Characteristics of the ID-NOW™ test for the rapid detection of SARS-CoV-2
Language English Country Czech Republic Media print
Document type Journal Article
PubMed
37185021
PII: 134136
Knihovny.cz E-resources
- Keywords
- ID-NOW™, POCT, SARS-CoV-2, sensitivity,
- MeSH
- COVID-19 * diagnosis MeSH
- Adult MeSH
- Clinical Laboratory Techniques MeSH
- Humans MeSH
- SARS-CoV-2 * genetics MeSH
- Sensitivity and Specificity MeSH
- COVID-19 Testing MeSH
- Check Tag
- Adult MeSH
- Humans MeSH
- Publication type
- Journal Article MeSH
OBJECTIVE: To determine the sensitivity and specificity of the Abbott ID-NOW™ test in the diagnosis of COVID-19. The test is based on the detection of the SARS-CoV-2 gene by isothermal amplification technology. METHODS: From 303 individuals, two nasopharyngeal swabs and one oropharyngeal swab were collected to be tested in parallel by the ID-NOW™ test and PCR test (Allplex™ SARS-CoV-2 Assay). A subgroup of 107 individuals presented to the public collection point for covid-19 at the Motol University Hospital during the dominance of the Delta variant, and the others were tested via the Adult Emergency Admission Department during the dominance of the Omicron variant. RESULTS: Of 297 valid samples, 43 were positive by the PCR assay and 33 were positive by the ID-NOW™ test (sensitivity 76.74%; 95% CI 61.37 to 88.24%). ID-NOW™ detected three samples as positive, but the positivity was not confirmed by PCR (specificity 98.82%; 95% CI 96.59 to 99.76%). A significant increase in sensitivity up to 100% is observed for samples with a higher viral load (with a PCR threshold cycle value below 30 or from patients with symptoms of COVID-19). The Delta or Omicron variant has no significant effect on the sensitivity of the test. CONCLUSION: Due to its ease of use and speed of result, ID-NOW™ is a suitable diagnostic tool for prompt assessment of a patient's infectivity. If, despite the negative ID-NOW™ result, the patient has symptoms of COVID-19, it is advised to perform a classic PCR test for SARS-CoV-2.