A Novel Temperature-Controlled Radiofrequency Catheter Ablation System Used to Treat Patients With Paroxysmal Atrial Fibrillation
Language English Country United States Media print-electronic
Document type Journal Article, Multicenter Study, Randomized Controlled Trial, Research Support, Non-U.S. Gov't
PubMed
33516712
DOI
10.1016/j.jacep.2020.11.009
PII: S2405-500X(20)31190-7
Knihovny.cz E-resources
- Keywords
- DiamondTemp, atrial fibrillation, catheter ablation, contact force, radiofrequency,
- MeSH
- Atrial Fibrillation * surgery MeSH
- Catheter Ablation * adverse effects MeSH
- Catheters MeSH
- Quality of Life MeSH
- Humans MeSH
- Prospective Studies MeSH
- Temperature MeSH
- Treatment Outcome MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Multicenter Study MeSH
- Research Support, Non-U.S. Gov't MeSH
- Randomized Controlled Trial MeSH
OBJECTIVES: DIAMOND-AF (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation) was a prospective, multicenter, noninferiority, randomized trial that compared the safety and effectiveness of the DTA system versus those of a force-sensing RF ablation system (control) for the treatment of patients with drug-refractory, recurrent, symptomatic paroxysmal atrial fibrillation (AF). BACKGROUND: Irrigated radiofrequency (RF) ablation catheters lose tissue temperature acuity, which is vital in assessing lesion formation. DiamondTemp Ablation (DTA) was designed to re-establish accurate tissue temperature measurements during ablation. METHODS: A total of 482 patients with paroxysmal AF were randomized (239 DTA, 243 control) to undergo pulmonary vein isolation and were followed up at 23 sites. Patients were screened for disease progression, cardiac characteristics, and prior interventions. Primary endpoints were effectiveness (freedom from atrial arrhythmia recurrence) and safety (composite of procedure- and device-related serious adverse events). RESULTS: The primary safety event rate was 3.3% in the DTA group versus 6.6% in the control group (p < 0.001 vs. 6.5% noninferiority margin). Primary effectiveness was met in 79.1% of DTA subjects and 75.7% of control subjects (p < 0.001 vs. -12.5% noninferiority margin). Secondary endpoint analysis found that off-drug effectiveness favored DTA compared with the control (142 [59.4%] vs. 120 [49.4%], respectively; p = 0.03). Total RF time and individual RF ablation duration were significantly shorter with less saline infused through the DTA catheter (p < 0.001). Both arms saw clinically meaningful improvements in quality of life at 12 months. CONCLUSIONS: Safety and efficacy of the DTA system proved noninferior to force-sensing RF ablation in a paroxysmal AF population. Efficiencies were observed using DTA with shorter total RF times, individual RF ablation durations, and less saline infusion. (DiamondTemp™ Ablation System for the Treatment of Paroxysmal Atrial Fibrillation; NCT03334630).
Clinique du Tonkin Valleurbanne France
Clinique Pasteur Toulouse France
Icahn School of Medicine at Mount Sinai New York New York USA
Institute for Clinical and Experimental Medicine Prague Czech Republic
Jackson Heart Clinic Jackson Mississippi USA
Medtronic Inc Minneapolis Minnesota USA
Na Homolce Hospital Prague Czech Republic
St Anne's University Hospital Brno Czech Republic
Texas Cardiac Arrhythmia Institute at St David's Medical Center Austin Texas USA
References provided by Crossref.org
ClinicalTrials.gov