History of blood pressure measurement in newborns and infants

. 2023 Nov 28 ; 72 (5) : 543-555.

Language English Country Czech Republic Media print

Document type Journal Article, Review

The development of methods for measuring blood pressure (BP) in newborns and small children has a rich history. Methods for BP measuring in adults had to be adapted to this age group. For measuring BP in direct invasive way, a suitable approach had to be found to access the arterial circulation through the umbilical and later radialis artery. Currently, results obtained from direct invasive BP measurement are considered the "gold standard". The development of non-invasive methods for BP measuring in newborns and children began with the use of von Basch's sphygmomanometer (1880). In 1899, Gustav Gärtner constructed the device, which was the basis for the flush method. After the discovery of the palpation and auscultation methods, these methods were also used for BP measurement in newborns and children, however, the BP values obtained in these ways were typically underestimated using excessively wide cuffs. From the auscultation method, methods utilizing ultrasound and infrasound to detect arterial wall movement and blood flow were later developed. The oscillometric method for BP measurement was introduced by E. J. Marey so early as in 1876. In 1912, P. Balard used the oscillometric technique to measure blood pressure in a large group of newborns. Through different types of oscillometers using various methods for detecting vascular oscillations (such as xylol method, impedance and volume plethysmography, etc.), the development has continued to assessment of vascular oscillations by modern sensor technology and software. For continuous non-invasive blood pressure measurement, the volume-clamp method, first described by Jan Peňáz in 1968, was developed. After modification for use in newborns, application of the cuff to the wrist instead of the finger, it is primarily used in clinical physiological studies to evaluate beat-to-beat BP and heart rate pressure variability, such as in the determination of the baroreflex sensitivity.

See more in PubMed

Hales S. Statical essays: containing Haemastaticks. An account of some hydraulick and hydrostatical experiments made on the blood and blood vessels of animals. Royal Society of London. 1733;2:1–385. doi: 10.5962/bhl.title.106596. DOI

Faivre J. Études experimentales sur les lésions organiques du coeur. Imprimerie d’Aimé Vingtrinier; Lyon: 1856. pp. 1–54.

Haselhorst G. Zum plazentaaren Kreislauf unter der Geburt. Zeitschrift fur Geburtshilfe und Gynäkologie. 1929;95:32–42.

Woodbury RA, Robinow M, Hamilton F. Blood pressure studies on infants. Am J Physiol. 1938;122:472–479. doi: 10.1152/ajplegacy.1938.122.2.472. DOI

Robinow M, Hamilton WF, Woodbury RA, Volpitto PP. Accuracy of clinical determinations of blood pressure in children with values under normal and abnormal conditions. Am J Dis Child. 1939;58:102–118. doi: 10.1001/archpedi.1939.01990070114011. DOI

Wallgren G, Karlberg P, Lind J. Studies of the circulatory adaptation immediately after birth. Acta Paediatr. 1960;49:843–849. doi: 10.1111/j.1651-2227.1960.tb16094.x. PubMed DOI

Rudolph AM, Drorbough JE, Auld PM, Rudolph AJ, Nadas AS, Smith CA, Hubbell JP. Studies on the circulation in the neonatal period. The circulation in the respiratory distress syndrome. Pediatrics. 1961;27:551–566. doi: 10.1542/peds.27.4.551. PubMed DOI

Todres ID, Rogers MC, Shannon DC, Moylan FM, Ryan JF. Percutaneous catheterization of the radial artery in the critically ill neonate. J Pediat. 1975;87:273–275. doi: 10.1016/S0022-3476(75)80601-9. PubMed DOI

Herisson JBES. Measurement of the pulse. The sphygmomanometer: An instrument which makes every action of the arteries apparent to the eye. Lancet. 1834;23:22–27. doi: 10.1016/S0140-6736(02)96308-7. DOI

Vierordt K. Physiologie des Kinderalters. Tübingen. 1881:496.

Basch von S. Über die Messung des Blutdruckes am Menschen. Zeitschr Klin Med. 1880;2:79–96.

