Transcervical microwave ablation in type 2 uterine fibroids via a hysteroscopic approach: analysis of ablation profiles
G. Zia, J. Sebek, J. Schenck, P. Prakash
Language English Country Great Britain
Document type Journal Article
- MeSH
- Leiomyoma * surgery MeSH
- Humans MeSH
- Microwaves * MeSH
- Myometrium MeSH
- Radiofrequency Ablation MeSH
- Check Tag
- Humans MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
Type 2 uterine fibroids are challenging to resect surgically as ≥ 50% volume of myoma lies within the myometrium. A hysteroscopic approach for ablating fibroids is minimally-invasive, but places a considerable burden on the operator to accurately place the ablation applicator within the target. We investigated the sensitivity of transcervical microwave ablation outcome with respect to position of the ablation applicator within 1 - 3 cm type 2 fibroids.Methods:A finite element computer model was developed to simulate 5.8 GHz microwave ablation of fibroids and validated with experiments inex vivotissue. The ablation outcome was evaluated with respect to applicator insertion angles (30°, 45°, 60°) , depth and offset from the fibroid center (±2 mm for 3 cm fibroid and ±1 mm for 1 cm fibroid) with 35 W and 15 W applied power for 3 cm and 1 cm fibroids, respectively. Power deposition was stopped when thermal dose of 40 cumulative equivalent minutes at 43 °C (CEM43) was accrued in adjacent myometrium.Results:Within the range of all evaluated insertion angles, depths and offsets, the ablation coverage was less sensitive to variation in angle as compared to depth and offset, and ranged from 34.9 - 83.6% for 3 cm fibroid in 140 - 400 s and 34.1 - 67.9% for 1 cm fibroid in 30 - 50 s of heating duration. Maximum achievable ablation coverage in both fibroid cases reach ∼ 90% if thermal dose is allowed to exceed 40 CEM43 in myometrium.Conclusion:The study demonstrates the technical feasibility of transcervical microwave ablation for fibroid treatment and the relationship between applicator position within the fibroid and fraction of fibroid that can be ablated while limiting thermal dose in adjacent myometrium.
Department of Circuit Theory Czech Technical University Prague Prague Czech Republic
Hologic Inc Marlborough Massachusetts United States of America
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc22012413
- 003
- CZ-PrNML
- 005
- 20220506131533.0
- 007
- ta
- 008
- 220425s2021 xxk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1088/2057-1976/abffe4 $2 doi
- 035 __
- $a (PubMed)33975302
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxk
- 100 1_
- $a Zia, Ghina $u Department of Electrical and Computer Engineering, Kansas State University, Manhattan, Kansas, United States of America $1 https://orcid.org/0000000302641533
- 245 10
- $a Transcervical microwave ablation in type 2 uterine fibroids via a hysteroscopic approach: analysis of ablation profiles / $c G. Zia, J. Sebek, J. Schenck, P. Prakash
- 520 9_
- $a Type 2 uterine fibroids are challenging to resect surgically as ≥ 50% volume of myoma lies within the myometrium. A hysteroscopic approach for ablating fibroids is minimally-invasive, but places a considerable burden on the operator to accurately place the ablation applicator within the target. We investigated the sensitivity of transcervical microwave ablation outcome with respect to position of the ablation applicator within 1 - 3 cm type 2 fibroids.Methods:A finite element computer model was developed to simulate 5.8 GHz microwave ablation of fibroids and validated with experiments inex vivotissue. The ablation outcome was evaluated with respect to applicator insertion angles (30°, 45°, 60°) , depth and offset from the fibroid center (±2 mm for 3 cm fibroid and ±1 mm for 1 cm fibroid) with 35 W and 15 W applied power for 3 cm and 1 cm fibroids, respectively. Power deposition was stopped when thermal dose of 40 cumulative equivalent minutes at 43 °C (CEM43) was accrued in adjacent myometrium.Results:Within the range of all evaluated insertion angles, depths and offsets, the ablation coverage was less sensitive to variation in angle as compared to depth and offset, and ranged from 34.9 - 83.6% for 3 cm fibroid in 140 - 400 s and 34.1 - 67.9% for 1 cm fibroid in 30 - 50 s of heating duration. Maximum achievable ablation coverage in both fibroid cases reach ∼ 90% if thermal dose is allowed to exceed 40 CEM43 in myometrium.Conclusion:The study demonstrates the technical feasibility of transcervical microwave ablation for fibroid treatment and the relationship between applicator position within the fibroid and fraction of fibroid that can be ablated while limiting thermal dose in adjacent myometrium.
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a leiomyom $x chirurgie $7 D007889
- 650 12
- $a mikrovlny $7 D008872
- 650 _2
- $a myometrium $7 D009215
- 650 _2
- $a radiofrekvenční ablace $7 D000078703
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Sebek, Jan $u Department of Electrical and Computer Engineering, Kansas State University, Manhattan, Kansas, United States of America $u Department of Circuit Theory, Czech Technical University in Prague, Prague, Czech Republic $1 https://orcid.org/0000000221166697 $7 xx0214491
- 700 1_
- $a Schenck, Jessica $u Hologic, Inc., Marlborough, Massachusetts, United States of America
- 700 1_
- $a Prakash, Punit $u Department of Electrical and Computer Engineering, Kansas State University, Manhattan, Kansas, United States of America
- 773 0_
- $w MED00208693 $t Biomedical physics & engineering express $x 2057-1976 $g Roč. 7, č. 4 (2021)
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/33975302 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y p $z 0
- 990 __
- $a 20220425 $b ABA008
- 991 __
- $a 20220506131525 $b ABA008
- 999 __
- $a ok $b bmc $g 1789839 $s 1163614
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2021 $b 7 $c 4 $e 20210601 $i 2057-1976 $m Biomedical physics & engineering express $n Biomed Phys Eng Express $x MED00208693
- LZP __
- $a Pubmed-20220425