Randomized, Controlled Trial of the Safety and Effectiveness of a Contact Force-Sensing Irrigated Catheter for Ablation of Paroxysmal Atrial Fibrillation: Results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) Study
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu časopisecké články, multicentrická studie, randomizované kontrolované studie, práce podpořená grantem
PubMed
26260733
DOI
10.1161/circulationaha.114.014092
PII: CIRCULATIONAHA.114.014092
Knihovny.cz E-zdroje
- Klíčová slova
- ablation techniques, atrial fibrillation, catheter ablation,
- MeSH
- Brugadův syndrom MeSH
- fibrilace síní diagnóza chirurgie MeSH
- katetrizační ablace škodlivé účinky metody MeSH
- lidé středního věku MeSH
- lidé MeSH
- následné studie MeSH
- nemoci převodního systému srdečního MeSH
- perikarditida diagnóza etiologie MeSH
- převodní systém srdeční abnormality MeSH
- prospektivní studie MeSH
- senioři MeSH
- srdeční arytmie diagnóza etiologie MeSH
- srdeční katetrizace škodlivé účinky metody MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH
BACKGROUND: Contact force (CF) is a major determinant of lesion size and transmurality and has the potential to improve efficacy of atrial fibrillation ablation. This study sought to evaluate the safety and effectiveness of a novel irrigated radiofrequency ablation catheter that measures real-time CF in the treatment of patients with paroxysmal atrial fibrillation. METHODS AND RESULTS: A total of 300 patients with symptomatic, drug-refractory, paroxysmal atrial fibrillation were enrolled in a prospective, multicenter, randomized, controlled trial and randomized to radiofrequency ablation with either a novel CF-sensing catheter or a non-CF catheter (control). The primary effectiveness end point consisted of acute electrical isolation of all pulmonary veins and freedom from recurrent symptomatic atrial arrhythmia off all antiarrhythmic drugs at 12 months. The primary safety end point included device-related serious adverse events. End points were powered to show noninferiority. All pulmonary veins were isolated in both groups. Effectiveness was achieved in 67.8% and 69.4% of subjects in the CF and control arms, respectively (absolute difference, -1.6%; lower limit of 1-sided 95% confidence interval, -10.7%; P=0.0073 for noninferiority). When the CF arm was stratified into optimal CF (≥90% ablations with ≥10 g) and nonoptimal CF groups, effectiveness was achieved in 75.9% versus 58.1%, respectively (P=0.018). The primary safety end point occurred in 1.97% and 1.40% of CF patients and control subjects, respectively (absolute difference, 0.57%; upper limit of 1-sided 95% confidence interval, 3.61%; P=0.0004 for noninferiority). CONCLUSIONS: The CF ablation catheter met the primary safety and effectiveness end points. Additionally, optimal CF was associated with improved effectiveness. CLINICAL TRIAL REGISTRATION: http://www.clinicaltrials.gov. Unique identifier: NCT01278953.
Citace poskytuje Crossref.org
ClinicalTrials.gov
NCT01278953