Validity of six consumer-level activity monitors for measuring steps in patients with chronic heart failure
T. Vetrovsky, M. Siranec, J. Marencakova, JJ. Tufano, V. Capek, V. Bunc, J. Belohlavek,
Jazyk angličtina Země Spojené státy americké
Typ dokumentu časopisecké články, práce podpořená grantem
NLK
Directory of Open Access Journals
od 2006
Free Medical Journals
od 2006
Public Library of Science (PLoS)
od 2006
PubMed Central
od 2006
Europe PubMed Central
od 2006
ProQuest Central
od 2006-12-01
Open Access Digital Library
od 2006-01-01
Open Access Digital Library
od 2006-01-01
Open Access Digital Library
od 2006-10-01
Medline Complete (EBSCOhost)
od 2008-01-01
Nursing & Allied Health Database (ProQuest)
od 2006-12-01
Health & Medicine (ProQuest)
od 2006-12-01
Public Health Database (ProQuest)
od 2006-12-01
ROAD: Directory of Open Access Scholarly Resources
od 2006
- MeSH
- akcelerometrie přístrojové vybavení metody MeSH
- ambulantní monitorování přístrojové vybavení metody MeSH
- chronická nemoc MeSH
- chůze fyziologie MeSH
- cvičení fyziologie MeSH
- dospělí MeSH
- fitness náramky MeSH
- lidé MeSH
- reprodukovatelnost výsledků MeSH
- senioři MeSH
- srdeční selhání patofyziologie MeSH
- zátěžový test přístrojové vybavení metody MeSH
- Check Tag
- dospělí MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
INTRODUCTION: Although numerous activity trackers have been validated in healthy populations, validation is lacking in chronic heart failure patients who normally walk at a slower pace, making it difficult for researchers and clinicians to implement activity monitors during physical activity interventions. METHODS: Six consumer-level activity monitors were validated in a 3-day field study in patients with chronic heart failure and healthy individuals under free living conditions. Furthermore, the same devices were evaluated in a lab-based study during treadmill walking at speeds of 2.4, 3.0, 3.6, and 4.2 km·h-1. Concordance correlation coefficients (CCC) were used to evaluate the agreement between the activity monitors and the criterion, and mean absolute percentage errors (MAPE) were calculated to assess differences between each device and the criterion (MAPE <10% was considered as a threshold for validity). RESULTS: In the field study of healthy individuals, all but one of the activity monitors showed a substantial correlation (CCC ≥0.95) with the criterion device and MAPE <10%. In patients with heart failure, the correlation of only two activity monitors (Garmin vívofit 3 and Withings Go) was classified as at least moderate (CCC ≥0.90) and none of the devices had MAPE <10%. In the lab-based study at speeds 4.2 and 3.6 km·h-1, all activity monitors showed substantial to almost perfect correlations (CCC ≥0.95) with the criterion and MAPE in the range 1%-3%. However, at slower speeds of 3.0 and 2.4 km·h-1, the accuracy of all devices substantially deteriorated: their correlation with the criterion decreased below 90% and their MAPE increased to 4-8% and 10-45%, respectively. CONCLUSIONS: Even though none of the tested activity monitors fall within arbitrary thresholds for validity, most of them perform reasonably well enough to be useful tools that clinicians can use to simply motivate chronic heart failure patients to walk more.
2nd Faculty of Medicine Charles University Prague Czech Republic
Faculty of Physical Education and Sport Charles University Prague Czech Republic
Citace poskytuje Crossref.org
- 000
- 00000naa a2200000 a 4500
- 001
- bmc20005910
- 003
- CZ-PrNML
- 005
- 20200518132203.0
- 007
- ta
- 008
- 200511s2019 xxu f 000 0|eng||
- 009
- AR
- 024 7_
- $a 10.1371/journal.pone.0222569 $2 doi
- 035 __
- $a (PubMed)31518367
- 040 __
- $a ABA008 $b cze $d ABA008 $e AACR2
- 041 0_
- $a eng
- 044 __
- $a xxu
- 100 1_
- $a Vetrovsky, Tomas $u Faculty of Physical Education and Sport, Charles University, Prague, Czech Republic.
