This article presents the protocol on Quality Controls in PET/CT and PET/MRI published online in May 2022 by the European Federation of Organisations for Medical Physics (EFOMP), which was developed by the Working group for PET/CT and PET/MRI Quality Control (QC) protocol. The main objective of this protocol was to comprehensively provide simple and practical procedures that may be integrated into clinical practice to identify changes in the PET/CT/MRI system's performance and avoid short- and long-term quality deterioration. The protocol describes the quality control procedures on radionuclide calibrators, weighing scales, PET, CT and MRI systems using selected and measurable parameters that are directly linked to clinical images quality. It helps to detect problems before they can impact clinical studies in terms of safety, image quality, quantification accuracy and patient radiation dose. CT and MRI QCs are described only in the context of their use for PET (attenuation correction and anatomical localization) imaging. Detailed step-by-step instructions have been provided, limiting any misinterpretations or interpersonal variations as much as possible. This paper presents the main characteristics of the protocol illustrated together with a brief summary of the content of each chapter. A regular QC based on the proposed protocol would guarantee that PET/CT and PET/MRI systems operate under optimal conditions, resulting in the best performance for routine clinical tasks.
PURPOSE: An EFOMP Working Group (WG) was created in 2020 to establish recommendations for PET/CT/MRI Quality Control (QC). The WG's intention was to create a document containing a set of measurements suitable for routine practice. In order to map the current situation in PET facilities, the WG prepared a survey addressed to European Medical Physics Experts (MPE). METHODS: The survey was conducted using an electronic questionnaire with 10 sections, for a total of 43 multiple choice or open questions. Data regarding general information, model of installed scanners, contract of maintenance and phantoms available were collected. The focal part of the questionnaire concerned the QC protocol adopted and accreditation programs. RESULTS: 123 answers from 24 countries were collected. 90.2% of the respondents are affiliated as staff MPEs; 45% have non-digital TOF PET/CT scanners with a contract of maintenance (97.6%). In 98.4% and 86.8% of responding centres a sealed source for daily QC and the NEMA Image Quality Phantom were present. 94.3% of respondents perform daily QC according to manufacturer recommendations, while NEMA Tests are not performed routinely (51.2%). 56.1% of the respondents have scanners accredited by a national or international organization. 56% of the centres perform annual CT tests, while more than 90% do not perform any MRI QCs. CONCLUSIONS: The results of the survey show that there is a lack of harmonization in the PET QC procedures across Europe. The information obtained will guide the WG in proposing a guideline containing a set of measurements suitable for the clinical routine.
- MeSH
- fantomy radiodiagnostické MeSH
- lidé MeSH
- magnetická rezonanční tomografie MeSH
- PET/CT * MeSH
- pozitronová emisní tomografie * metody MeSH
- průzkumy a dotazníky MeSH
- řízení kvality MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- Geografické názvy
- Evropa MeSH