Quantitative selected ion flow tube mass spectrometry: the influence of ionic diffusion and mass discrimination

. 2001 Jul ; 12 (7) : 863-72.

Status PubMed-not-MEDLINE Jazyk angličtina Země Spojené státy americké Médium print

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid11444610
Odkazy

PubMed 11444610
DOI 10.1016/s1044-0305(01)00253-7
PII: S1044-0305(01)00253-7
Knihovny.cz E-zdroje

Selected ion flow tube mass spectrometry, (SIFT-MS), involves the partial conversion of mass-selected precursor ions to product ions in their reactions with the trace gases in an air sample that is introduced into helium carrier gas in a flow tube. The precursor and product ions are then detected and counted by a downstream quadrupole mass spectrometer. Quantification of particular trace gases is thus achieved from the ratio of the total count rate of the product ions to that for the precursor ions. However, it is important to appreciate that in this ion chemistry the light precursor ions (usually H3O+ ions) are invariably converted to heavier product ions. Hence, the product ions diffuse to the flow tube walls more slowly and thus they are more efficiently transported to the downstream mass spectrometer sampling orifice. This phenomenon we refer to as diffusion enhancement. Further, it is a well-known fact that discrimination can occur against ions of large mass-to-charge ratio, (m/z), in quadrupole mass spectrometers. If not accounted for, diffusion enhancement usually results in erroneously high trace gas concentrations and mass discrimination results in erroneously low concentrations. In this experimental investigation, we show how both these counteracting effects can be accounted for to increase the accuracy of SIFT-MS quantification. This is achieved by relating the currents of ions of various m/z that arrive at the downstream mass spectrometer sampling orifice disc to their count rates at the ion detector after mass analysis. Thus, both diffusion enhancement and mass discrimination are parameterized as a function of m/z and these are combined to provide an overall discrimination factor for the particular analytical instrument.

Zobrazit více v PubMed

Rapid Commun Mass Spectrom. 1999;13(17):1733-8 PubMed

Rapid Commun Mass Spectrom. 1999;13(14):1354-9 PubMed

Clin Chem. 1983 Jan;29(1):5-15 PubMed

Med Biol Eng Comput. 1996 Nov;34(6):409-19 PubMed

Rapid Commun Mass Spectrom. 2000;14(13):1136-40 PubMed

Rapid Commun Mass Spectrom. 2000;14(20):1898-906 PubMed

J Appl Physiol (1985). 1999 Nov;87(5):1584-8 PubMed

Kidney Int. 1997 Jul;52(1):223-8 PubMed

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...