Measurement of homocysteine and other aminothiols in plasma: advantages of using tris(2-carboxyethyl)phosphine as reductant compared with tri-n-butylphosphine
Jazyk angličtina Země Velká Británie, Anglie Médium print
Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem
PubMed
11568092
Knihovny.cz E-zdroje
- MeSH
- fosfiny * MeSH
- homocystein krev MeSH
- indikátory a reagencie MeSH
- lidé MeSH
- oxidace-redukce MeSH
- sulfhydrylové sloučeniny krev MeSH
- vysokoúčinná kapalinová chromatografie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- srovnávací studie MeSH
- Názvy látek
- fosfiny * MeSH
- homocystein MeSH
- indikátory a reagencie MeSH
- sulfhydrylové sloučeniny MeSH
- tri-n-butylphosphine MeSH Prohlížeč
- tris(2-carboxyethyl)phosphine MeSH Prohlížeč
BACKGROUND: Aminothiols have been implicated in the pathogenesis of arteriosclerosis, and reliable methods are needed to determine their concentrations in body fluids. We present a comparison of two analytical methods and focus on the reduction of low-molecular weight and protein-mixed disulfides of homocysteine, cysteine, cysteinyl-glycine, and glutathione. METHODS: The plasma total aminothiol profile was determined by HPLC with fluorescence detection after derivatization with ammonium 7-fluorobenzo-2-oxa-1,3-diazole-4-sulfonate. Disulfides and protein-bound aminothiols were reduced by either tri-n-butylphosphine (the TBP method) or tris(2-carboxyethyl)phosphine (the TCEP method); the effects of temperature, time of reduction, and concentration of reductants were evaluated. RESULTS: The intraassay imprecision (CV) was <3% for all aminothiols using both methods. The interassay CVs for total cysteine (tCys), total cysteinyl-glycine (tCys-Gly), and total homocysteine (tHcy) were <4% and <8% for the TCEP and TBP methods, respectively, whereas for total glutathione (tGSH) the interassay CV was >12% for both methods. Deming regression and Bland-Altman difference plots showed positive biases for total aminothiol concentrations determined by the TCEP method relative to the TBP method. The mean proportional biases were 65%, 27%, 6%, and 60% for tCys, tCys-Gly, tHcy, and tGSH, respectively. The calculated concentrations of total aminothiols by the TCEP method were less influenced by changes in temperature and concentration of reducing agent or by calibrator matrix. CONCLUSIONS: The agreement between the TCEP and TBP methods was considerably lower for the determination of tCys, tCys-Gly, and tGSH than for tHcy. For total-aminothiol determination, the TCEP method yields better reproducibility and is more robust than the TBP method.
Cysteine restriction-specific effects of sulfur amino acid restriction on lipid metabolism
High cysteine diet reduces insulin resistance in SHR-CRP rats
Metabolism of sulfur compounds in homocystinurias
Enzyme Replacement Therapy Ameliorates Multiple Symptoms of Murine Homocystinuria
Vascular presentation of cystathionine beta-synthase deficiency in adulthood