Are all HCL systems the same? long term outcomes of three HCL systems in children with type 1 diabetes: real-life registry-based study

. 2023 ; 14 () : 1283181. [epub] 20231016

Jazyk angličtina Země Švýcarsko Médium electronic-ecollection

Typ dokumentu časopisecké články, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid37908748

OBJECTIVE: To compare parameters of glycemic control among three types of hybrid closed loop (HCL) systems in children with T1D (CwD) using population-wide data from the national pediatric diabetes registry ČENDA. METHODS: CwD aged <19 years treated with Medtronic MiniMed 780G (780G), Tandem t:slim X2 (Control-IQ) or do-it-yourself AndroidAPS (AAPS) systems for >12 months and monitored by CGM >70% of the time were included. HbA1c, times in glycemic ranges, and Glycemia Risk Index (GRI) were used for cross-sectional comparison between the HCL systems. RESULTS: Data from 512 CwD were analyzed. 780G, Control-IQ and AAPS were used by 217 (42.4%), 211 (41.2%), and 84 (16.4%) CwD, respectively. The lowest HbA1c value was observed in the AAPS group (44 mmol/mol; IQR 8.0, p<0.0001 vs any other group), followed by Control-IQ and 780G groups (48 (IQR 11) and 52 (IQR 10) mmol/mol, respectively). All of the systems met the recommended criteria for time in range (78% in AAPS, 76% in 780G, and 75% in Control-IQ users). CwD using AAPS spent significantly more time in hypoglycemia (5% vs 2% in 780G and 3% in Control-IQ) and scored the highest GRI (32, IQR 17). The lowest GRI (27, IQR 15) was seen in 780G users. CONCLUSION: Although all HCL systems proved effective in maintaining recommended long-term glycemic control, we observed differences that illustrate strengths and weaknesses of particular systems. Our findings could help in individualizing the choice of HCL systems.

Zobrazit více v PubMed

Boughton CK, Hovorka R. New closed-loop insulin systems. Diabetologia (2021) 64:1007–15. doi: 10.1007/s00125-021-05391-w PubMed DOI PMC

Carlson AL, Sherr JL, Shulman DI, Garg SK, Pop-Busui R, Bode BW, et al. . Safety and glycemic outcomes during the miniMedTM advanced hybrid closed-loop system pivotal trial in adolescents and adults with type 1 diabetes. Diabetes Technol Ther (2022) 24:178–89. doi: 10.1089/dia.2021.0319 PubMed DOI PMC

Da Silva J, Lepore G, Battelino T, Arrieta A, Castañeda J, Grossman B, et al. . Real-world performance of the miniMedTM 780G system: first report of outcomes from 4120 users. Diabetes Technol Ther (2022) 24:113–9. doi: 10.1089/dia.2021.0203 PubMed DOI PMC

Pinsker JE, Müller L, Constantin A, Leas S, Manning M, McElwee Malloy M, et al. . Real-world patient-reported outcomes and glycemic results with initiation of control-IQ technology. Diabetes Technol Ther (2021) 23:120–7. doi: 10.1089/dia.2020.0388 PubMed DOI PMC

Sumnik Z, Pavlikova M, Neuman V, Petruzelkova L, Konecna P, Venhacova P, et al. . Glycemic control by treatment modalities: national registry-based population data in children and adolescents with type 1 diabetes. Horm Res Paediatr (2023). doi: 10.1159/000530833 PubMed DOI

Braune K, O’Donnell S, Cleal B, Lewis D, Tappe A, Willaing I, et al. . Real-world use of do-it-yourself artificial pancreas systems in children and adolescents with type 1 diabetes: online survey and analysis of self-reported clinical outcomes. JMIR Mhealth Uhealth (2019) 7:e14087. doi: 10.2196/14087 PubMed DOI PMC

Collyns OJ, Meier RA, Betts ZL, Chan DSH, Frampton C, Frewen CM, et al. . Improved glycemic outcomes with medtronic miniMed advanced hybrid closed-loop delivery: results from a randomized crossover trial comparing automated insulin delivery with predictive low glucose suspend in people with type 1 diabetes. Diabetes Care (2021) 44:969–75. doi: 10.2337/dc20-2250 PubMed DOI

Breton MD, Kanapka LG, Beck RW, Ekhlaspour L, Forlenza GP, Cengiz E, et al. . A randomized trial of closed-loop control in children with type 1 diabetes. N Engl J Med (2020) 383:836–45. doi: 10.1056/NEJMoa2004736 PubMed DOI PMC

Wadwa RP, Reed ZW, Buckingham BA, DeBoer MD, Ekhlaspour L, Forlenza GP, et al. . Trial of hybrid closed-loop control in young children with type 1 diabetes. N Engl J Med (2023) 388:991–1001. doi: 10.1056/NEJMoa2210834 PubMed DOI PMC

Choudhary P, Kolassa R, Keuthage W, Kroeger J, Thivolet C, Evans M, et al. . Advanced hybrid closed loop therapy versus conventional treatment in adults with type 1 diabetes (ADAPT): a randomised controlled study. Lancet Diabetes Endocrinol (2022) 10:720–31. doi: 10.1016/S2213-8587(22)00212-1 PubMed DOI

