Safety and efficacy of catheter ablation on patients with persistent atrial fibrillation by targeting repetitive activation patterns and focal impulses
T. de Potter, A. Sarkozy, M. Duytschaever, A. Bulava
Jazyk angličtina Země Spojené státy americké
Typ dokumentu multicentrická studie, časopisecké články
NLK
CINAHL Plus with Full Text (EBSCOhost)
od 1978-01-01 do Před 1 rokem
Medline Complete (EBSCOhost)
od 1978-01-01 do Před 1 rokem
PubMed
37208935
DOI
10.1111/pace.14708
Knihovny.cz E-zdroje
- MeSH
- fibrilace síní * MeSH
- katetrizační ablace * metody MeSH
- lidé středního věku MeSH
- lidé MeSH
- prospektivní studie MeSH
- recidiva MeSH
- senioři MeSH
- supraventrikulární tachykardie * MeSH
- venae pulmonales * chirurgie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
BACKGROUND: The study is intended to evaluate the acute and long-term effectiveness and peri-procedural safety in ablation of persistent atrial fibrillation (PsAF) using the CartoFinder algorithm guided ablation (CFGA) targeting on repetitive activation patterns (RAPs) and focal impulses (FIs) identified in dynamic maps. METHODS: This is a prospective, single-arm, multicenter study. A 64-pole multielectrode basket catheter was used for intracardiac global electrogram (EGM) mapping. The RAPs or FIs were repeatedly mapped and ablated for up to five iterations by the CartoFinder algorithm to achieve sinus rhythm (SR) or organized atrial tachycardia (AT), which were followed by PVI. All patients were followed up for 12 months after procedure. RESULTS: Sixty-four PsAF patients (age, 60.7 ± 9.1 years; male, 76.6%; median PsAF duration, 6.0 months) underwent CFGA on RAPs/FIs. Six patients (9.4%) reported primary adverse event (PAE) including groin hematoma (2), complete heart block (1), tamponade (1), pericarditis (1), and pseudoaneurysm (1). Repeated mapping and ablation on RAPs/FIs resulted in the cycle length (CL) increase from 191.0 ± 167.6 ms at baseline to 365.7 ± 296.7 ms in the LA and from 167.8 ± 41.6 ms to 379.4 ± 293.5 ms in the RA and 30.2% (19/63) AF termination to SR or organized AT. The 12-month arrhythmia-free and symptomatic AF-free rates were 60.9% and 75.0%, respectively. Patients with acute AF termination showed a higher 12-month arrhythmia-free rate (76.9%) than those without (50.0%, p = .04). CONCLUSIONS: The study demonstrated that the CartoFinder algorithm can be used for global activation mapping during PsAF ablation. Patients with acute AF termination had a lower 12-month AF recurrence rate compared to patients without.
Cardiology Department AZ Sint Jan Bruges Belgium
Cardiology Department OLV Hospital Aalst Belgium
Cardiology Department Universitair Ziekenhuis Antwerpen Edegem Belgium
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc23016993
- 003
- CZ-PrNML
- 005
- 20231026105347.0
- 007
- ta
- 008
- 231013s2023 xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1111/pace.14708 $2 doi
- 035 __
- $a (PubMed)37208935
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a de Potter, Tom $u Cardiology Department, OLV Hospital, Aalst, Belgium
- 245 10
- $a Safety and efficacy of catheter ablation on patients with persistent atrial fibrillation by targeting repetitive activation patterns and focal impulses / $c T. de Potter, A. Sarkozy, M. Duytschaever, A. Bulava
- 520 9_
- $a BACKGROUND: The study is intended to evaluate the acute and long-term effectiveness and peri-procedural safety in ablation of persistent atrial fibrillation (PsAF) using the CartoFinder algorithm guided ablation (CFGA) targeting on repetitive activation patterns (RAPs) and focal impulses (FIs) identified in dynamic maps. METHODS: This is a prospective, single-arm, multicenter study. A 64-pole multielectrode basket catheter was used for intracardiac global electrogram (EGM) mapping. The RAPs or FIs were repeatedly mapped and ablated for up to five iterations by the CartoFinder algorithm to achieve sinus rhythm (SR) or organized atrial tachycardia (AT), which were followed by PVI. All patients were followed up for 12 months after procedure. RESULTS: Sixty-four PsAF patients (age, 60.7 ± 9.1 years; male, 76.6%; median PsAF duration, 6.0 months) underwent CFGA on RAPs/FIs. Six patients (9.4%) reported primary adverse event (PAE) including groin hematoma (2), complete heart block (1), tamponade (1), pericarditis (1), and pseudoaneurysm (1). Repeated mapping and ablation on RAPs/FIs resulted in the cycle length (CL) increase from 191.0 ± 167.6 ms at baseline to 365.7 ± 296.7 ms in the LA and from 167.8 ± 41.6 ms to 379.4 ± 293.5 ms in the RA and 30.2% (19/63) AF termination to SR or organized AT. The 12-month arrhythmia-free and symptomatic AF-free rates were 60.9% and 75.0%, respectively. Patients with acute AF termination showed a higher 12-month arrhythmia-free rate (76.9%) than those without (50.0%, p = .04). CONCLUSIONS: The study demonstrated that the CartoFinder algorithm can be used for global activation mapping during PsAF ablation. Patients with acute AF termination had a lower 12-month AF recurrence rate compared to patients without.
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a lidé středního věku $7 D008875
- 650 _2
- $a senioři $7 D000368
- 650 12
- $a fibrilace síní $7 D001281
- 650 _2
- $a prospektivní studie $7 D011446
- 650 _2
- $a výsledek terapie $7 D016896
- 650 _2
- $a recidiva $7 D012008
- 650 12
- $a supraventrikulární tachykardie $7 D013617
- 650 12
- $a katetrizační ablace $x metody $7 D017115
- 650 12
- $a venae pulmonales $x chirurgie $7 D011667
- 655 _2
- $a multicentrická studie $7 D016448
- 655 _2
- $a časopisecké články $7 D016428
- 700 1_
- $a Sarkozy, Andrea $u Cardiology Department, Universitair Ziekenhuis Antwerpen, Edegem, Belgium $u University of Antwerp, Antwerpen, Belgium
- 700 1_
- $a Duytschaever, Mattias $u Cardiology Department, AZ Sint Jan, Bruges, Belgium
- 700 1_
- $a Bulava, Alan $u Ceske Budejovice Hospital, Ceske Budejovice, Czech Republic $u Faculty of Health and Social Sciences, University of South Bohemia in Ceske Budejovice, Ceske Budejovice, Czech Republic
- 773 0_
- $w MED00003671 $t Pacing and clinical electrophysiology : PACE $x 1540-8159 $g Roč. 46, č. 7 (2023), s. 729-737
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/37208935 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y - $z 0
- 990 __
- $a 20231013 $b ABA008
- 991 __
- $a 20231026105342 $b ABA008
- 999 __
- $a ok $b bmc $g 2000487 $s 1203355
- BAS __
- $a 3
- BAS __
- $a PreBMC-MEDLINE
- BMC __
- $a 2023 $b 46 $c 7 $d 729-737 $e 20230520 $i 1540-8159 $m Pacing and clinical electrophysiology $n Pacing Clin Electrophysiol $x MED00003671
- LZP __
- $a Pubmed-20231013