Comparison of toric intraocular lens tilt and decentration measurement using dynamic Purkinje-meter and anterior segment optical coherence tomography
E. Palkovicova, J. Cendelin, J. Novak
Jazyk angličtina Země Česko
Typ dokumentu klinické zkoušky, srovnávací studie, časopisecké články
NLK
Directory of Open Access Journals
od 2001
Free Medical Journals
od 1998
Medline Complete (EBSCOhost)
od 2007-06-01
ROAD: Directory of Open Access Scholarly Resources
od 2001
PubMed
38868995
DOI
10.5507/bp.2024.017
Knihovny.cz E-zdroje
- MeSH
- algoritmy MeSH
- lidé středního věku MeSH
- lidé MeSH
- nitrooční čočky * MeSH
- optická koherentní tomografie MeSH
- optické zobrazování * přístrojové vybavení metody MeSH
- senioři MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- klinické zkoušky MeSH
- srovnávací studie MeSH
AIMS: To present a new method of dynamic Purkinje-metry and to verify it by comparison with a commercially available anterior segment optical coherence tomography CASIA2. PATIENTS AND METHODS: A dynamic Purkinje-meter with a movable fixation target was assembled. A coaxial circular pattern formed by infrared LEDs was projected onto the eye and evoked Purkinje images (1st, 3rd, 4th = P1, P3, P4). The measurement was performed on 29 eyes with an implanted toric IOL (intraocular lens), under mydriatic conditions, with reference to the visual axis. The IOL tilt was calculated from the position of a fixation target at the moment of P3 and P4 superposition. The IOL decentration was determined based on the relative position of P1 during on-axis fixation and of P3 and P4 superposition during off-axis fixation. A custom-developed software was used for distance measurements. Using CASIA2, the IOL position was fully calculated by the device. RESULTS: The mean absolute difference between CASIA2 and Purkinje-meter values was 0.6° ± 0.4° for the tilt magnitude and 10° ± 10° for the tilt direction, and 0.11 mm ± 0.08 mm for the decentration magnitude and 16° ± 14° for the decentration direction. There was no statistically significant difference between the values determined by the two methods for the tilt and decentration direction. The differences were statistically significant for the tilt and decentration magnitude. CONCLUSION: The values of IOL tilt and decentration direction are similar for both devices. The values of IOL tilt and decentration magnitude measured by Purkinje-meter are higher than those from CASIA2, but overall, they correspond to the values presented in other published studies.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25008980
- 003
- CZ-PrNML
- 005
- 20250521143020.0
- 007
- ta
- 008
- 250411s2025 xr ad f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.5507/bp.2024.017 $2 doi
- 035 __
- $a (PubMed)38868995
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xr
- 100 1_
- $a Palkovičová, Eliška $u Department of Natural Sciences, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic $7 _AN097408
- 245 10
- $a Comparison of toric intraocular lens tilt and decentration measurement using dynamic Purkinje-meter and anterior segment optical coherence tomography / $c E. Palkovicova, J. Cendelin, J. Novak
- 520 9_
- $a AIMS: To present a new method of dynamic Purkinje-metry and to verify it by comparison with a commercially available anterior segment optical coherence tomography CASIA2. PATIENTS AND METHODS: A dynamic Purkinje-meter with a movable fixation target was assembled. A coaxial circular pattern formed by infrared LEDs was projected onto the eye and evoked Purkinje images (1st, 3rd, 4th = P1, P3, P4). The measurement was performed on 29 eyes with an implanted toric IOL (intraocular lens), under mydriatic conditions, with reference to the visual axis. The IOL tilt was calculated from the position of a fixation target at the moment of P3 and P4 superposition. The IOL decentration was determined based on the relative position of P1 during on-axis fixation and of P3 and P4 superposition during off-axis fixation. A custom-developed software was used for distance measurements. Using CASIA2, the IOL position was fully calculated by the device. RESULTS: The mean absolute difference between CASIA2 and Purkinje-meter values was 0.6° ± 0.4° for the tilt magnitude and 10° ± 10° for the tilt direction, and 0.11 mm ± 0.08 mm for the decentration magnitude and 16° ± 14° for the decentration direction. There was no statistically significant difference between the values determined by the two methods for the tilt and decentration direction. The differences were statistically significant for the tilt and decentration magnitude. CONCLUSION: The values of IOL tilt and decentration direction are similar for both devices. The values of IOL tilt and decentration magnitude measured by Purkinje-meter are higher than those from CASIA2, but overall, they correspond to the values presented in other published studies.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a lidé středního věku $7 D008875
- 650 _2
- $a senioři $7 D000368
- 650 12
- $a nitrooční čočky $7 D007910
- 650 _2
- $a optická koherentní tomografie $7 D041623
- 650 12
- $a optické zobrazování $x přístrojové vybavení $x metody $7 D061848
- 650 _2
- $a algoritmy $7 D000465
- 655 _2
- $a klinické zkoušky $7 D016430
- 655 _2
- $a srovnávací studie $7 D003160
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a srovnávací studie $7 D003160
- 700 1_
- $a Cendelín, Jiří $u Department of Natural Sciences, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic $u Department of Ophthalmology, Second Faculty of Medicine, Charles University in Prague and Motol University Hospital, Prague, Czech Republic $7 xx0073752
- 700 1_
- $a Novák, Jiří $u Department of Natural Sciences, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic $u Department of Physics, Faculty of Civil Engineering, Czech Technical University in Prague, Prague, Czech Republic $7 _AN060799
- 773 0_
- $w MED00012606 $t Biomedical papers of the Medical Faculty of the University Palacky, Olomouc, Czechoslovakia $x 1804-7521 $g Roč. 169, č. 1 (2025), s. 56-65
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/38868995 $y Pubmed
- 910 __
- $a ABA008 $b A 1502 $c 958 $y p $z 0
- 990 __
- $a 20250411 $b ABA008
- 991 __
- $a 20250521143017 $b ABA008
- 999 __
- $a ok $b bmc $g 2324856 $s 1246058
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2025 $b 169 $c 1 $d 56-65 $e 20240611 $i 1804-7521 $m Biomedical papers of the Medical Faculty of the University Palacky, Olomouc, Czechoslovakia $n Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub $x MED00012606
- LZP __
- $b NLK124 $a Pubmed-20250411