Petruzelkova L, Jiranova P, Soupal J, Kozak M, Plachy L, Neuman V, et al. . Pre-school and school-aged children benefit from the switch from a sensor-augmented pump to an AndroidAPS hybrid closed loop: A retrospective analysis. Pediatr Diabetes (2021) 22:594–604. doi: 10.1111/pedi.13190 PubMed DOI

Braune K, Lal RA, Petruželková L, Scheiner G, Winterdijk P, Schmidt S, et al. . Open-source automated insulin delivery: international consensus statement and practical guidance for health-care professionals. Lancet Diabetes Endocrinol (2022) 10:58–74. doi: 10.1016/S2213-8587(21)00267-9 PubMed DOI PMC

Breton MD, Kovatchev BP. One year real-world use of the control-IQ advanced hybrid closed-loop technology. Diabetes Technol Ther (2021) 23:601–8. doi: 10.1089/dia.2021.0097 PubMed DOI PMC

Gawrecki A, Zozulinska-Ziolkiewicz D, Michalak MA, Adamska A, Michalak M, Frackowiak U, et al. . Safety and glycemic outcomes of do-it-yourself AndroidAPS hybrid closed-loop system in adults with type 1 diabetes. PLoS One (2021) 16:e0248965. doi: 10.1371/journal.pone.0248965 PubMed DOI PMC

Šumník Z, Venháčová J, Škvor J, Pomahačová R, Konečná P, Neumann D, et al. . Five years of improving diabetes control in Czech children after the establishment of the population-based childhood diabetes register ČENDA. Pediatr Diabetes (2020) 21:77–87. doi: 10.1111/pedi.12929 PubMed DOI

Šumník Z, Pavlíková M, Pomahačová R, Venháčová P, Petruželková L, Škvor J, et al. . Use of continuous glucose monitoring and its association with type 1 diabetes control in children over the first 3 years of reimbursement approval: Population data from the ČENDA registry. Pediatr Diabetes (2021) 22:439–47. doi: 10.1111/pedi.13184 PubMed DOI

Battelino T, Danne T, Bergenstal RM, Amiel SA, Beck R, Biester T, et al. . Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care (2019) 42:1593–603. doi: 10.2337/dci19-0028 PubMed DOI PMC

Klonoff DC, Wang J, Rodbard D, Kohn MA, Li C, Liepmann D, et al. . A glycemia risk index (GRI) of hypoglycemia and hyperglycemia for continuous glucose monitoring validated by clinician ratings. J Diabetes Sci Technol (2022) 193229682210852. doi: 10.1177/19322968221085273 PubMed DOI PMC

Cefalu M, Ridgeway G, Mccaffrey D, Morral A, Griffin AB, Burgette L. Toolkit for weighting and analysis of nonequivalent groups. R Package (2006).

Sundberg F, deBeaufort C, Krogvold L, Patton S, Piloya T, Smart C, et al. . ISPAD Clinical Practice Consensus Guidelines 2022: Managing diabetes in preschoolers. Pediatr Diabetes (2022) 23:1496–511. doi: 10.1111/pedi.13427 PubMed DOI PMC

Arrieta A, Battelino T, Scaramuzza AE, Da Silva J, Castañeda J, Cordero TL, et al. . Comparison of MiniMed 780G system performance in users aged younger and older than 15 years: Evidence from 12 870 real-world users. Diabetes Obes Metab (2022) 24:1370–9. doi: 10.1111/dom.14714 PubMed DOI PMC

Schiaffini R, Deodati A, Nicoletti MC, Carducci C, Ciampalini P, Lorubbio A, et al. . Comparison of two advanced hybrid closed loop in a pediatric population with type 1 diabetes: a real-life observational study. Acta Diabetol (2022) 59:959–64. doi: 10.1007/s00592-022-01886-z PubMed DOI

Bassi M, Patti L, Silvestrini I, Strati MF, Ponzano M, Minuto N, et al. . One-year follow-up comparison of two hybrid closed-loop systems in Italian children and adults with type 1 diabetes. Front Endocrinol (Lausanne) (2023) 14:1099024. doi: 10.3389/fendo.2023.1099024 PubMed DOI PMC

Tsujino D, Nishimura R, Onda Y, Seo C, Ando K, Morimoto A, et al. . The relationship between HbA1c values and the occurrence of hypoglycemia as assessed by continuous glucose monitoring in patients with type 1 diabetes. Diabetol Metab Syndr (2016) 8:53. doi: 10.1186/s13098-016-0167-z PubMed DOI PMC

Lombardo F, Passanisi S, Alibrandi A, Bombaci B, Bonfanti R, Delvecchio M, et al. . MiniMed 780G six-month use in children and adolescents with type 1 diabetes: clinical targets and predictors of optimal glucose control. Diabetes Technol Ther (2023) 25:404–13. doi: 10.1089/dia.2022.0491 PubMed DOI

Najít záznam

Citační ukazatele

Nahrávání dat ...

    Možnosti archivace