Diagnosis of Intoxication by the Organophosphate VX: Comparison Between an Electrochemical Sensor and Ellman´s Photometric Method

. 2008 Sep 01 ; 8 (9) : 5229-5237. [epub] 20080901

Status PubMed-not-MEDLINE Jazyk angličtina Země Švýcarsko Médium electronic

Typ dokumentu časopisecké články

Perzistentní odkaz   https://www.medvik.cz/link/pmid27873811

An electrochemical sensor is introduced as a tool applicable for diagnosis of intoxication by cholinesterase inhibitors caused by the well-known nerve agent VX. The traditional Ellman method was chosen for comparison with the sensor's analytical parameters. Both methods are based on estimation of blood cholinesterase inhibition as a marker of intoxication. While Ellman´s method provided a limit of detection of 5.2´10-7 M for blood containing VX, the electrochemical sensor was able to detect 4.0´10-7 M. Good correlation between both methods was observed (R = 0.92). The electrochemical sensor could be considered a convenient tool for a fast yet accurate method, easily available for field as well as laboratory use. Time and cost savings are key features of the sensor-based assay.

Zobrazit více v PubMed

Milatovic D., Gupta R.C., Aschner M. Anticholinesterase toxicity and oxidative stress. Scientif. World J. 2006;6:295–310. PubMed PMC

Goldman L.R. Managing pesticide chronic health risks: U.S. policies. J. Agromed. 2007;12:67–75. PubMed

Eyer F., Meischner V., Kiderlen D., Thiermann H., Worek F., Haberkorn M., Felgenhauer N., Zilker T., Eyer P. Human parathion poisoning. A toxicokinetic analysis. Toxicol. Rev. 2003;22:143–163. PubMed

Mega M.S., Cummings J.L., O'Connor S.M., Dinov I.D., Reback E., Felix J., Masterman D.L., Phelp M.E., Small G.W., Toga A.W. Cognitive and metabolic responses to metrifonate therapy in Alzheimer disease. Neuropsychiary Neuropsychol. Behav. Neurol. 2001;14:63–68. PubMed

Goozner B., Lutwick L.I., Bourke E. Chemical terrorism: a primer for 2002. J. Assoc. Acad. Minor. Phys. 2002;13:14–18. PubMed

Ballared C.G., Greig N.H., Guillozet-Bongaarts A.L., Enz A., Darvesh S. Cholinesterases: roles in the brain during health and disease. Curr. Alzheimer Res. 2005;2:307–318. PubMed

Jameson R.R., Seidler F.J., Slotkin T.A. Nonenzymatic functions of acetylcholinesterase splice variants in the developmental neurotoxicity of organophosphates: chlorpyrifos, chlorpyrifos oxon, and diazinon. Environ. Health Perspect. 2007;115:65–70. PubMed PMC

Goel A., Aggarwal P. Pesticide poisoning. Natl. Med. J. India. 2007;20:182–191. PubMed

Worek F., Koller M., Thiermann H., Szinicz L. Diagnostic aspects of organophosphate poisoning. Toxicology. 2005;214:182–189. PubMed

Margariti M.G., Tsakalof A.F., Tsatsakis A.M. Analytical methods of biological monitoring for exposure to pesticides: recent update. Ther. Drug. Monit. 2007;29:150–163. PubMed

Kim B.M., El-Aty A.M.A., Hwang T.E., Jin L.T., Kim Y.S., Shim J.H. Development of an acetylcholinesterase-based detection kit for the determination of organophosporus and carbamae pesticide residues in agricultural samples. Bull. Korean Chem. Soc. 2007;28:929–935.

No H.Y., Kim Y.A., Lee Y.T., Lee H.S. Cholinesterase-based dipstick assay for the detection of organophosphate and carbamate pesticides. Anal. Chim. Acta. 2007;594:37–43. PubMed

Pohanka M., Jun D., Kalasz H., Kuca K. Cholinesterase biosensor construction – a review. Prot. Pept. Letters. 2008;15:795–798. PubMed

Pohanka M., Kuca K., Jun D. Aflatoxin assay using an amperometric sensor strip and acetylcholinesterase as recognition element. Sens. Lett. 2008;6:450–453.

Pohanka M., Kuca K., Jun D. Sensor system based on acetylcholinesterase in homogenous phase for analysis of paraoxon. Anal. Lett. 2008 In press.

Pohanka M., Jun D., Kuca K. Amperometric biosensor for evaluation of competitive cholinesterase inhibition by the reactivator HI-6. Anal. Lett. 2007;40:2351–2359.

Baydin A., Aygun D., Yazici M., Karatas A., Deniz T., Yardan T. Is there a relationship between the blood cholinesterase and QTc interval in the patients with acute organophosphate poisoning? Int. J. Clin. Pract. 2007;61:927–930. PubMed

Ellman G.L., Courtney D.K, Andres V., Featherstone R.M. A new and rapid colorimetric determination of acetylcholinesterase activity. Biochem. Pharmacol. 1961;7:88–95. PubMed

Pohanka M., Pavlis O., Skladal M. Rapid characterization of monoclonal antibodies using the piezoelectric immunosensor. Sensors. 2007;7:341–353.

Pohanka M., Skladal P. Serological diagnosis of tularemia in mice using the amperometric immunosensor. Electroanalysis. 2007;19:2507–2512.

Pohanka M., Treml F., Hubalek M., Bandouchova H, Beklova M., Pikula J. Piezoelectric biosensor for a simple serological diagnosis of tularemia in infected European brown hares (Lepus europaeus) Sensors. 2007;7:2825–2834. PubMed PMC

Pohanka M., Zboril P. Amperometric biosensor for D-lactate assay. Food Technol. Biotechnol. 2008;46:107–110.

Adam V., Mikelova R., Hubalek J., Hanustiak P., Beklova M., Hodek P., Horna A., Trnkova L., Stiborova M., Zeman L., Kizek R. Utilizing of square wave voltammetry to detect flavonoids in the presence of human urine. Sensors. 2007;7:2402–2418. PubMed PMC

Krizkova S., Beklova M., Pikula J., Adam V., Horna A., Kizek R. Hazards of secondary bromadiolone intoxications evaluated using high-performance liquid chromatography with electrochemical detection. Sensors. 2007;7:1271–1286.

Supalkova V., Huska D., Diopan V., Hanustiak P., Zika O., Stejskal K., Baloun J., Pikula J., Havel L., Zehnalek J., Adam V., Trnkova L., Beklova M., Kizek R. Electroanalysis of plant thiols. Sensors. 2007;7:932–952.

Krizkova S., Ryant P., Krystofova O., Adam V., Galiova M., Beklova M., Babula P., Kaiser J., Novotny K., Novotny J., Liska M., Malina R., Zehnalek J., Hubalek J., Havel L., Kizek R. Multi-instrumental analysis of tissues of sunflower plants treated with silver(I) ions – plants as bioindicators of environmental pollution. Sensors. 2008;8:445–463. PubMed PMC

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...