In a pig bone model mimicking the human trapezium, we assessed the cementless fixation of trapezial cups in cancellous and cortical bone. Thirty-two saddle-shaped pig forefoot bones were prepared for cup fixation. Conical Konos cups (Beznoska, Kladno, Czech Republic) size 9 mm or 10 mm diameter were impacted. Cup migration was evaluated with repeated radiostereometry after intervals of cyclic-load tests. Migration increased for every load-interval up to 750 N and was higher with cancellous bone fixation than with cortical bone fixation. In cancellous bone, 9 mm cups migrated more than 10 mm cups. At the highest load (1050 N), the cumulative implant survival was 88% for 10 mm cups with cortical bone fixation and 13% for 9 mm cups with cancellous bone fixation. We conclude that mechanical fixation of conical Konos cups was better in cortical than in cancellous bone. Our results further suggest that the largest possible cup diameter should be used.
- MeSH
- acetabulum chirurgie MeSH
- náhrada kyčelního kloubu * MeSH
- prasata MeSH
- protézy - design MeSH
- selhání protézy MeSH
- trapézová kost * MeSH
- zvířata MeSH
- Check Tag
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
- Geografické názvy
- Česká republika MeSH