Nejvíce citovaný článek - PubMed ID 32857492
Understanding Gas Phase Ion Chemistry Is the Key to Reliable Selected Ion Flow Tube-Mass Spectrometry Analyses
RATIONALE: Data are required for SIFT-MS analysis of perfluoroalkyl and polyfluoroalkyl substances (PFAS), which are persistent in the environment and cause adverse health effects. Specifically, the rate coefficients and product ion branching ratios of the reactions of H3O+, NO+, O2 +•, O-•, OH-, O2 -•, NO2 - and NO3 - with PFAS vapours are needed. METHODS: The dual polarity SIFT-MS instrument (Voice200) was used to generate these eight reagent ions and inject them into the flow tube with N2 carrier gas at a temperature of 393 K. Vapours of pentafluoropropionic acid, heptafluorobutyric acid, nonafluoro-1-hexanol, perfluoro-2-methyl-2-pentene, perfluorohexanoic acid, perfluoro(2-methyl-3-oxahexanoic) acid, tridecafluoro-1-octanol and nonafluorobutane-1-sulfonic acid were introduced in dry and humid air. Full-scan mass spectra were collected for all reagents at variable PFAS concentrations and analysed numerically. RESULTS: Rate coefficients were determined for 64 reactions, for which 55 positive and 71 negative product ions were identified. The branching ratios for the primary reaction channels were extracted from the data, and the secondary chemistry with H2O molecules was qualitatively assessed. The thermochemical data were calculated for the H3O+ reactions using density functional theory (DFT). CONCLUSIONS: An important observation is that secondary reactions with water molecules remove the positive product ions, making them unsuitable for practical SIFT-MS analysis of PFAS vapours. In contrast, most negative reaction product ions are not significantly affected by humidity and are thus preferred for the SIFT-MS analyses of PFAS substances in various gaseous matrices.
- Klíčová slova
- ion molecule reactions, perfluoroalkyl and polyfluoroalkyl substances (PFAS), selected ion flow tube,
- Publikační typ
- časopisecké články MeSH
Selected ion flow tube mass spectrometry (SIFT-MS) is now recognized as the most versatile analytical technique for the identification and quantification of trace gases down to the parts-per-trillion by volume, pptv, range. This statement is supported by the wide reach of its applications, from real-time analysis, obviating sample collection of very humid exhaled breath, to its adoption in industrial scenarios for air quality monitoring. This review touches on the recent extensions to the underpinning ion chemistry kinetics library and the alternative challenge of using nitrogen carrier gas instead of helium. The addition of reagent anions in the Voice200 series of SIFT-MS instruments has enhanced the analytical capability, thus allowing analyses of volatile trace compounds in humid air that cannot be analyzed using reagent cations alone, as clarified by outlining the anion chemistry involved. Case studies are reviewed of breath analysis and bacterial culture volatile organic compound (VOC), emissions, environmental applications such as air, water, and soil analysis, workplace safety such as transport container fumigants, airborne contamination in semiconductor fabrication, food flavor and spoilage, drugs contamination and VOC emissions from packaging to demonstrate the stated qualities and uniqueness of the new generation SIFT-MS instrumentation. Finally, some advancements that can be made to improve the analytical capability and reach of SIFT-MS are mentioned.
- Klíčová slova
- SIFT‐MS, VOCs, cation and anion gas phase chemistry, nitrogen carrier gas, selected ion flow tube mass spectrometry, volatile organic compounds,
- Publikační typ
- časopisecké články MeSH
- přehledy MeSH
Selected ion flow tube mass spectrometry, SIFT-MS, has been widely used in industry and research since its introduction in the mid-1990s. Previously described quantitation methods have been advanced to include a gas standard for a more robust and repeatable analytical performance. The details of this approach to calculate the concentrations from ion-molecule reaction kinetics based on reaction times and instrument calibration functions determined from known concentrations in the standard mix are discussed. Important practical issues such as the overlap of product ions are outlined, and best-practice approaches are presented to enable them to be addressed during method development. This review provides a fundamental basis for a plethora of studies in broad application areas that are possible with SIFT-MS instruments.
- Klíčová slova
- SIFT-MS, VOCs, ion molecule reactions, negative reagent ions, nitrogen carrier gas, quantitation, selected ion flow tube mass spectrometry, volatile organic compounds,
- Publikační typ
- časopisecké články MeSH
Selected ion flow tube mass spectrometry (SIFT-MS) instruments have significantly developed since this technique was introduced more than 20 years ago. Most studies of the ion-molecule reaction kinetics that are essential for accurate analyses of trace gases and vapors in air and breath were conducted in He carrier gas at 300 K, while the new SIFT-MS instruments (optimized to quantify concentrations down to parts per trillion by volume) operate with N2 carrier gas at 393 K. Thus, we pose the question of how to reuse the data from the extensive body of previous literature using He at room temperature in the new instruments operating with N2 carrier gas at elevated temperatures. Experimentally, we found the product ions to be qualitatively similar, although there were differences in the branching ratios, and some reaction rate coefficients were lower in the heated N2 carrier gas. The differences in the reaction kinetics may be attributed to temperature, an electric field in the current flow tubes, and the change from He to N2 carrier gas. These results highlight the importance of adopting an updated reaction kinetics library that accounts for the new instruments' specific conditions. In conclusion, almost all previous rate coefficients may be used after adjustment for higher temperatures, while some product ion branching ratios need to be updated.
- Publikační typ
- časopisecké články MeSH
The analysis of volatile organic compounds (VOCs) within breath for noninvasive disease detection and monitoring is an emergent research field that has the potential to reshape current clinical practice. However, adoption of breath testing has been limited by a lack of standardization. This protocol provides a comprehensive workflow for online and offline breath analysis using selected ion flow tube mass spectrometry (SIFT-MS). Following the suggested protocol, 50 human breath samples can be analyzed and interpreted in <3 h. Key advantages of SIFT-MS are exploited, including the acquisition of real-time results and direct compound quantification without need for calibration curves. The protocol includes details of methods developed for targeted analysis of disease-specific VOCs, specifically short-chain fatty acids, aldehydes, phenols, alcohols and alkanes. A procedure to make custom breath collection bags is also described. This standardized protocol for VOC analysis using SIFT-MS is intended to provide a basis for wider application and the use of breath analysis in clinical studies.
- MeSH
- dechové testy metody MeSH
- dospělí MeSH
- hmotnostní spektrometrie metody MeSH
- ionty MeSH
- lidé středního věku MeSH
- lidé MeSH
- mladiství MeSH
- mladý dospělý MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- těkavé organické sloučeniny analýza MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mladiství MeSH
- mladý dospělý MeSH
- senioři nad 80 let MeSH
- senioři MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- Názvy látek
- ionty MeSH
- těkavé organické sloučeniny MeSH