Flow injection analysis vs. ultra high performance liquid chromatography coupled with tandem mass spectrometry for determination of imatinib in human plasma
Language English Country Netherlands Media print-electronic
Document type Comparative Study, Journal Article, Research Support, Non-U.S. Gov't
PubMed
20713033
DOI
10.1016/j.cca.2010.08.014
PII: S0009-8981(10)00511-5
Knihovny.cz E-resources
- MeSH
- Benzamides MeSH
- Blood Chemical Analysis methods MeSH
- Leukemia, Myelogenous, Chronic, BCR-ABL Positive blood drug therapy MeSH
- Imatinib Mesylate MeSH
- Humans MeSH
- Piperazines blood therapeutic use MeSH
- Flow Injection Analysis methods MeSH
- Pyrimidines blood therapeutic use MeSH
- Reproducibility of Results MeSH
- Tandem Mass Spectrometry methods MeSH
- Chromatography, High Pressure Liquid methods MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Research Support, Non-U.S. Gov't MeSH
- Comparative Study MeSH
- Names of Substances
- Benzamides MeSH
- Imatinib Mesylate MeSH
- Piperazines MeSH
- Pyrimidines MeSH
BACKGROUND: The aim of this study was to develop, validate and compare flow injection analysis (FIA) and ultra-high-performance liquid chromatography (LC)/tandem mass spectrometry methods for the determination of imatinib in plasma from patients with chronic myeloid leukemia. METHODS: The plasma for analysis by both methods was deproteinated by methanol containing d8-imatinib. The separation was achieved on a 1.7 μm C18 column with a linear gradient (4 mM ammonium formiate and acetonitrile, pH 3.2). FIA was performed at flow rate of 0.03 mL/min (0.1% formic acid in methanol). Multiple reaction monitoring mode on the tandem mass spectrometer (API 4000, AB Sciex) in positive ESI were used for detection. RESULTS: The total analysis times were 3.2 (LC) and 0.75 min (FIA). Both methods were successfully validated and applied to the plasma patients samples. The limits of quantification were 4.1 and 30.8 ng/mL; imprecisions were less than 5.7% and recovery ranged between 93 and 105%, for the LC and FIA, respectively. The methods revealed an agreement with a mean difference of 1.46 ng/mL (SD 28.95 ng/mL). CONCLUSIONS: The high-throughput methods that were developed are suitable for the therapeutic drug monitoring of imatinib in plasma. They can be used in routine clinical practice.