Feasibility Evaluation of Metamaterial Microwave Sensors for Non-Invasive Blood Glucose Monitoring

L. Malena, O. Fiser, PR. Stauffer, T. Drizdal, J. Vrba, D. Vrba

. 2021 ; 21 (20) : . [pub] 20211016

Jazyk angličtina Země Švýcarsko

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/bmc22003454

Grantová podpora
LTC19031; SGS21/179/OHK4/3T/17 Ministry of Education, Youth and Sports of the Czech Republic; Student Grant Competition of CTU

The use of microwave technology is currently under investigation for non-invasive estimation of glycemia in patients with diabetes. Due to their construction, metamaterial (MTM)-based sensors have the potential to provide higher sensitivity of the phase shift of the S21 parameter (∠S21) to changes in glucose concentration compared to standard microstrip transmission line (MSTL)-based sensors. In this study, a MSTL sensor and three MTM sensors with 5, 7, and 9 MTM unit cells are exposed to liquid phantoms with different dielectric properties mimicking a change in blood glucose concentration from 0 to 14 mmol/L. Numerical models were created for the individual experiments, and the calculated S-parameters show good agreement with experimental results, expressed by the maximum relative error of 8.89% and 0.96% at a frequency of 1.99 GHz for MSTL and MTM sensor with nine unit cells, respectively. MTM sensors with an increasing number of cells show higher sensitivity of 0.62° per mmol/L and unit cell to blood glucose concentration as measured by changes in ∠S21. In accordance with the numerical simulations, the MTM sensor with nine unit cells showed the highest sensitivity of the sensors proposed by us, with an average of 3.66° per mmol/L at a frequency of 1.99 GHz, compared to only 0.48° per mmol/L for the MSTL sensor. The multi-cell MTM sensor has the potential to proceed with evaluation of human blood samples.

Citace poskytuje Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc22003454
003      
CZ-PrNML
005      
20220127150221.0
007      
ta
008      
220113s2021 sz f 000 0|eng||
009      
AR
024    7_
$a 10.3390/s21206871 $2 doi
035    __
$a (PubMed)34696084
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a sz
100    1_
$a Malena, Lukas $u Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Prague, Czech Republic
245    10
$a Feasibility Evaluation of Metamaterial Microwave Sensors for Non-Invasive Blood Glucose Monitoring / $c L. Malena, O. Fiser, PR. Stauffer, T. Drizdal, J. Vrba, D. Vrba
520    9_
$a The use of microwave technology is currently under investigation for non-invasive estimation of glycemia in patients with diabetes. Due to their construction, metamaterial (MTM)-based sensors have the potential to provide higher sensitivity of the phase shift of the S21 parameter (∠S21) to changes in glucose concentration compared to standard microstrip transmission line (MSTL)-based sensors. In this study, a MSTL sensor and three MTM sensors with 5, 7, and 9 MTM unit cells are exposed to liquid phantoms with different dielectric properties mimicking a change in blood glucose concentration from 0 to 14 mmol/L. Numerical models were created for the individual experiments, and the calculated S-parameters show good agreement with experimental results, expressed by the maximum relative error of 8.89% and 0.96% at a frequency of 1.99 GHz for MSTL and MTM sensor with nine unit cells, respectively. MTM sensors with an increasing number of cells show higher sensitivity of 0.62° per mmol/L and unit cell to blood glucose concentration as measured by changes in ∠S21. In accordance with the numerical simulations, the MTM sensor with nine unit cells showed the highest sensitivity of the sensors proposed by us, with an average of 3.66° per mmol/L at a frequency of 1.99 GHz, compared to only 0.48° per mmol/L for the MSTL sensor. The multi-cell MTM sensor has the potential to proceed with evaluation of human blood samples.
650    12
$a krevní glukóza $7 D001786
650    12
$a selfmonitoring glykemie $7 D015190
650    _2
$a studie proveditelnosti $7 D005240
650    _2
$a lidé $7 D006801
650    _2
$a mikrovlny $7 D008872
650    _2
$a monitorování fyziologických funkcí $7 D008991
655    _2
$a časopisecké články $7 D016428
700    1_
$a Fiser, Ondrej $u Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Prague, Czech Republic
700    1_
$a Stauffer, Paul R $u Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, PA 19107, USA
700    1_
$a Drizdal, Tomas $u Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Prague, Czech Republic
700    1_
$a Vrba, Jan $u Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Prague, Czech Republic
700    1_
$a Vrba, David $u Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Prague, Czech Republic
773    0_
$w MED00008309 $t Sensors (Basel, Switzerland) $x 1424-8220 $g Roč. 21, č. 20 (2021)
856    41
$u https://pubmed.ncbi.nlm.nih.gov/34696084 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y p $z 0
990    __
$a 20220113 $b ABA008
991    __
$a 20220127150217 $b ABA008
999    __
$a ok $b bmc $g 1751033 $s 1154603
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2021 $b 21 $c 20 $e 20211016 $i 1424-8220 $m Sensors $n Sensors Basel $x MED00008309
GRA    __
$a LTC19031; SGS21/179/OHK4/3T/17 $p Ministry of Education, Youth and Sports of the Czech Republic; Student Grant Competition of CTU
LZP    __
$a Pubmed-20220113

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...