First-in-human clinical series of a novel conformable large-lattice pulsed field ablation catheter for pulmonary vein isolation

. 2024 Mar 30 ; 26 (4) : .

Jazyk angličtina Země Velká Británie, Anglie Médium print

Typ dokumentu časopisecké články, multicentrická studie, práce podpořená grantem, klinické zkoušky

Perzistentní odkaz   https://www.medvik.cz/link/pmid38584468

Grantová podpora
Affera, Inc

AIMS: Pulsed field ablation (PFA) has significant advantages over conventional thermal ablation of atrial fibrillation (AF). This first-in-human, single-arm trial to treat paroxysmal AF (PAF) assessed the efficiency, safety, pulmonary vein isolation (PVI) durability and one-year clinical effectiveness of an 8 Fr, large-lattice, conformable single-shot PFA catheter together with a dedicated electroanatomical mapping system. METHODS AND RESULTS: After rendering the PV anatomy, the PFA catheter delivered monopolar, biphasic pulse trains (5-6 s per application; ∼4 applications per PV). Three waveforms were tested: PULSE1, PULSE2, and PULSE3. Follow-up included ECGs, Holters at 6 and 12 months, and symptomatic and scheduled transtelephonic monitoring. The primary and secondary efficacy endpoints were acute PVI and post-blanking atrial arrhythmia recurrence, respectively. Invasive remapping was conducted ∼75 days post-ablation. At three centres, PVI was performed by five operators in 85 patients using PULSE1 (n = 30), PULSE2 (n = 20), and PULSE3 (n = 35). Acute PVI was achieved in 100% of PVs using 3.9 ± 1.4 PFA applications per PV. Overall procedure, transpired ablation, PFA catheter dwell and fluoroscopy times were 56.5 ± 21.6, 10.0 ± 6.0, 19.1 ± 9.3, and 5.7 ± 3.9 min, respectively. No pre-defined primary safety events occurred. Upon remapping, PVI durability was 90% and 99% on a per-vein basis for the total and PULSE3 cohort, respectively. The Kaplan-Meier estimate of one-year freedom from atrial arrhythmias was 81.8% (95% CI 70.2-89.2%) for the total, and 100% (95% CI 80.6-100%) for the PULSE3 cohort. CONCLUSION: Pulmonary vein isolation (PVI) utilizing a conformable single-shot PFA catheter to treat PAF was efficient, safe, and effective, with durable lesions demonstrated upon remapping.

Zobrazit více v PubMed

Joglar JA, Chung MK, Armbruster AL, Benjamin EJ, Chyou JY, Cronin EMet al. . 2023 ACC/AHA/ACCP/HRS guideline for the diagnosis and management of atrial fibrillation: a report of the American Heart Association/American College of Cardiology joint committee on clinical practice guidelines. J Am Coll Cardiol 2024;83:109–279. PubMed PMC

Howard B, Haines DE, Verma A, Kirchhof N, Barka N, Onal Bet al. . Characterization of phrenic nerve response to pulsed field ablation. Circ Arrhythm Electrophysiol 2022;15:e010127. PubMed PMC

Neven K, van Es R, van Driel V, van Wessel H, Fidder H, Vink Aet al. . Acute and long-term effects of full-power electroporation ablation directly on the porcine esophagus. Circ Arrhythm Electrophysiol 2017;10:e004672. PubMed

Koruth JS, Kuroki K, Iwasawa J, Enomoto Y, Viswanathan R, Brose Ret al. . Preclinical evaluation of pulsed field ablation: electrophysiological and histological assessment of thoracic vein isolation. Circ Arrhythm Electrophysiol 2019;12:e007781. PubMed PMC

Koruth JS, Kuroki K, Kawamura I, Brose R, Viswanathan R, Buck EDet al. . Pulsed field ablation vs radiofrequency ablation: esophageal injury in a novel porcine model. Circ Arrhythm Electrophysiol 2020;13:e008303. PubMed PMC

Koruth JS, Kuroki J, Kawamura I, Stoffregen WC, Dukkipati SR, Neuzil Pet al. . Focal pulsed field ablation for pulmonary vein isolation and linear atrial lesions: a preclinical assessment of safety and durability. Circ Arrhythm Electrophysiol 2020;13:e008716. PubMed

Howard B, Haines DE, Verma A, Packer D, Kirchhof N, Barka Net al. . Reduction in pulmonary vein stenosis and collateral damage with pulsed field ablation compared with radiofrequency ablation in a canine model. Circ Arrhythm Electrophysiol 2020;13:e008337. PubMed PMC

Reddy VY, Neuzil P, Koruth JS, Petru J, Funosako M, Cochet Het al. . Pulsed field ablation for pulmonary vein isolation in atrial fibrillation. J Am Coll Cardiol 2019;74:315–26. PubMed

