Precision of post-operative localization of deep brain stimulation electrodes
A. Lasica, P. Filip, K. Burdová, J. Mana, F. Růžička, D. Urgošík, K. Mueller, D. Kiakou, R. Jech
Language English Country England, Great Britain
Document type Journal Article
Grant support
LX22NPO5107
Ministerstvo Školství, Mládeže a Tělovýchovy
CZ-DRO-VFN64165
Ministerstvo Zdravotnictví Ceské Republiky
NLK
Directory of Open Access Journals
from 2011
Free Medical Journals
from 2011
Nature Open Access
from 2011-12-01
PubMed Central
from 2011
Europe PubMed Central
from 2011
ProQuest Central
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Open Access Digital Library
from 2011-01-01
Health & Medicine (ProQuest)
from 2011-01-01
ROAD: Directory of Open Access Scholarly Resources
from 2011
Springer Nature OA/Free Journals
from 2011-12-01
- MeSH
- Deep Brain Stimulation * methods instrumentation MeSH
- Electrodes, Implanted * MeSH
- Middle Aged MeSH
- Humans MeSH
- Magnetic Resonance Imaging MeSH
- Subthalamic Nucleus surgery MeSH
- Parkinson Disease * therapy MeSH
- Aged MeSH
- Software MeSH
- Check Tag
- Middle Aged MeSH
- Humans MeSH
- Male MeSH
- Aged MeSH
- Female MeSH
- Publication type
- Journal Article MeSH
Tools for post-operative localization of deep brain stimulation (DBS) electrodes may be of major benefit in the evaluation of the stimulation area. However, little is known about their precision. This study compares 3 different software packages used for DBS electrode localization. T1-weighted MRI images before and after the implantation of the electrodes into the subthalamic nucleus for DBS in 105 Parkinson's disease patients were processed using the pipelines implemented in Lead-DBS, SureTune4, and Brainlab. Euclidean distance between active contacts determined by individual software packages and in repeated processing by the same and by a different operator was calculated. Furthermore, Dice coefficient for overlap of volume of tissue activated (VTA) was determined for Lead-DBS. Medians of Euclidean distances between estimated active contact locations in inter-software package comparison ranged between 1.5 mm and 2 mm. Euclidean distances in within-software package intra- and inter-rater assessments were 0.6-1 mm and 1-1.7 mm, respectively. Median intra- and inter-rater Dice coefficients for VTAs were 0.78 and 0.75, respectively. Since the median distances are close to the size of the target nucleus, any clinical use should be preceded by careful review of the outputs.
Center for Magnetic Resonance Research University of Minnesota Minneapolis MN USA
Department of Radiology Na Homolce Hospital Prague Czech Republic
Max Planck Institute for Human Cognitive and Brain Sciences Leipzig Germany
References provided by Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc25015633
- 003
- CZ-PrNML
- 005
- 20250731091133.0
- 007
- ta
- 008
- 250708s2025 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1038/s41598-025-01449-6 $2 doi
- 035 __
- $a (PubMed)40436963
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Lasica, Andrej $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic
- 245 10
- $a Precision of post-operative localization of deep brain stimulation electrodes / $c A. Lasica, P. Filip, K. Burdová, J. Mana, F. Růžička, D. Urgošík, K. Mueller, D. Kiakou, R. Jech
- 520 9_
- $a Tools for post-operative localization of deep brain stimulation (DBS) electrodes may be of major benefit in the evaluation of the stimulation area. However, little is known about their precision. This study compares 3 different software packages used for DBS electrode localization. T1-weighted MRI images before and after the implantation of the electrodes into the subthalamic nucleus for DBS in 105 Parkinson's disease patients were processed using the pipelines implemented in Lead-DBS, SureTune4, and Brainlab. Euclidean distance between active contacts determined by individual software packages and in repeated processing by the same and by a different operator was calculated. Furthermore, Dice coefficient for overlap of volume of tissue activated (VTA) was determined for Lead-DBS. Medians of Euclidean distances between estimated active contact locations in inter-software package comparison ranged between 1.5 mm and 2 mm. Euclidean distances in within-software package intra- and inter-rater assessments were 0.6-1 mm and 1-1.7 mm, respectively. Median intra- and inter-rater Dice coefficients for VTAs were 0.78 and 0.75, respectively. Since the median distances are close to the size of the target nucleus, any clinical use should be preceded by careful review of the outputs.
- 650 _2
- $a lidé $7 D006801
- 650 12
- $a hluboká mozková stimulace $x metody $x přístrojové vybavení $7 D046690
- 650 12
- $a Parkinsonova nemoc $x terapie $7 D010300
- 650 _2
- $a magnetická rezonanční tomografie $7 D008279
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 12
- $a implantované elektrody $7 D004567
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a nucleus subthalamicus $x chirurgie $7 D020531
- 650 _2
- $a lidé středního věku $7 D008875
- 650 _2
- $a senioři $7 D000368
- 650 _2
- $a software $7 D012984
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Filip, Pavel $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic $u Center for Magnetic Resonance Research (CMRR), University of Minnesota, Minneapolis, MN, USA
- 700 1_
- $a Burdová, Kristína $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic
- 700 1_
- $a Mana, Josef $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic
- 700 1_
- $a Růžička, Filip $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic
- 700 1_
- $a Urgošík, Dušan $u Department of Radiology, Na Homolce Hospital, Prague, Czech Republic
- 700 1_
- $a Mueller, Karsten $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic $u Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
- 700 1_
- $a Kiakou, Dimitra $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic $u Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
- 700 1_
- $a Jech, Robert $u Department of Neurology, First Faculty of Medicine, General University Hospital in Prague, Charles University, Kateřinská 30, 120 00, Prague, Czech Republic. jech@cesnet.cz
- 773 0_
- $w MED00182195 $t Scientific reports $x 2045-2322 $g Roč. 15, č. 1 (2025), s. 18652
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/40436963 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20250708 $b ABA008
- 991 __
- $a 20250731091128 $b ABA008
- 999 __
- $a ok $b bmc $g 2366471 $s 1252758
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2025 $b 15 $c 1 $d 18652 $e 20250528 $i 2045-2322 $m Scientific reports $n Sci Rep $x MED00182195
- GRA __
- $a LX22NPO5107 $p Ministerstvo Školství, Mládeže a Tělovýchovy
- GRA __
- $a CZ-DRO-VFN64165 $p Ministerstvo Zdravotnictví Ceské Republiky
- LZP __
- $a Pubmed-20250708