AIMS: To report 5-year outcomes of EFFORTLESS registry patients with early generation subcutaneous implantable cardioverter-defibrillator (S-ICD) devices. METHODS AND RESULTS: Kaplan-Meier, trend and multivariable analyses were performed for mortality and late (years 2-5) complications, appropriate shock (AS) and inappropriate shock (IAS) rates. Nine hundred and eighty-four of 994 enrolled patients with diverse diagnoses (28% female, 48 ± 17 years, body mass index 27 ± 6 kg/m2, ejection fraction 43 ± 18%) underwent S-ICD implantation. Median follow-up was 5.1 years (interquartile range 4.7-5.5 years). All-cause mortality was 9.3% (95% confidence interval 7.2-11.3%) at 5 years; 703 patients remained in follow-up on study completion, 171 withdrew including 87 (8.8%) with device explanted, and 65 (6.6%) lost to follow-up. Of the explants, only 20 (2.0%) patients needed a transvenous device for pacing indications. First and final shock efficacy for discrete ventricular arrhythmias was consistent at 90% and 98%, respectively, with storm episode final shock efficacy at 95.2%. Time to therapy remained unaltered. Overall 1- and 5-year complication rates were 8.9% and 15.2%, respectively. Early complications did not predict later complications. There were no structural lead failures. Inappropriate shock rates at 1 and 5 years were 8.7% and 16.9%, respectively. Self-terminating inappropriately sensed episodes predicted late IAS. Predictors of late AS included self-terminating appropriately sensed episodes and earlier AS. CONCLUSION: In this diverse S-ICD registry population, spontaneous shock efficacy was consistently high over 5 years. Very few patients underwent S-ICD replacement with a transvenous device for pacing indications. Treated and self-terminating arrhythmic episodes predict future shock events, which should encourage more personalized device optimization.
- MeSH
- defibrilátory implantabilní * škodlivé účinky MeSH
- lidé MeSH
- náhlá srdeční smrt epidemiologie etiologie prevence a kontrola MeSH
- následné studie MeSH
- srdeční arytmie MeSH
- tepový objem MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) was developed to defibrillate ventricular arrhythmias, avoiding drawbacks of transvenous leads. The global EFFORTLESS S-ICD (Evaluation oF FactORs ImpacTing CLinical Outcome and Cost EffectiveneSS of the S-ICD) registry is collecting outcomes in 985 patients during a 5-year follow-up. OBJECTIVES: The primary goal of the EFFORTLESS registry is to determine the safety of the S-ICD by evaluating complications and inappropriate shock rate. METHODS: This is the first report on the full patient cohort and study endpoints with follow-up ≥1 year. The predefined endpoints are 30- and 360-day complications, and shocks for atrial fibrillation or supraventricular tachycardia. RESULTS: Patients were followed for 3.1 ± 1.5 years and 82 completed the study protocol 5-year visit. Average age was 48 years, 28% were women, ejection fraction was 43 ± 18%, and 65% had a primary prevention indication. The S-ICD system and procedure complication rate was 4.1% at 30 days and 8.4% at 360 days. The 1-year complication rate trended toward improvement from the first to last quartile of enrollment (11.3% [quartile 1]) to 7.8% [quartile 2], 6.6% [quartile 3], and 7.4% [quartile 4]; quartile 1 vs. quartiles 2 to 4; p = 0.06). Few device extractions occurred due to need for antitachycardia (n = 5), or biventricular (n = 4) or bradycardia pacing (n = 1). Inappropriate shocks occurred in 8.1% at 1 year and 11.7% after 3.1 years. At implant, 99.5% of patients had a successful conversion of induced ventricular tachycardia or ventricular fibrillation. The 1- and 5-year rates of appropriate shock were 5.8% and 13.5%, respectively. Conversion success for discrete spontaneous episodes was 97.4% overall. CONCLUSIONS: This registry demonstrates that the S-ICD fulfills predefined endpoints for safety and efficacy. Midterm performance rates on complications, inappropriate shocks, and conversion efficacy were comparable to rates observed in transvenous implantable cardioverter-defibrillator studies. (Evaluation oF Factors ImpacTing CLinical Outcome and Cost EffectiveneSS of the S-ICD [The EFFORTLESS S-ICD Registry]; NCT01085435).
