To select a Kjeldahl procedure suitable for the determination of total protein in reference materials used in laboratory medicine, we reviewed in our previous article Kjeldahl methods adopted by clinical chemistry and found an indirect two-step analysis by total Kjeldahl nitrogen corrected for its nonprotein nitrogen and a direct analysis made on isolated protein precipitates. In this article, we compare both procedures on various reference materials. An indirect Kjeldahl method gave falsely lower results than a direct analysis. Preliminary performance parameters qualify the direct Kjeldahl analysis as a suitable primary reference procedure for the certification of total protein in reference laboratories.
- MeSH
- klinická chemie přístrojové vybavení metody MeSH
- lidé MeSH
- proteiny chemie MeSH
- referenční hodnoty MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
We found previously that albumin-calibrated total protein in certified reference materials causes unacceptable positive bias in analysis of human sera. The simplest way to cure this defect is the use of human-based serum/plasma standards calibrated by the Kjeldahl method. Such standards, commutative with serum samples, will compensate for bias caused by lipids and bilirubin in most human sera. To find a suitable primary reference procedure for total protein in reference materials, we reviewed Kjeldahl methods adopted by laboratory medicine. We found two methods recommended for total protein in human samples: an indirect analysis based on total Kjeldahl nitrogen corrected for its nonprotein nitrogen and a direct analysis made on isolated protein precipitates. The methods found will be assessed in a subsequent article.