Hofmann E. Über Verblutung aus der Nabelschnur. Oesterr Jahrb Pädiatr. 1877;II:192.

Vierordt H. Anatomische Physiologische und Physikalische Daten und Tabellen zum Gebrauche fur Mediciner. Verlag Von Gustav Fischer; Jena: 1893. p. 400.

Zadek I. Die Messung des Blutdrucks am Menschen mittels des Basch’schen Apparatus. Z Klin Med. 1880/1881;2:509–551..

Eckert A. Beobachtung aus den Elisabeth-Kinderhospital in St. Petersburgh. Wratsch; 1882. p. 220.

Gärtner G. Über einen neuen Blutdruckmesser (Tonometer) Wien Med Woch. 1899;49:1412–1418.

Neu M. Experimentalle und Klinische Blutdruckuntersuchungen mit Gärtners Tonometer. Inaug Diss Universitätsbuchhandlung; Heidelberg: 1902. p. 172.

Trumpp J. Blutdruckmessungen an gesunden und kranken Säuglingen. Jahrb Kinderheilk. 1906;63:43–59.

Sladkoff. Blood Pressure in Children. St Petersburg: 1903.

Gundobin . Deutsche Ausgabe. von Rubinstein, Berlin: Allgemeine Med Verlagsanstalt; 1912. Die Besonderheiten des Kindesalters.

Cappe BE, Pallin IM. Systolic blood pressure determination in the newborn and infant. Anesthesiology. 1952;13:648–650. doi: 10.1097/00000542-195211000-00014. PubMed DOI

Goldring D, Wohltmann H. Flush method for blood pressure determination in newborn infants. J Pediat. 1952;40:285–289. doi: 10.1016/S0022-3476(52)80257-4. PubMed DOI

Ribeiro MAS, Fiori HH, Luz JH, Piva JP, Ribeiro NM, Fiori RM. Comparison of noninvasive techniques to measure blood pressure in newborns. J Pediatria (Rio J) 2011;87:57–62. doi: 10.1590/S0021-75572011000100010. PubMed DOI

O’Brien ET, O’Malley K. ABC of blood pressure measurement. Infancy and childhood. Br Med J. 1979:1048–1049. doi: 10.1136/bmj.2.6197.1048. PubMed DOI PMC

Forfar JO, Kibel MA. Blood pressure in the newborn estimated by the flush method. Arch Dis Child. 1956;31:126–130. doi: 10.1136/adc.31.156.126. PubMed DOI PMC

Virnig NL, Reynolds JW. Reliability of flush blood pressure measurements in the sick newborn infant. J Pediatr. 1974;84:594–598. doi: 10.1016/S0022-3476(74)80688-8. PubMed DOI

Riva-Rocci S. Un nuovo sfigmomanometro. Gazzetta Medica di Torino. 1896;50:981–1017.

Von Recklinghausen H. Über Blutdruckmessung beim Menschen. Arch Exp Pathol Pharmakol. 1901;46:78–132. doi: 10.1007/BF01977772. DOI

Reis RA, Chaloupka AJ. Blood pressure in the newborn following normal and pathological labor. Surg Gyn Obst. 1923;37:206–211.

Holland WW, Young IM. Neonatal blood pressure in relation to maturity, mode of delivery, and condition at birth. Br Med J. 1956;2:1331–1333. doi: 10.1136/bmj.2.5005.1331. PubMed DOI PMC

Korotkov NS. To the question of methods of determining the blood pressure. Reports of the Imperial Military Academy (Izvestie Imp Voiennomedicinskoi Akademii) 1905;11:365.