- 245 10
- $a Validity of six consumer-level activity monitors for measuring steps in patients with chronic heart failure / $c T. Vetrovsky, M. Siranec, J. Marencakova, JJ. Tufano, V. Capek, V. Bunc, J. Belohlavek,
- 520 9_
- $a INTRODUCTION: Although numerous activity trackers have been validated in healthy populations, validation is lacking in chronic heart failure patients who normally walk at a slower pace, making it difficult for researchers and clinicians to implement activity monitors during physical activity interventions. METHODS: Six consumer-level activity monitors were validated in a 3-day field study in patients with chronic heart failure and healthy individuals under free living conditions. Furthermore, the same devices were evaluated in a lab-based study during treadmill walking at speeds of 2.4, 3.0, 3.6, and 4.2 km·h-1. Concordance correlation coefficients (CCC) were used to evaluate the agreement between the activity monitors and the criterion, and mean absolute percentage errors (MAPE) were calculated to assess differences between each device and the criterion (MAPE <10% was considered as a threshold for validity). RESULTS: In the field study of healthy individuals, all but one of the activity monitors showed a substantial correlation (CCC ≥0.95) with the criterion device and MAPE <10%. In patients with heart failure, the correlation of only two activity monitors (Garmin vívofit 3 and Withings Go) was classified as at least moderate (CCC ≥0.90) and none of the devices had MAPE <10%. In the lab-based study at speeds 4.2 and 3.6 km·h-1, all activity monitors showed substantial to almost perfect correlations (CCC ≥0.95) with the criterion and MAPE in the range 1%-3%. However, at slower speeds of 3.0 and 2.4 km·h-1, the accuracy of all devices substantially deteriorated: their correlation with the criterion decreased below 90% and their MAPE increased to 4-8% and 10-45%, respectively. CONCLUSIONS: Even though none of the tested activity monitors fall within arbitrary thresholds for validity, most of them perform reasonably well enough to be useful tools that clinicians can use to simply motivate chronic heart failure patients to walk more.
- 650 _2
- $a akcelerometrie $x přístrojové vybavení $x metody $7 D061725
- 650 _2
- $a dospělí $7 D000328
- 650 _2
- $a senioři $7 D000368
- 650 _2
- $a chronická nemoc $7 D002908
- 650 _2
- $a cvičení $x fyziologie $7 D015444
- 650 _2
- $a zátěžový test $x přístrojové vybavení $x metody $7 D005080
- 650 _2
- $a ženské pohlaví $7 D005260
- 650 _2
- $a fitness náramky $7 D000072936
- 650 _2
- $a srdeční selhání $x patofyziologie $7 D006333
- 650 _2
- $a lidé $7 D006801
- 650 _2
- $a mužské pohlaví $7 D008297
- 650 _2
- $a ambulantní monitorování $x přístrojové vybavení $x metody $7 D018670
- 650 _2
- $a reprodukovatelnost výsledků $7 D015203
- 650 _2
- $a chůze $x fyziologie $7 D016138
- 655 _2
- $a časopisecké články $7 D016428
- 655 _2
- $a práce podpořená grantem $7 D013485
- 700 1_
- $a Siranec, Michal $u 2nd Department of Medicine - Department of Cardiovascular Medicine, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic.
- 700 1_
- $a Marencakova, Jitka $u Faculty of Physical Education and Sport, Charles University, Prague, Czech Republic.
- 700 1_
- $a Tufano, James J $u Faculty of Physical Education and Sport, Charles University, Prague, Czech Republic.
- 700 1_
- $a Capek, Vaclav $u Second Faculty of Medicine, Charles University, Prague, Czech Republic.
- 700 1_
- $a Bunc, Vaclav $u Faculty of Physical Education and Sport, Charles University, Prague, Czech Republic.
- 700 1_
- $a Belohlavek, Jan $u 2nd Department of Medicine - Department of Cardiovascular Medicine, First Faculty of Medicine, Charles University and General University Hospital in Prague, Prague, Czech Republic.
- 773 0_
- $w MED00180950 $t PloS one $x 1932-6203 $g Roč. 14, č. 9 (2019), s. e0222569
- 856 41
- $u https://pubmed.ncbi.nlm.nih.gov/31518367 $y Pubmed
- 910 __
- $a ABA008 $b sig $c sign $y a $z 0
- 990 __
- $a 20200511 $b ABA008
- 991 __
- $a 20200518132203 $b ABA008
- 999 __
- $a ok $b bmc $g 1524768 $s 1095966
- BAS __
- $a 3
- BAS __
- $a PreBMC
- BMC __
- $a 2019 $b 14 $c 9 $d e0222569 $e 20190913 $i 1932-6203 $m PLoS One $n PLoS One $x MED00180950
- LZP __
- $a Pubmed-20200511