Test-retest reliability of a pendant-worn sensor device in measuring chair rise performance in older persons
W. Zhang, GR. Regterschot, H. Schaabova, H. Baldus, W. Zijlstra,
Jazyk angličtina Země Švýcarsko
Typ dokumentu hodnotící studie, časopisecké články, práce podpořená grantem
NLK
Directory of Open Access Journals
od 2001
PubMed Central
od 2003
Europe PubMed Central
od 2003
ProQuest Central
od 2001-01-01
Open Access Digital Library
od 2001-01-01
Open Access Digital Library
od 2003-01-01
Health & Medicine (ProQuest)
od 2001-01-01
ROAD: Directory of Open Access Scholarly Resources
od 2001
PubMed
24841248
DOI
10.3390/s140508705
Knihovny.cz E-zdroje
- MeSH
- akcelerometrie přístrojové vybavení MeSH
- aktigrafie přístrojové vybavení MeSH
- ambulantní monitorování přístrojové vybavení MeSH
- analýza selhání vybavení MeSH
- design vybavení MeSH
- geriatrické hodnocení metody MeSH
- lidé MeSH
- pohyb fyziologie MeSH
- postura těla fyziologie MeSH
- posturální rovnováha fyziologie MeSH
- reprodukovatelnost výsledků MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- senzitivita a specificita MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- hodnotící studie MeSH
- práce podpořená grantem MeSH
Chair rise performance is incorporated in clinical assessments to indicate fall risk status in older persons. This study investigated the test-retest reliability of a pendant-sensor-based assessment of chair rise performance. Forty-one older persons (28 females, 13 males, age: 72-94) were assessed in two sessions with 3 to 8 days in between. Repeated chair rise transfers were measured after different instructions. Relative and absolute test-retest reliability of chair rise measurements in individual tests and average over all tests were evaluated by means of intra-class correlation coefficients (ICCs) and standard error of measurement (SEM) as a percentage of the measurement mean. Systematic bias between the measurements in test and retest was examined with paired t-tests. Heteroscedasticity of the measurements was visually checked with Bland-Altman plots. In the different test conditions, the ICCs ranged between 0.63 and 0.93, and the SEM% ranged between 5.7% and 21.2%. The relative and absolute reliability of the average over all tests were ICC = 0.86 and SEM% = 9.5% for transfer duration, ICC = 0.93 and SEM% = 9.2% for maximum vertical acceleration, and ICC = 0.89 and SEM% = 10.0% for peak power. The results over all tests indicated that a fall risk assessment application based on pendant-worn-sensor measured chair rise performance in daily life might be feasible.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc15023402
- 003
- CZ-PrNML
- 005
- 20150728091009.0
- 007
- ta
- 008
- 150709s2014 sz f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.3390/s140508705 $2 doi
- 035 __
- $a (PubMed)24841248
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a sz
- 100 1_
- $a Zhang, Wei $u Personal Health Solutions Department, Philips Research Europe, High Tech Campus 34, Eindhoven 5656AE, The Netherlands. wei.zhang01@philips.com.
- 245 10
- $a Test-retest reliability of a pendant-worn sensor device in measuring chair rise performance in older persons / $c W. Zhang, GR. Regterschot, H. Schaabova, H. Baldus, W. Zijlstra,
- 520 9_
- $a Chair rise performance is incorporated in clinical assessments to indicate fall risk status in older persons. This study investigated the test-retest reliability of a pendant-sensor-based assessment of chair rise performance. Forty-one older persons (28 females, 13 males, age: 72-94) were assessed in two sessions with 3 to 8 days in between. Repeated chair rise transfers were measured after different instructions. Relative and absolute test-retest reliability of chair rise measurements in individual tests and average over all tests were evaluated by means of intra-class correlation coefficients (ICCs) and standard error of measurement (SEM) as a percentage of the measurement mean. Systematic bias between the measurements in test and retest was examined with paired t-tests. Heteroscedasticity of the measurements was visually checked with Bland-Altman plots. In the different test conditions, the ICCs ranged between 0.63 and 0.93, and the SEM% ranged between 5.7% and 21.2%. The relative and absolute reliability of the average over all tests were ICC = 0.86 and SEM% = 9.5% for transfer duration, ICC = 0.93 and SEM% = 9.2% for maximum vertical acceleration, and ICC = 0.89 and SEM% = 10.0% for peak power. The results over all tests indicated that a fall risk assessment application based on pendant-worn-sensor measured chair rise performance in daily life might be feasible.
- 650 _2
- $a akcelerometrie $x přístrojové vybavení $7 D061725
- 650 _2
- $a aktigrafie $x přístrojové vybavení $7 D056044
- 650 _2
- $a senioři $7 D000368
- 650 _2
- $a senioři nad 80 let $7 D000369
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a analýza selhání vybavení $7 D019544
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a geriatrické hodnocení $x metody $7 D015577
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a ambulantní monitorování $x přístrojové vybavení $7 D018670
- 650 _2
- $a pohyb $x fyziologie $7 D009068
- 650 _2
- $a posturální rovnováha $x fyziologie $7 D004856
- 650 _2
- $a postura těla $x fyziologie $7 D011187
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a senzitivita a specificita $7 D012680
- 655 _2
- $a hodnotící studie $7 D023362
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Regterschot, G Ruben H $u Center for Human Movement Sciences, University Medical Center Groningen, University of Groningen, Antonius Deusinglaan 1, Groningen 9713 AV, The Netherlands. g.r.h.regterschot@umcg.nl.
- 700 1_
- $a Schaabova, Hana $u Faculty of Biomedical Engineering, Czech Technical University, Nám. Sítná 3105, Kladno 2 272 01, Czech Republic. hankasch@centrum.cz.
- 700 1_
- $a Baldus, Heribert $u Personal Health Solutions Department, Philips Research Europe, High Tech Campus 34, Eindhoven 5656AE, The Netherlands. heribert.baldus@philips.com.
- 700 1_
- $a Zijlstra, Wiebren $u Center for Human Movement Sciences, University Medical Center Groningen, University of Groningen, Antonius Deusinglaan 1, Groningen 9713 AV, The Netherlands. zijlstra@dshs-koeln.de.
- 773 0_
- $w MED00008309 $t Sensors (Basel, Switzerland) $x 1424-8220 $g Roč. 14, č. 5 (2014), s. 8705-17
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/24841248 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20150709 $b ABA008
- 991 __
- $a 20150728091054 $b ABA008
- 999 __
- $a ok $b bmc $g 1083739 $s 906395
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2014 $b 14 $c 5 $d 8705-17 $i 1424-8220 $m Sensors $n Sensors Basel $x MED00008309
- LZP __
- $a Pubmed-20150709