Determination of 8-iso-prostaglandin F(2alpha) in exhaled breath condensate using combination of immunoseparation and LC-ESI-MS/MS

. 2008 May 01 ; 867 (1) : 8-14. [epub] 20080229

Jazyk angličtina Země Nizozemsko Médium print-electronic

Typ dokumentu časopisecké články, práce podpořená grantem, validační studie

Perzistentní odkaz   https://www.medvik.cz/link/pmid18373964
Odkazy

PubMed 18373964
DOI 10.1016/j.jchromb.2008.02.019
PII: S1570-0232(08)00138-4
Knihovny.cz E-zdroje

Rapid and precise method for the determination of 8-iso-prostaglandin F(2alpha), an essential marker of the oxidative stress, in exhaled breath condensate (EBC) was developed. The protocol consisted of stable isotope dilution, immunoseparation combined with selective and sensitive LC-ESI-MS/MS operated in multiple reaction monitoring (MRM) mode. The imprecision of the developed method was below 8.8%, the parameter of mean inaccuracy was determined as <9.6% (0-250pg of 8-iso-prostaglandin F(2alpha)/ml EBC). The limit of detection (LOD) was 1 pg/ml EBC and limit of quantification (LOQ) 5 pg/ml EBC. A significant difference in 8-iso-prostaglandin F(2alpha) content between the group of asbestosis patients and healthy volunteers was found.