Design and Development of a Novel Invasive Blood Pressure Simulator for Patient's Monitor Testing

. 2020 Jan 01 ; 20 (1) : . [epub] 20200101

Jazyk angličtina Země Švýcarsko Médium electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid31906383

Grantová podpora
SV450994/2101 VSB Fonds
TL01000302 Technology Agency of the Czech Republic

This paper presents a newly-designed and realized Invasive Blood Pressure (IBP) device for the simulation on patient's monitors. This device shows improvements and presents extended features with respect to a first prototype presented by the authors and similar systems available in the state-of-the-art. A peculiarity of the presented device is that all implemented features can be customized from the developer and from the point of view of the end user. The realized device has been tested, and its performances in terms of accuracy and of the back-loop measurement of the output for the blood pressure regulation utilization have been described. In particular, an accuracy of ±1 mmHg at 25 °C, on a range from -30 to 300 mmHg, was evaluated under different test conditions. The designed device is an ideal tool for testing IBP modules, for zero setting, and for calibrations. The implemented extended features, like the generation of custom waveforms and the Universal Serial Bus (USB) connectivity, allow use of this device in a wide range of applications, from research to equipment maintenance in clinical environments to educational purposes. Moreover, the presented device represents an innovation, both in terms of technology and methodologies: It allows quick and efficient tests to verify the proper functioning of IBP module of patients' monitors. With this innovative device, tests can be performed directly in the field and faster procedures can be implemented by the clinical maintenance personnel. This device is an open source project and all materials, hardware, and software are fully available for interested developers or researchers.

Zobrazit více v PubMed

Eason M.P. Simulation devices in cardiothoracic and vascular anesthesia. Semin. Cardiothorac. Vasc. Anesth. 2005;9:309–323. doi: 10.1177/108925320500900404. PubMed DOI

Green M., Tariq R., Green P. Improving patient safety through simulation training in anesthesiology: Where are we? Anesthesiol. Res. Pract. 2016;2016:12. doi: 10.1155/2016/4237523. PubMed DOI PMC

Cerny M., Penhaker M., Gala M., Babusiak B. Biomedical engineering education under european union support. In: AbuOsman N.A., Abas W.A.W., AbdulWahab A.K., Ting H.N., editors. 5th Kuala Lumpur International Conference on Biomedical Engineering. Vol. 35. Springer; Berlin/Heidelberg, Germany: 2011. pp. 16–20.

Johnson S.M., Owens T.L., O’Neil J.N. Making the clinical connection from textbook to bedside during MDY1: An integrative approach for medical physiology education employing human simulation. Adv. Physiol. Educ. 2019;43:128–133. doi: 10.1152/advan.00109.2018. PubMed DOI

Fox F. Method and Apparatus for Zeroing and Calibrating an Invasive Blood Pressure Monitoring System. No. 4,342,218. U.S. Patent. 1982 Aug 3;

Kim C.H., SaGong G., Nam K., Jeon R. Blood pressure simulator using hybrid controller. J. Sens. Sci. Technol. 2007;16:44–51.

Krejcar O., Slanina Z., Stambachr J., Silber P., Frischer R. Noninvasive continuous blood pressure measurement and GPS position monitoring of patients; Proceedings of the 2009 IEEE 70th Vehicular Technology Conference Fall; Anchorage, AK, USA. 20–23 September 2009; pp. 3–7. DOI

Kijonka J., Penhaker M. Electronic invasive blood pressure simulator device for patient monitor testing. Elektron. Elektrotechnika. 2012;122:49–54. doi: 10.5755/j01.eee.122.6.1820. DOI

Penhaker M., Kijonka J. Electrical Power Systems and Computers. Springer; Berlin/Heidelberg, Germany: 2011. Invasive blood pressure simulator electronics device bed side monitor testing; pp. 823–830.

Kijonka J., Penhaker M., Cernohorsky J. 2011 1st Middle East Conference on Biomedical Engineering. IEEE; Piscataway, NJ, USA: 2011. Invasive blood pressure curves simulation device.

Kijonka J., Penhaker M. XII Mediterranean Conference on Medical and Biological Engineering and Computing. Springer; Berlin/Heidelberg, Germany: 2010. Embedded programmable invasive blood pressure simulator.

Kijonka J., Penhaker M. Master’s Thesis. Department of Cybernetics and Biomedical Engineering, Faculty of Electrical Engineering and Computer Science, VSB-Technical University of Ostrava; Ostrava, Czech Republic: 2010. Invasive Blood Pressure Simulator, Testing, Adjusting and Calibration of the Device.

Fida B., Bernabucci I., Bibbo D., Conforto S., Schmid M. Pre-processing effect on the accuracy of event-based activity segmentation and classification through inertial sensors. Sensors. 2015;15:23095–23109. doi: 10.3390/s150923095. PubMed DOI PMC

NovaSensor Disposable Medical Pressure Sensor NPC-100 and NPC-120 Series. [(accessed on 31 December 2019)]; Available online: http://www.amphenol-sensors.com/en/products/pressure-mems/mems-sensors/3150-npc-100-and-120-series#features.

Garani G., Adam G.K. Qualitative modeling at the design of concrete manufacturing machinery. Int. J. Comput. Appl. 2008;30:325–330.

Proto A., Bibbo D., Conforto S., Schmid M. A new microcontroller-based system to optimize the digital conversion of signals originating from load cells built-in into pedals; Proceedings of the IEEE 2014 Biomedical Circuits and Systems Conference (BioCAS); Lausanne, Switzerland. 22–24 October 2014; pp. 300–303.

Babusiak B., Borik S., Gala M. Bio-amplifier with programmable gain and adjustable leads for basic measurement of bioelectric signals. In: Pietka E., Kawa J., Wieclawek W., editors. Information Technologies in Biomedicine. Vol. 4. Springer; Cham, Germany: 2014. pp. 249–260.

Babusiak B., Borik S. Bluetooth communication for battery powered medical devices. J. Electr. Eng. Elektrotech. Cas. 2016;67:65–68. doi: 10.1515/jee-2016-0010. DOI

Babusiak B., Borik S. Low energy wireless communication for medical devices; Proceedings of the 38th International Conference on Telecommunications and Signal Processing (TSP); Prague, Czech. 9–11 July 2015; pp. 444–447.