Activity standardisation of 45Ca and 204Tl using the new TDCR system at CMI
J. Sochorová, P. Auerbach, Z. Dutka,
Jazyk angličtina Země Anglie, Velká Británie
Typ dokumentu časopisecké články
- MeSH
- analýza selhání vybavení MeSH
- dávka záření MeSH
- design vybavení MeSH
- internacionalita MeSH
- poločas MeSH
- radioizotopy thallia analýza normy MeSH
- radioizotopy vápníku analýza normy MeSH
- radiometrie přístrojové vybavení normy MeSH
- referenční hodnoty MeSH
- referenční standardy MeSH
- Publikační typ
- časopisecké články MeSH
A new triple-to-double coincidence ratio (TDCR) system has been established at CMI as an additional technique for primary activity standardisation. Details of the detectors and the electronics are given. Construction of the optical chamber was validated using measurement of a (3)H standard-the efficiency achieved with Ultima Gold was 58%. Several methods of efficiency variation (coloured bands painted on vial, positioning and photomultipliers defocussing) are compared. Activity concentrations of (204)Tl and (45)Ca were determined with the TDCR method and with the efficiency tracer technique. Advantages of the TDCR method compared to the efficiency tracing method and agreement between results are shown.
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc13000891
- 003
- CZ-PrNML
- 005
- 20130109103019.0
- 007
- ta
- 008
- 130108s2012 enk f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1016/j.apradiso.2012.02.098 $2 doi
- 035 __
- $a (PubMed)22476016
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a enk
- 100 1_
- $a Sochorová, Jana $u Czech Metrology Institute, IIR, Radiová 1, 102 00 Praha 10, Czech Republic. jsochorova2@cmi.cz
- 245 10
- $a Activity standardisation of 45Ca and 204Tl using the new TDCR system at CMI / $c J. Sochorová, P. Auerbach, Z. Dutka,
- 520 9_
- $a A new triple-to-double coincidence ratio (TDCR) system has been established at CMI as an additional technique for primary activity standardisation. Details of the detectors and the electronics are given. Construction of the optical chamber was validated using measurement of a (3)H standard-the efficiency achieved with Ultima Gold was 58%. Several methods of efficiency variation (coloured bands painted on vial, positioning and photomultipliers defocussing) are compared. Activity concentrations of (204)Tl and (45)Ca were determined with the TDCR method and with the efficiency tracer technique. Advantages of the TDCR method compared to the efficiency tracing method and agreement between results are shown.
- 650 _2
- $a radioizotopy vápníku $x analýza $x normy $7 D002132
- 650 _2
- $a design vybavení $7 D004867
- 650 _2
- $a analýza selhání vybavení $7 D019544
- 650 _2
- $a poločas $7 D006207
- 650 _2
- $a internacionalita $7 D038622
- 650 _2
- $a dávka záření $7 D011829
- 650 _2
- $a radiometrie $x přístrojové vybavení $x normy $7 D011874
- 650 _2
- $a referenční standardy $7 D012015
- 650 _2
- $a referenční hodnoty $7 D012016
- 650 _2
- $a radioizotopy thallia $x analýza $x normy $7 D013794
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Auerbach, Pavel $7 gn_A_00009980
- 700 1_
- $a Dutka, Zdeněk
- 773 0_
- $w MED00000499 $t Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine $x 1872-9800 $g Roč. 70, č. 9 (2012), s. 2200-3
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/22476016 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20130108 $b ABA008
- 991 __
- $a 20130109103124 $b ABA008
- 999 __
- $a ok $b bmc $g 963673 $s 799055
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2012 $b 70 $c 9 $d 2200-3 $i 1872-9800 $m Applied radiation and isotopes $n Appl Radiat Isot $x MED00000499
- LZP __
- $a Pubmed-20130108