Booth J. A short history of blood pressure measurement. Proc R Soc Med. 1977;70:793–799. doi: 10.1177/003591577707001112. PubMed DOI PMC

Browne FJ, Dodds GH. The cause of hypertension in pre-eclamptic toxaemia. Lancet. 1936;1:1059–1060. doi: 10.1016/S0140-6736(01)37020-4. DOI

Long M, Dunlop JR, Holland WW. Blood pressure recording in children. Arch Dis Childhood. 1971;46:636–640. doi: 10.1136/adc.46.249.636. PubMed DOI PMC

Elseed AM, Shinebourne EA, Joseph MC. Assessment of techniques for measurement of blood pressure in infants and children. Arch Dis Childhood. 1973;48:932–936. doi: 10.1136/adc.48.12.932. PubMed DOI PMC

Satomura S. Ultrasonic Doppler method for the inspection of cardiac function. J Acoust Soc Am. 1957;29:1181–1185. doi: 10.1121/1.1908737. DOI

McCutcheon EP, Rushmer RF. Korotkoff sounds in experimental critique. Circulat Res. 1967;20:149–161. doi: 10.1161/01.RES.20.2.149. PubMed DOI

Stegal MF, Kardon MB, Kemmerer WT. Indirect measurement of arterial blood pressure by Doppler ultrasonic sphygmomanometry. J Apl Physiol. 1968;23:793–798. doi: 10.1152/jappl.1968.25.6.793. PubMed DOI

Hernandez A, Goldring D, Hartmann AF., Jr Measurement of blood pressure in infants and children by the Doppler ultrasonic technique. Pediatrics. 1971;48:788–794. doi: 10.1542/peds.48.5.788. PubMed DOI

Janis KM, Kemmerer WT, Hagood CO., Jr Doppler blood pressure measuement in infants and small children. J Pediatr Surg. 1971;6:70–72. doi: 10.1016/0022-3468(71)90672-5. PubMed DOI

Kellerová E, Kittová M, Kováčik P. Neinvazívna metóda merania krvného tlaku u novorodencov na princípe Dopplerovho fenoménu ultrazvuku. (Article in Slovak) Bratisl Lek Listy. 1978;70:409–418. PubMed

Kittnar M, Kellerová E. The age dependence of some cardiovascular parameters in children. Physiol Bohemoslov. 1978;27:253.

Javorka K, Zavarská L’. Zmeny systémového tlaku krvi a kardiorespiračních parametrov u nedonosených novorodencov počas fototerapie. (Article in Slovak) Cesk Pediat. 1990;45:230–232. PubMed

Javorka K, Zavarská L’. Tlak krvi u hypotrofických a nedonosených novorodencov. (Article in Slovak) Cesk Pediat. 1983;38:430–434. PubMed

Kellerova E. Metódy neinvazívneho merania krvného tlaku (in Slovak) In: Javorka K, Buchanec J, Kellerova E, editors. Krvný Obeh Plodov, Novorodencov, Detí a Adolescentov. Osveta, Martin: 1992. pp. 77–86.

Edwards RC, Goldberg AD, Bannister R, Raftery EB. The infrasound blood-pressure recorder. A clinical evaluation. Lancet. 1976;2:398–400. doi: 10.1016/S0140-6736(76)92410-7. PubMed DOI

Savage JM, Dillon MJ, Taylor JFN. Clinical evaluation and comparison of the Infrasonde, Arteriosonde, and mercury sphygmomanometer in measurement of blood pressure in children. Arch Dis Childhood. 1979;54:184–189. doi: 10.1136/adc.54.3.184. PubMed DOI PMC

Marey EJ. Travaux du laboratoire de M. Physiologie expérimentale. Vol. 2. Marey; Paris: 1876. Pression et vitesse du sang; pp. 236–371. DOI

Major RH. The history of taking the blood pressure. Ann Med His. 1930:47–55. PubMed PMC

Posey JA, Geddes LA, Moore WH. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure. Cardiovasc Res Cent Bull. 1969;8:15–25. PubMed

Hill L, Barnard H. A simple and accurate form of sphygmometer or arterial pressure gauge contrived for clinical use. Brit Med J. 1897 Oct 2;:904. doi: 10.1136/bmj.2.1918.904. PubMed DOI PMC

Erlanger J. A new instrument for determining the minimum and maximum blood-pressures in man. Johns Hopkins Hosp Rep. 1904;12:53–100.

Pachon V. Exposé des titres et travaux scientifiques. Paris: Libraires de l’Academie de Medicine; 1911. p. 122.