Reddy VY, Dukkipati SR, Neuzil P, Anic A, Petru J, Funasako Met al. . Pulsed field ablation of paroxysmal atrial fibrillation: one-year outcomes of IMPULSE, PEFCAT & PEFCAT II. JACC Clin Electrophysiol 2021;7:614–27. PubMed

Reddy VY, Anic A, Koruth J, Petru J, Funasako M, Minami Ket al. . Pulsed field ablation in patients with persistent atrial fibrillation. J Am Coll Cardiol 2020;76:1068–80. PubMed

Reddy VY, Peichl P, Anter E, Rackauskas G, Petru J, Funasako Met al. . A focal ablation catheter toggling between radiofrequency and pulsed field energy to treat atrial fibrillation. JACC Clin Electrophysiol 2023;9:1786–801. PubMed

Cochet H, Nakatani Y, Sridi-Cheniti S, Cheniti G, Ramirez FD, Nakashima Tet al. . Pulsed field ablation selectively spares the oesophagus during pulmonary vein isolation for atrial fibrillation. Europace 2021;23:1391–9. PubMed PMC

Kuroki K, Whang W, Eggert C, Lam J, Leavitt J, Kawamura Iet al. . Ostial dimensional changes after pulmonary vein isolation: pulsed field ablation vs radiofrequency ablation. Heart Rhythm 2020;17:1528–35. PubMed

Mansour M, Gerstenfeld EP, Patel C, Natale A, Whang W, Cuoco FAet al. . Pulmonary vein narrowing after pulsed field versus thermal ablation. Europace 2024;26:euae038. PubMed PMC

Ekanem E, Reddy VY, Schmidt B, Reichlin T, Neven K, Metzner Aet al. . Multi-national survey on the methods, efficacy, and safety on the post-approval clinical use of pulsed field ablation (MANIFEST-PF). Europace 2022;24:1256–66. PubMed PMC

Turagam MK, Neuzil P, Petru J, Funasako M, Koruth JS, Reinders Det al. . PV isolation using a spherical array PFA catheter: application repetition and lesion durability (PULSE-EU study). JACC Clin Electrophysiol 2023;9:638–48. PubMed

Serban T, Mannhart D, Abid Q-U-A, Höchli A, Lazar S, Krisai Pet al. . Durability of pulmonary vein isolation for atrial fibrillation: a meta-analysis and systematic review. Europace 2023;25:euad335. PubMed PMC

Verma A, Haines DE, Boersma LV, Sood N, Natale A, Marchlinski FEet al. . Pulsed field ablation for the treatment of atrial fibrillation: PULSED AF pivotal trial. Circulation 2023;147:1422–32. PubMed PMC

Duytschaever M, De Potter E, Grimaldi M, Anic A, Vijgen J, Neuzil Pet al. . Paroxysmal AF ablation using a novel variable-loop biphasic pulsed field ablation catheter integrated with a 3d mapping system: 1-year outcomes of the multicenter inspIRE study. Circ Arrhythm Electrophysiol 2023;16:e011780. PubMed PMC

Reddy VY, Gerstenfeld EP, Natale A, Whang W, Cuoco FA, Patel Cet al. . Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation. N Engl J Med 2023;389:1660–71. PubMed

Tohoku S, Chun KRJ, Bordignon S, Chen S, Schaack D, Urbanek Let al. . Findings from repeat ablation using high-density mapping after pulmonary vein isolation with pulsed field ablation. Europace 2023;25:433–40. PubMed PMC

Turagam MK, Neuzil P, Schmidt B, Reichlin T, Neven K, Metzner Aet al. . Safety and effectiveness of pulsed field ablation to treat atrial fibrillation: one-year outcomes from the MANIFEST-PF registry. Circulation 2023;148:35–46. PubMed

Kueffer T, Bordignon S, Neven K, Blaauw Y, Hansen J, Adelino R, et al. . Durability of pulmonary vein isolation using pulsed-field ablation: results from the multicenter EU-PORIA registry. JACC Clin Electrophysiol 2024:S2405-500X(23)00917-9. Online ahead of print. PubMed

Schmidt B, Bordignon S, Neven K, Reichlin T, Blaauw Y, Hansen Jet al. . EUropean real-world outcomes with Pulsed field ablatiOn in patients with symptomatic atRIAl fibrillation: lessons from the multi-centre EU-PORIA registry. Europace 2023;25:euad185. PubMed PMC

Koruth J, Kawamura I, Dukkipati SR, Neuzil P, Reddy VY. Preclinical assessment of the feasibility, safety and lesion durability of a novel ‘single-shot’ pulsed field ablation catheter for pulmonary vein isolation. Europace 2023;25:1369–78. PubMed PMC