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- komorová tachykardie terapie MeSH
- lidé středního věku MeSH
- lidé MeSH
- následné studie MeSH
- registrace * MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- pozorovací studie MeSH
AIMS: The subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced to overcome complications related to transvenous leads. Adoption of the S-ICD requires implanters to learn a new implantation technique. The aim of this study was to assess the learning curve for S-ICD implanters with respect to implant-related complications, procedure time, and inappropriate shocks (IASs). METHODS AND RESULTS: In a pooled cohort from two clinical S-ICD databases, the IDE Trial and the EFFORTLESS Registry, complications, IASs at 180 days follow-up and implant procedure duration were assessed. Patients were grouped in quartiles based on experience of the implanter and Kaplan-Meier estimates of complication and IAS rates were calculated. A total of 882 patients implanted in 61 centres by 107 implanters with a median of 4 implants (IQR 1,8) were analysed. There were a total of 59 patients with complications and 48 patients with IAS. The complication rate decreased significantly from 9.8% in Quartile 1 (least experience) to 5.4% in Quartile 4 (most experience) (P = 0.02) and non-significantly for IAS from 7.9 to 4.8% (P = 0.10). Multivariable analysis demonstrated a hazard ratio of 0.78 (P = 0.045) for complications and 1.01 (P = 0.958) for IAS. Dual-zone programming increased with experience of the individual implanter (P < 0.001), which reduced IAS significantly in the multivariable model (HR 0.44, P = 0.01). Procedure time decreased from 75 to 65 min (P < 0.001). The complication rate and procedure time stabilized after Quartile 2 (>13 implants). CONCLUSION: There is a short and significant learning curve associated with physicians adopting the S-ICD. Performance stabilizes after 13 implants.
- MeSH
- databáze faktografické MeSH
- defibrilátory implantabilní * MeSH
- dospělí MeSH
- elektrická defibrilace * MeSH
- implantace protézy výchova MeSH
- Kaplanův-Meierův odhad MeSH
- kardiochirurgické výkony výchova MeSH
- křivka učení * MeSH
- lidé středního věku MeSH
- lidé MeSH
- multivariační analýza MeSH
- následné studie MeSH
- proporcionální rizikové modely MeSH
- registrace MeSH
- rizikové faktory MeSH
- senioři MeSH
- výsledek terapie MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- pozorovací studie MeSH
The first clinical results from the Evaluation of Factors Impacting Clinical Outcome and Cost Effectiveness of the subcutaneous implantable cardioverter defibrillator (EFFORTLESS S-ICD) Registry on the entirely S-ICD system are promising, but the impact of the S-ICD system on patients' quality of life (QoL) is not known. We evaluated the QoL of patients with an S-ICD against an unrelated cohort with a transvenous (TV)-ICD system during 6 months of follow-up. Consecutively implanted patients with an S-ICD system were matched with patients with a TV-ICD system on a priori selected variables including baseline QoL. QoL was measured with the Short-Form Health Survey at baseline, 3, and 6 months after implant and compared using multivariable modeling with repeated measures. Patients with an S-ICD (n = 167) versus a TV-ICD system (n = 167) did not differ significantly on physical (p = 0.8157) and mental QoL scores (p = 0.9080) across baseline, 3, and 6 months after implantation in adjusted analyses. The evolution in physical (p = 0.0503) and mental scores (p = 0.3772) during follow-up was similar for both cohorts, as indicated by the nonsignificant interaction effect for ICD system by time. Both patients with an S-ICD system and a TV-ICD system experienced significant improvements in physical and mental QoL between time of implant and 3 months (both p's <0.0001) and between time of implant and 6 months (both p's <0.0001) but not between 3 and 6 months (both p's >0.05). In conclusion, these first results show that the QoL of patients with an S-ICD versus TV-ICD system is similar and that patients with either system experience improvements in QoL on the short term.
- MeSH
- činnosti denního života MeSH
- defibrilátory implantabilní * MeSH
- dospělí MeSH
- fibrilace komor komplikace terapie MeSH
- implantace protézy metody MeSH
- komorová tachykardie komplikace terapie MeSH
- kvalita života * MeSH
- lidé středního věku MeSH
- lidé MeSH
- multivariační analýza MeSH
- náhlá srdeční smrt etiologie prevence a kontrola MeSH
- prospektivní studie MeSH
- registrace * MeSH
- senioři MeSH
- Check Tag
- dospělí MeSH
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- pozorovací studie MeSH
- srovnávací studie MeSH