Ramsey M. Noninvasive automatic determination of mean arterial pressure. Med Biol Eng Comput. 1979;17:11–18. doi: 10.1007/BF02440948. PubMed DOI

Nunn DE, Beveridge RW. Apparatus and method for measuring blood pressure. 4427013 Patent N. 1984 January 24;

Jílek J, Štork M. Oscilometrické monitory krevního tlaku: metody měření, validace a koncept databáze oscilometrických tlakových vln. (in Czech) Lékař a Technika. 2004;35:160–164. doi: 10.3200/REVU.35.4.160-164. DOI

References on the physical growth and development of the normal child. Bureau Publ. N. 179. Edit. US Dept. Of Labor and US Children’s Bureau. US Government Printing Office; Washington: 1927. p. 355.

Koessler L. Thése. Paris: 1912. L’oscillométrie appliqué a l’étude de la tension artérielle chez les enfants; p. 192.

Kafka HL. Blood pressure measurement in infants. Arch Dis Child. 1974;49:970. doi: 10.1136/adc.49.12.970. PubMed DOI PMC

Balard P. Modifications évolutives du pouls et de la tension artérielle chez le nouveau-né, dans le premiers jours de la vie, étudiées par l’oscillométrie. Compt Rend Hebd Séances Memoires Soc Biol Paris. 1912a;73:483–485..

Balard P. Sur la cause de la diminuation de frequence du pouls chez le nouveau-né dans le premiérs heures de la vie. Vol. 73. Compt Rend Hebd Séances Memoires Soc Biol; Paris: 1912b. pp. 486–488.

Balard P. Le pouls et la tension artérielle de l’enfant et du nouveau-né. Gazette des hopitaux, Paris. 1913;86:837–841.

Balard P. La tension artérielle et l’oscillométrie chez le nouveau-né. Nourisson Paris. 1921;9:304–319.

Ashworth AM, Neligan GA, Rogers JE. Sphygmomanometer for the newborn. Lancet. 1959;1:801–803. doi: 10.1016/S0140-6736(59)91994-4. PubMed DOI

Ashworth AM, Neligan GA. Changes in the systolic blood pressure of normal babies during the first twenty-four hours of life. Lancet. 1959;1:804–807. doi: 10.1016/S0140-6736(59)91995-6. PubMed DOI

Schaffer AI. Neonatal blood pressure studies. AMA Am J Dis Child. 1955;89:204–209. doi: 10.1001/archpedi.1955.02050110246012. PubMed DOI

Lang VO, Hilber HM., Jr Zur Blutdruckmessung im Säuglingsalter. Z Kinder-Heilk. 1969;105:156–164. doi: 10.1007/BF00438937. PubMed DOI

Celander O, Thunell G. A plethysmographic method for measuring the systolic and diastolic blood pressure in newborn infants. Acta Paediatr. 1960;49:497–502. doi: 10.1111/j.1651-2227.1960.tb07764.x. PubMed DOI

Mason DT, Braunwald E. A simplified plethysmographic system for the measurement of systemic arterial pressure and peripheral blood flow. Amer Heart J. 1962;64:796–804. doi: 10.1016/0002-8703(62)90178-3. DOI

Gundersen J, Dahlin K. Measurement of systolic blood pressure in fingers of newborn infants. Acta Paediatr Scand. 1975;64:741–744. doi: 10.1111/j.1651-2227.1975.tb03913.x. PubMed DOI

Movius AJ, Bratton SL, Sorensen GK. Use of pulse oximetry for blood pressure measurement after cardiac surgery. Arch Dis Child. 1998;78:457–460. doi: 10.1136/adc.78.5.457. PubMed DOI PMC

Langbaum M, Eyal FG. A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry. J Pediatr. 1994;125:591–595. doi: 10.1016/S0022-3476(94)70016-8. PubMed DOI

Gupta JM, Scopes JW. Observations on blood pressure in newborn infants. Arch Dis Childhood. 1965;40:637–644. doi: 10.1136/adc.40.214.637. PubMed DOI PMC

Dionne JM, Bremner SA, Baygani SK, Batton B, Ergenekon E, Bhatt-Mehta V, Dempsey E, et al. Method of Blood Pressure Measurement in Neonates and Infants: A Systematic Review and Analysis. J Pediatr. 2020;221:23–31. doi: 10.1016/j.jpeds.2020.02.072. PubMed DOI

Penaz J. Photoelectric measurement of blood pressure, volume and flow in the finger. Dresden. Digest 10th Int Conf Med Biol Engng; 1973; p. 104.

