Interlaboratory Comparison Investigations (ICIs) for human biomonitoring of chromium as part of the quality assurance programme under HBM4EU
Language English Country Germany Media print-electronic
Document type Journal Article
PubMed
34954563
DOI
10.1016/j.jtemb.2021.126912
PII: S0946-672X(21)00202-9
Knihovny.cz E-resources
- Keywords
- Chromium, Electrothermal atomic absorption spectrometry (EAAS), HBM4EU, Human biomonitoring (HBM), Inductively coupled plasma mass spectrometry (ICP-MS), Interlaboratory comparison investigation (ICI), Occupational exposure, Plasma, Red blood cells, Urine, Whole blood,
- MeSH
- Biological Monitoring * MeSH
- Chromium analysis MeSH
- Humans MeSH
- Environmental Monitoring MeSH
- Occupational Exposure * analysis MeSH
- Spectrophotometry, Atomic MeSH
- Check Tag
- Humans MeSH
- Publication type
- Journal Article MeSH
- Names of Substances
- Chromium MeSH
BACKGROUND: The pan-European human biomonitoring initiative HBM4EU targets the harmonization of human biomonitoring (HBM) procedures and data for both environmental and occupational exposure, including chromium. The determination of chromium in urine (U-Cr), plasma (P-Cr) and whole blood (WB-Cr) is a common HBM application in employees occupationally exposed to chromium (VI) compounds. METHODS: European laboratories which have registered as candidate laboratories for chromium analysis within HBM4EU were invited to participate in a quality assurance/qualitycontrol (QA/QC) programme comprising interlaboratory comparison investigations (ICI) for the parameters U-Cr, P-Cr and WB-Cr. Participating laboratories received two samples of different concentrations in each of four rounds and were asked to analyse the samples using their standard analytical procedure. The data were evaluated by the Z-score approach and were reported to the participants after each round. RESULTS: The majority of the 29 participating laboratories obtained satisfactory results, although low limits of quantification were required to quantify chromium concentrations in some of the ICI materials. The robust relative standard deviation of the participants' results (study RSDR) obtained from all ICI runs ranged from 6 to 16 % for U-Cr, 7-18 % for P-Cr and 4-47 % for WB-Cr. The application of both inductively coupled plasma mass spectrometry (ICP-MS) and electrothermal atomic absorption spectrometry (EAAS) appeared appropriate for the determination of chromium in urine, plasma and whole blood with regard to occupational exposure levels. CONCLUSION: This QA/QC programme succeeded in establishing a network of laboratories with high analytical comparability and accuracy for the analysis of chromium across Europe.
Aarhus University Department of Environmental Science Frederiksborgvej 399 4000 Roskilde Denmark
Department of Environmental Sciences Jožef Stefan Institute Jamova 39 1000 Ljubljana Slovenia
Environmental Exposure and Epidemiology Norwegian Institute of Public Health Oslo Norway
L'Université Nantes Angers Le Mans F 44307 Nantes France
National Center for Environmental Health Instituto de Salud Carlos 3 Majadahonda Spain
References provided by Crossref.org