Flow injection analysis vs. ultra high performance liquid chromatography coupled with tandem mass spectrometry for determination of imatinib in human plasma
Jazyk angličtina Země Nizozemsko Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem
PubMed
20713033
DOI
10.1016/j.cca.2010.08.014
PII: S0009-8981(10)00511-5
Knihovny.cz E-zdroje
- MeSH
- benzamidy MeSH
- biochemická analýza krve metody MeSH
- chronická myeloidní leukemie krev farmakoterapie MeSH
- imatinib mesylát MeSH
- lidé MeSH
- piperaziny krev terapeutické užití MeSH
- průtoková injekční analýza metody MeSH
- pyrimidiny krev terapeutické užití MeSH
- reprodukovatelnost výsledků MeSH
- tandemová hmotnostní spektrometrie metody MeSH
- vysokoúčinná kapalinová chromatografie metody MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- srovnávací studie MeSH
- Názvy látek
- benzamidy MeSH
- imatinib mesylát MeSH
- piperaziny MeSH
- pyrimidiny MeSH
BACKGROUND: The aim of this study was to develop, validate and compare flow injection analysis (FIA) and ultra-high-performance liquid chromatography (LC)/tandem mass spectrometry methods for the determination of imatinib in plasma from patients with chronic myeloid leukemia. METHODS: The plasma for analysis by both methods was deproteinated by methanol containing d8-imatinib. The separation was achieved on a 1.7 μm C18 column with a linear gradient (4 mM ammonium formiate and acetonitrile, pH 3.2). FIA was performed at flow rate of 0.03 mL/min (0.1% formic acid in methanol). Multiple reaction monitoring mode on the tandem mass spectrometer (API 4000, AB Sciex) in positive ESI were used for detection. RESULTS: The total analysis times were 3.2 (LC) and 0.75 min (FIA). Both methods were successfully validated and applied to the plasma patients samples. The limits of quantification were 4.1 and 30.8 ng/mL; imprecisions were less than 5.7% and recovery ranged between 93 and 105%, for the LC and FIA, respectively. The methods revealed an agreement with a mean difference of 1.46 ng/mL (SD 28.95 ng/mL). CONCLUSIONS: The high-throughput methods that were developed are suitable for the therapeutic drug monitoring of imatinib in plasma. They can be used in routine clinical practice.