Wesseling KH, Wit de B, Hoeven GMA, Goudoever J, Settels JJ. Physiocal, calibrating finger vascular physiology for finapres. Homeostasis Health Dis. 1995;36:67–82.

Jagomäki K, Talts J, Raamat R, Länsimies E. Continuous non-invasive measurement of mean blood pressure in fingers by volume - clamp and differential oscillometric method. Clin Physiol. 1996;16:551–560. doi: 10.1111/j.1475-097X.1996.tb01020.x. PubMed DOI

Lemson J, Hofhuizen CM, Schraa O, Settels JJ, Scheffer G-J, Van der Hoeven JG. The reliability of continuous noninvasive finger blood pressure measurement in critically ill children. Anesth Analg. 2009;108:814–821. doi: 10.1213/ane.0b013e318194f401. PubMed DOI

Drouin E, Gournay V, Calamel J, Mouzard A, Rozé JC. Feasibility of using finger arterial pressure in neonates. Arch Dis Child Fetal Neonatal Ed. 1997;77:F139–F140. doi: 10.1136/fn.77.2.F139. PubMed DOI PMC

Andriessen P, Schoffelen RL, Berendsen RC, de Beer NA, Oei SG, Wijn PF, Blanco CE. Noninvasive Assessment of Blood Pressure Variability in Preterm Infants. Pediatr Res. 2004;55:220–223. doi: 10.1203/01.PDR.0000104152.85296.4F. PubMed DOI

Yiallourou SR, Walker AM, Horne RS. Validation of a new noninvasive method to measure blood pressure and assess baroreflex sensitivity in preterm infants during sleep. Sleep. 2006;29:1083–1088. doi: 10.1093/sleep/29.8.1083. PubMed DOI

Andriessen P, Oetomo SB, Peters C, Vermeulen B, Wijn PF, Blanco CE. Baroreceptor reflex sensitivity in human neonates: The effect of postmenstrual age. J Physiol. 2005;568:333–341. doi: 10.1113/jphysiol.2005.093641. PubMed DOI PMC

Yiallourou SR, Sands SA, Walker AM, Horne RS. Postnatal development of baroreflex sensitivity in infancy. J Physiol. 2010;588:2193–2203. doi: 10.1113/jphysiol.2010.187070. PubMed DOI PMC

Fyfe KL, Yiallourou SR, Wong FY, Odoi A, Walker AM, Horne RS. Gestational age at birth affects maturation of baroreflex control. J Pediat. 2015;166:559–565. doi: 10.1016/j.jpeds.2014.11.026. PubMed DOI

Haskova K, Javorka M, Czippelova B, Zibolen M, Javorka K. Baroreflex sensitivity in premature infants - relation to the parameters characterizing intrauterine and postnatal condition. Physiol Res. 2017;66(Suppl 2):S257–S264. doi: 10.33549/physiolres.933681. PubMed DOI

Javorka K, Haskova K, Czippelova B, Zibolen M, Javorka M. Baroreflex sensitivity and blood pressure in premature infants - dependence on gestational age, postnatal age and sex. Physiol Res. 2021;70(Suppl 3):S349–S356. doi: 10.33549/physiolres.934829. PubMed DOI PMC

Javorka K, Haskova K, Czippelova B, Zibolen M, Javorka M: Blood pressure variability and baroreflex sensitivity in premature newborns-an effect of postconceptional and gestational age. Front Pediatr. 2021;9:653573. doi: 10.3389/fped.2021.653573. PubMed DOI PMC

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...