Yavin HD, Higuchi K, Younis A, Anter E. Lattice-tip catheter for single-shot pulmonary vein isolation with pulsed field ablation. J Interv Card Electrophysiol 2023;66:1741–8. PubMed

Yavin H, Shapira-Daniels A, Barkagan M, Sroubek J, Shim D, Melidone Ret al. . Pulsed field ablation using a lattice electrode for focal energy delivery: biophysical characterization, lesion durability, and safety evaluation. Circ Arrhythm Electrophysiol 2020;13:e008580. PubMed PMC

Deneke T, Jais P, Scaglione M, Schmitt R, DI Biase L, Christopoulos Get al. . Silent cerebral events/lesions related to atrial fibrillation ablation: a clinical review. J Cardiovasc Electrophysiol 2015;26:455–63. PubMed

Calvert P, Kollias G, Pürerfellner H, Narasimhan C, Osorio J, Lip GYHet al. . Silent cerebral lesions following catheter ablation for atrial fibrillation: a state-of-the-art review. Europace 2023;25:euad151. PubMed PMC

Haeusler KG, Eichner FA, Heuschmann PU, Fiebach JB, Engelhorn T, Callans Det al. . Detection of brain lesions after catheter ablation depends on imaging criteria: insights from AXAFA-AFNET 5 trial. Europace 2023;25:euad323. PubMed PMC

Haeusler KG, Eichner FA, Heuschmann PU, Fiebach JB, Engelhorn T, Blank Bet al. . MRI-detected brain lesions and cognitive function in patients with atrial fibrillation undergoing left atrial catheter ablation in the randomized AXAFA-AFNET 5 trial. Circulation 2022;145:906–15. PubMed

Howard B, Verma A, Tzou WS, Mattison L, Kos B, Miklavčič Det al. . Effects of electrode-tissue proximity on cardiac lesion formation using pulsed field ablation. Circ Arrhythm Electrophysiol 2022;15:e011110. PubMed PMC

Mattison L, Verma A, Tarakji KG, Reichlin T, Hindricks G, Sack KLet al. . Effect of contact force on pulsed field ablation lesions in porcine cardiac tissue. J Cardiovasc Electrophysiol 2023;34:693–9. PubMed

Di Biase L, Marazzato J, Govari A, Altman A, Beeckler C, Keyes Jet al. . Pulsed field ablation index–guided ablation for lesion formation: impact of contact force and number of applications in the ventricular model. Circ Arrhythm Electrophysiol 2024;17:e012717. PubMed PMC

Stojadinović P, Wichterle D, Peichl P, Nakagawa H, Čihák R, Hašková Jet al. . Autonomic changes are more durable after radiofrequency than pulsed electric field pulmonary vein ablation. JACC Clin Electrophysiol 2022;8:895–904. PubMed

Musikantow DR, Neuzil P, Petru J, Koruth JS, Kralovec S, Miller MAet al. . Pulsed field ablation to treat atrial fibrillation: autonomic nervous system effects. JACC Clin Electrophysiol 2023;9:481–93. PubMed

Tohoku S, Schmidt B, Schaack D, Bordignon S, Hirokami J, Chen Set al. . Impact of pulsed-field ablation on intrinsic cardiac autonomic nervous system after pulmonary vein isolation. JACC Clin Electrophysiol 2023;9:1864–75. PubMed

Del Monte A, Cespón Fernández M, Vetta G, Della Rocca DG, Pannone L, Mouram Set al. . Quantitative assessment of transient autonomic modulation after single-shot pulmonary vein isolation with pulsed-field ablation. J Cardiovasc Electrophysiol 2023;34:2393–7. PubMed

Katritsis DG, Pokushalov E, Romanov A, Giazitzoglou E, Siontis GCM, Po SSet al. . Autonomic denervation added to pulmonary vein isolation for paroxysmal atrial fibrillation: a randomized clinical trial. J Am Coll Cardiol 2013;62:2318–25. PubMed

Reddy VY, Ekanem E. Multi-national survey on the safety of the post-approval clinical use of pulsed field ablation in 17,000+ patients (MANIFEST-17K). American Heart Association Annual Scientific Sessions 2023. (Presented Nov. 11, 2023 in Philadelphia, PA).

Venier S, Vaxelaire N, Jacon P, Carabelli A, Desbiolles A, Garban Fet al. . Severe acute kidney injury related to haemolysis after pulsed field ablation for atrial fibrillation. Europace 2023;26:euad371. PubMed PMC

Metzner A, Fiala M, Vijgen J, Ouss A, Gunawardene M, Hansen Jet al. . Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom study. Europace 2024;26:euae053. PubMed PMC