Non-Adaptive Methods for Fetal ECG Signal Processing: A Review and Appraisal
Non-invasive fetal electrocardiography appears to be one of the most promising fetal monitoring techniques during pregnancy and delivery nowadays. This method is based on recording electrical potentials produced by the fetal heart from the surface of the maternal abdomen. Unfortunately, in addition to the useful fetal electrocardiographic signal, there are other interference signals in the abdominal recording that need to be filtered. The biggest challenge in designing filtration methods is the suppression of the maternal electrocardiographic signal. This study focuses on the extraction of fetal electrocardiographic signal from abdominal recordings using a combination of independent component analysis, recursive least squares, and ensemble empirical mode decomposition. The method was tested on two databases, the Fetal Electrocardiograms, Direct and Abdominal with Reference Heartbeats Annotations and the PhysioNet Challenge 2013 database. The evaluation was performed by the assessment of the accuracy of fetal QRS complexes detection and the quality of fetal heart rate determination. The effectiveness of the method was measured by means of the statistical parameters as accuracy, sensitivity, positive predictive value, and F1-score. Using the proposed method, when testing on the Fetal Electrocardiograms, Direct and Abdominal with Reference Heartbeats Annotations database, accuracy higher than 80% was achieved for 11 out of 12 recordings with an average value of accuracy 92.75% [95% confidence interval: 91.19-93.88%], sensitivity 95.09% [95% confidence interval: 93.68-96.03%], positive predictive value 96.36% [95% confidence interval: 95.05-97.17%] and F1-score 95.69% [95% confidence interval: 94.83-96.35%]. When testing on the Physionet Challenge 2013 database, accuracy higher than 80% was achieved for 17 out of 25 recordings with an average value of accuracy 78.24% [95% confidence interval: 73.44-81.85%], sensitivity 81.79% [95% confidence interval: 76.59-85.43%], positive predictive value 87.16% [95% confidence interval: 81.95-90.35%] and F1-score 84.08% [95% confidence interval: 80.75-86.64%]. Moreover, the non-invasive ST segment analysis was carried out on the records from the Fetal Electrocardiograms, Direct and Abdominal with Reference Heartbeats Annotations database and achieved high accuracy in 7 from in total of 12 records (mean values μ < 0.1 and values of ±1.96σ < 0.1).
- MeSH
- Algorithms * MeSH
- Abdomen physiology MeSH
- Databases, Factual MeSH
- Electrocardiography methods MeSH
- Humans MeSH
- Mothers statistics & numerical data MeSH
- Fetal Monitoring methods MeSH
- Fetus physiology MeSH
- Signal Processing, Computer-Assisted instrumentation MeSH
- Heart Rate, Fetal physiology MeSH
- Pregnancy MeSH
- Check Tag
- Humans MeSH
- Pregnancy MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
Advanced signal processing methods are one of the fastest developing scientific and technical areas of biomedical engineering with increasing usage in current clinical practice. This paper presents an extensive literature review of the methods for the digital signal processing of cardiac bioelectrical signals that are commonly applied in today's clinical practice. This work covers the definition of bioelectrical signals. It also covers to the extreme extent of classical and advanced approaches to the alleviation of noise contamination such as digital adaptive and non-adaptive filtering, signal decomposition methods based on blind source separation and wavelet transform.
- Keywords
- biomedical signals, cardiac signals, electrocardiography, fetal electrocardiography, vectorcardiography,
- MeSH
- Algorithms * MeSH
- Electrocardiography * MeSH
- Humans MeSH
- Signal Processing, Computer-Assisted MeSH
- Heart MeSH
- Wavelet Analysis MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Review MeSH
The most commonly used method of fetal monitoring is based on heart activity analysis. Computer-aided fetal monitoring system enables extraction of clinically important information hidden for visual interpretation-the instantaneous fetal heart rate (FHR) variability. Today's fetal monitors are based on monitoring of mechanical activity of the fetal heart by means of Doppler ultrasound technique. The FHR is determined using autocorrelation methods, and thus it has a form of evenly spaced-every 250 ms-instantaneous measurements, where some of which are incorrect or duplicate. The parameters describing a beat-to-beat FHR variability calculated from such a signal show significant errors. The aim of our research was to develop new analysis methods that will both improve an accuracy of the FHR determination and provide FHR representation as time series of events. The study was carried out on simultaneously recorded (during labor) Doppler ultrasound signal and the reference direct fetal electrocardiogram Two subranges of Doppler bandwidths were separated to describe heart wall movements and valve motions. After reduction of signal complexity by determining the Doppler ultrasound envelope, the signal was analyzed to determine the FHR. The autocorrelation method supported by a trapezoidal prediction function was used. In the final stage, two different methods were developed to provide signal representation as time series of events: the first using correction of duplicate measurements and the second based on segmentation of instantaneous periodicity measurements. Thus, it ensured the mean heart interval measurement error of only 1.35 ms. In a case of beat-to-beat variability assessment the errors ranged from -1.9% to -10.1%. Comparing the obtained values to other published results clearly confirms that the new methods provides a higher accuracy of an interval measurement and a better reliability of the FHR variability estimation.
- Keywords
- Doppler ultrasound signal, beat-to-beat variability, fetal heart rate, fetal monitoring,
- MeSH
- Data Analysis MeSH
- Electrocardiography MeSH
- Humans MeSH
- Fetal Monitoring * MeSH
- Reproducibility of Results MeSH
- Heart Rate, Fetal * MeSH
- Heart Rate MeSH
- Pregnancy MeSH
- Ultrasonography, Doppler MeSH
- Check Tag
- Humans MeSH
- Pregnancy MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
Monitoring fetal heart rate (FHR) variability plays a fundamental role in fetal state assessment. Reliable FHR signal can be obtained from an invasive direct fetal electrocardiogram (FECG), but this is limited to labour. Alternative abdominal (indirect) FECG signals can be recorded during pregnancy and labour. Quality, however, is much lower and the maternal heart and uterine contractions provide sources of interference. Here, we present ten twenty-minute pregnancy signals and 12 five-minute labour signals. Abdominal FECG and reference direct FECG were recorded simultaneously during labour. Reference pregnancy signal data came from an automated detector and were corrected by clinical experts. The resulting dataset exhibits a large variety of interferences and clinically significant FHR patterns. We thus provide the scientific community with access to bioelectrical fetal heart activity signals that may enable the development of new methods for FECG signals analysis, and may ultimately advance the use and accuracy of abdominal electrocardiography methods.