaPlan software, and geometrical deviation of the selected points from the reference values were determined. The deviations were further investigated for both definitions including fiducial-based and co-registration-based in the case of the CIRS phantom images. The same co-registration accuracy assessment was also performed for a clinical case. Patient stereotactic imaging was done on 3-T Skyra, 1.5-T Avanto, and CT scanners.RESULTSThe accuracy of the CT scanner was determined as 0.10, 0.30, and 0.30 mm for X, Y, and Z coordinates, respectively. The total estimated uncertainty in distortion measurement in one coordinate was determined to be 0.32 mm and 0.14 mm, respectively, for methods using and not using CT as reference imaging. Slightly more significant distortions were observed when using the 3-T than either 1.5-T MR units. However, all scanners were comparable within the estimated measurement error. Observed deviation/distortion for individual X, Y, and Z stereotactic coordinates was typically within 0.50 mm for all 3 scanners and all 3 measurement methods employed. The total radial deviation/distortion was typically within 1.00 mm. Maximum total radial distortion was observed when the CIRS phantom was used; 1.08 ± 0.49 mm, 1.15 ± 0.48 mm, and 1.35 ± 0.49 mm for Symphony, Avanto, and Skyra, respectively. The co-registration process improved image stereotactic definition in a clinical case in which fiducial-based stereotactic definition was not accurate; this was demonstrated for 3-T stereotactic imaging in this study. The best results were shown for 3-T MR image co-registration with CT images improving image stereotactic definition by about 0.50 mm. The results obtained with patient data provided a similar trend of improvement in stereotactic definition by co-registration.CONCLUSIONSAll 3 methods/phantoms used were evaluated as satisfactory for the image distortion measurement. The method using the PTGR phantom had the lowest uncertainty as no reference CT imaging was needed. Image co-registration can improve stereotactic image definition when fiducial-based definition is not accurate.