Pulsed Field Ablation of Paroxysmal Atrial Fibrillation: 1-Year Outcomes of IMPULSE, PEFCAT, and PEFCAT II

. 2021 May ; 7 (5) : 614-627. [epub] 20210428

Jazyk angličtina Země Spojené státy americké Médium print-electronic

Typ dokumentu klinické zkoušky, časopisecké články, multicentrická studie, práce podpořená grantem

Perzistentní odkaz   https://www.medvik.cz/link/pmid33933412
Odkazy

PubMed 33933412
DOI 10.1016/j.jacep.2021.02.014
PII: S2405-500X(21)00196-1
Knihovny.cz E-zdroje

OBJECTIVES: This study sought to determine whether durable pulmonary vein isolation (PVI) using pulsed field ablation (PFA) translates to freedom from atrial fibrillation recurrence without an increase in adverse events. BACKGROUND: PFA is a nonthermal ablative modality that, in preclinical studies, is able to preferentially ablate myocardial tissue with minimal effect on surrounding tissues. Herein, we present 1-year clinical outcomes of PFA. METHODS: In 3 multicenter studies (IMPULSE [A Safety and Feasibility Study of the IOWA Approach Endocardial Ablation System to Treat Atrial Fibrillation], PEFCAT [A Safety and Feasibility Study of the FARAPULSE Endocardial Ablation System to Treat Paroxysmal Atrial Fibrillation], and PEFCAT II [Expanded Safety and Feasibility Study of the FARAPULSE Endocardial Multi Ablation System to Treat Paroxysmal Atrial Fibrillation]), paroxysmal atrial fibrillation patients underwent PVI using a basket or flower PFA catheter. Invasive remapping was performed at ∼2 to 3 months, and reconnected PVs were reisolated with PFA or radiofrequency ablation. After a 90-day blanking period, arrhythmia recurrence was assessed over 1-year follow-up. RESULTS: In 121 patients, acute PVI was achieved in 100% of PVs with PFA alone. PV remapping, performed in 110 patients at 93.0 ± 30.1 days, demonstrated durable PVI in 84.8% of PVs (64.5% of patients), and 96.0% of PVs (84.1% of patients) treated with the optimized biphasic energy PFA waveform. Primary adverse events occurred in 2.5% of patients (2 pericardial effusions or tamponade, 1 hematoma); in addition, there was 1 transient ischemic attack. The 1-year Kaplan-Meier estimates for freedom from any atrial arrhythmia for the entire cohort and for the optimized biphasic energy PFA waveform cohort were 78.5 ± 3.8% and 84.5 ± 5.4%, respectively. CONCLUSIONS: PVI with a "single-shot" PFA catheter results in excellent PVI durability and acceptable safety with a low 1-year rate of atrial arrhythmia recurrence. These data mitigate concern that the nonthermal ablative mechanism of PFA might mask undiscovered compromises to clinical success. (IMPULSE: A Safety and Feasibility Study of the IOWA Approach Endocardial Ablation System to Treat Atrial Fibrillation, NCT03700385; A Safety and Feasibility Study of the FARAPULSE Endocardial Ablation System to Treat Paroxysmal Atrial Fibrillation, NCT03714178; PEFCAT II Expanded Safety and Feasibility Study of the FARAPULSE Endocardial Multi Ablation System to Treat Paroxysmal Atrial Fibrillation [PEFCAT II], NCT04170608).

Komentář v

PubMed

Citace poskytuje Crossref.org

Nejnovějších 20 citací...

Zobrazit více v
Medvik | PubMed

Atrial Fibrillation Ablation During Hospitalization for Acute Heart Failure: Feasibility and Role of Pulsed Field Ablation

. 2025 Jan ; 36 (1) : 256-265. [epub] 20241126

Pulmonary Vein Morphology in Patients Undergoing Catheter Ablation of Atrial Fibrillation

. 2024 Oct ; 15 (5) : 616-622. [epub] 20240618

Safety of pulsed field ablation in more than 17,000 patients with atrial fibrillation in the MANIFEST-17K study

. 2024 Jul ; 30 (7) : 2020-2029. [epub] 20240708

Comparing pulsed field electroporation and radiofrequency ablation for the treatment of paroxysmal atrial fibrillation: design and rationale of the BEAT PAROX-AF randomized clinical trial

. 2024 May 02 ; 26 (5) : .

Predictors of Success for Pulmonary Vein Isolation With Pulsed-field Ablation Using a Variable-loop Catheter With 3D Mapping Integration: Complete 12-month Outcomes From inspIRE

. 2024 May ; 17 (5) : e012667. [epub] 20240424

First-in-human clinical series of a novel conformable large-lattice pulsed field ablation catheter for pulmonary vein isolation

. 2024 Mar 30 ; 26 (4) : .

Long-term outcomes of the pentaspline pulsed-field ablation catheter for the treatment of paroxysmal atrial fibrillation: results of the prospective, multicentre FARA-Freedom Study

. 2024 Mar 01 ; 26 (3) : .

Catheter ablation for non-paroxysmal atrial fibrillation. A review

. 2024 Mar ; 168 (1) : 1-14. [epub] 20240115

Coronary Artery Spasm During Pulsed Field vs Radiofrequency Catheter Ablation of the Mitral Isthmus

. 2024 Jan 01 ; 9 (1) : 72-77.

Clinical Outcomes by Sex After Pulsed Field Ablation of Atrial Fibrillation

. 2023 Dec 01 ; 8 (12) : 1142-1151.

Reversible Pulsed Electrical Fields as an In Vivo Tool to Study Cardiac Electrophysiology: The Advent of Pulsed Field Mapping

. 2023 Oct ; 16 (10) : e012018. [epub] 20230920

Paroxysmal Atrial Fibrillation Ablation Using a Novel Variable-Loop Biphasic Pulsed Field Ablation Catheter Integrated With a 3-Dimensional Mapping System: 1-Year Outcomes of the Multicenter inspIRE Study

. 2023 Mar ; 16 (3) : e011780. [epub] 20230203

Multi-national survey on the methods, efficacy, and safety on the post-approval clinical use of pulsed field ablation (MANIFEST-PF)

. 2022 Sep 01 ; 24 (8) : 1256-1266.

How does the level of pulmonary venous isolation compare between pulsed field ablation and thermal energy ablation (radiofrequency, cryo, or laser)?

. 2021 Nov 08 ; 23 (11) : 1757-1766.

Zobrazit více v PubMed

ClinicalTrials.gov
NCT04170608

Najít záznam

Citační ukazatele

Nahrávání dat ...

Možnosti archivace

Nahrávání dat ...