Measurement of stray radiation within a scanning proton therapy facility: EURADOS WG9 intercomparison exercise of active dosimetry systems
Jazyk angličtina Země Spojené státy americké Médium print
Typ dokumentu časopisecké články, práce podpořená grantem
PubMed
25979049
DOI
10.1118/1.4916667
Knihovny.cz E-zdroje
- MeSH
- dávka záření MeSH
- fantomy radiodiagnostické MeSH
- fotony MeSH
- neutrony MeSH
- protonová terapie přístrojové vybavení metody MeSH
- protony MeSH
- radiometrie přístrojové vybavení metody MeSH
- spektrální analýza přístrojové vybavení metody MeSH
- voda MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- Geografické názvy
- Evropa MeSH
- Názvy látek
- protony MeSH
- voda MeSH
PURPOSE: To characterize stray radiation around the target volume in scanning proton therapy and study the performance of active neutron monitors. METHODS: Working Group 9 of the European Radiation Dosimetry Group (EURADOS WG9-Radiation protection in medicine) carried out a large measurement campaign at the Trento Centro di Protonterapia (Trento, Italy) in order to determine the neutron spectra near the patient using two extended-range Bonner sphere spectrometry (BSS) systems. In addition, the work focused on acknowledging the performance of different commercial active dosimetry systems when measuring neutron ambient dose equivalents, H(∗)(10), at several positions inside (8 positions) and outside (3 positions) the treatment room. Detectors included three TEPCs--tissue equivalent proportional counters (Hawk type from Far West Technology, Inc.) and six rem-counters (WENDI-II, LB 6411, RadEye™ NL, a regular and an extended-range NM2B). Meanwhile, the photon component of stray radiation was deduced from the low-lineal energy transfer part of TEPC spectra or measured using a Thermo Scientific™ FH-40G survey meter. Experiments involved a water tank phantom (60 × 30 × 30 cm(3)) representing the patient that was uniformly irradiated using a 3 mm spot diameter proton pencil beam with 10 cm modulation width, 19.95 cm distal beam range, and 10 × 10 cm(2) field size. RESULTS: Neutron spectrometry around the target volume showed two main components at the thermal and fast energy ranges. The study also revealed the large dependence of the energy distribution of neutrons, and consequently of out-of-field doses, on the primary beam direction (directional emission of intranuclear cascade neutrons) and energy (spectral composition of secondary neutrons). In addition, neutron mapping within the facility was conducted and showed the highest H(∗)(10) value of ∼ 51 μSv Gy(-1); this was measured at 1.15 m along the beam axis. H(∗)(10) values significantly decreased with distance and angular position with respect to beam axis falling below 2 nSv Gy(-1) at the entrance of the maze, at the door outside the room and below detection limit in the gantry control room, and at an adjacent room (<0.1 nSv Gy(-1)). Finally, the agreement on H(∗)(10) values between all detectors showed a direct dependence on neutron spectra at the measurement position. While conventional rem-counters (LB 6411, RadEye™ NL, NM2-458) underestimated the H(∗)(10) by up to a factor of 4, Hawk TEPCs and the WENDI-II range-extended detector were found to have good performance (within 20%) even at the highest neutron fluence and energy range. Meanwhile, secondary photon dose equivalents were found to be up to five times lower than neutrons; remaining nonetheless of concern to the patient. CONCLUSIONS: Extended-range BSS, TEPCs, and the WENDI-II enable accurate measurements of stray neutrons while other rem-counters are not appropriate considering the high-energy range of neutrons involved in proton therapy.
Departament de Física Universitat Autònoma de Barcelona Bellaterra E 08193 Spain
Department of Radiation Dosimetry Nuclear Physics Institute Řež CZ 250 68 Czech Republic
Institute of Nuclear Physics PAN Radzikowskiego 152 Krakow 31 342 Poland
Ruđer Bošković Institute Bijenička c 54 Zagreb 10000 Croatia
University of Newcastle upon Tyne Newcastle upon Tyne Tyne and Wear NE1 7RU United Kingdom
Citace poskytuje Crossref.org
In Vitro Comparison of Passive and Active Clinical Proton Beams